CliniEco Medical Biological Indicator — vial box, 50-pack

CliniEco Medical Biological Indicator — 1-Hour Rapid Readout, Fluorescence, 50-Pack

$229.99 CAD
Sale price  $229.99 CAD Regular price 
Skip to product information
CliniEco Medical Biological Indicator — vial box, 50-pack

CliniEco Medical Biological Indicator — 1-Hour Rapid Readout, Fluorescence, 50-Pack

$229.99 CAD
Sale price  $229.99 CAD Regular price 
In stock

🚚 Free Shipping over $99 (Ontario) · $199 (rest of Canada)

Prefer a quote? B2B & facility pricing — bulk discounts and standing orders on request. We reply within 1 business day.

⭐ Bulk savings
Buy 5+ units → save 5%
Buy 10+ units → save 10%
Orders $500+ → save 10%
Auto-applied at checkout — most popular: bulk case orders

Rapid-readout sterilization monitoring — results in 1 hour, not 24 hours.

Each vial contains Geobacillus stearothermophilus ATCC 7953 spores in a self-contained format. After the sterilization cycle the vial is activated and incubated, and the result is read by fluorescence — a positive (failed) and negative (passed) result is reported by the reader, not by eye.

How it differs from our 24-hour indicator

  • Readout: 1 hour instead of 24 hours
  • Reader required — this model is read on the CliniEco Rapid Reader (4-well fluorescence reader). It is not a colour-change product, so it cannot be interpreted visually.
  • Same spore strain and same ISO reference as our 24-hour line, so both can sit in one monitoring programme.

Specifications

Model ZJ60
Readout time 1 hour
Format Self-contained vial, rapid readout (enzymatic / fluorescence)
Organism Geobacillus stearothermophilus ATCC 7953
Reference standard ISO 11138 and ISO 11140 (per manufacturer specification)
Incubation 58 ± 2 °C (use with the CliniEco Rapid Reader)
Pack size 50 vials per box
Intended use Routine monitoring of steam (moist heat) sterilization cycles
Storage Follow the manufacturer's instructions supplied with the box; keep the vials in their original packaging and do not add any label or tape to the vial or cap

Readout times across our biological indicator line

Model Readout Result read by
Rapid 1 小时 (ZJ60) 1 hour CliniEco Rapid Reader
Rapid 20 minutes (ZJ20) 20 minutes CliniEco Rapid Reader
Rapid 1 hour (ZJ60) 1 hour CliniEco Rapid Reader
Rapid 3 hours (ZJ180) 3 hours CliniEco Rapid Reader
24-hour self-contained (ZJ1440) Up to 24 hours (early read at 8 / 12 h) Visual colour change

Who it is for

Dental clinics, medical clinics and laboratories that run daily steam sterilization and need a same-day, documented result — for example where a load cannot be released until the biological indicator reads negative.

What you need with it

  • CliniEco Rapid Reader — 4-well fluorescence reader (required for this model)
  • Incubation at 58 ± 2 °C
  • Batch record: record the vial lot number against the sterilizer load

Documentation

Manufacturer's instructions for use are supplied with every box. A certificate of analysis per batch is available on request — the lot number on the box is traceable to the release documentation.

How to run a test

  1. Place the indicator in the load, in the area that is hardest to sterilize (for a typical dental load: with the instruments in the pouch or cassette).
  2. Run the sterilization cycle as normal.
  3. After the cycle, follow the vial activation step described in the manufacturer's instructions — the vial is activated, then incubated.
  4. Place the vial in the CliniEco Rapid Reader and incubate at 58 ± 2 °C. The reader reports the result at the readout time above.
  5. Record the result and the vial lot number against the sterilizer load log.

One indicator per load, every day. Do not add any label or tape to the vial or its cap — it interferes with the incubation and read.

Interpreting the result

  • Negative (pass) — the spores were killed; the load may be released.
  • Positive (fail) — spores survived. Do not release the load: re-clean, re-pack and re-sterilize, and check the sterilizer before using it again. Retain the indicator and the load record.
  • This model is read by the reader only. Unlike our 24-hour indicator, there is no colour change to read by eye — the result must come from the reader.

Frequently asked

Can I read it by eye? No. Rapid-readout models are read on the fluorescence reader.
Do I need the reader as well? Yes — the reader is required for this model. Our 24-hour self-contained indicator does not need a reader (visual colour change).
Why choose 20 minutes over the 24-hour indicator? When the load cannot be released until the result is known — same-day reprocessing, single-set practices, or where a 24-hour hold is operationally not possible.
Is it the same organism as the 24-hour indicator? Yes, Geobacillus stearothermophilus ATCC 7953 under ISO 11138-3 — so both can be used in one monitoring programme.
Storage? As printed on the box and in the manufacturer's instructions. Keep vials in the original packaging and do not relabel them.

Official manufacturer specifications

Readout time 1 hour
Incubation range 15–60 minutes
Organism Geobacillus stearothermophilus ATCC 7953
Population ≥ 1.0 × 10⁶ c.f.u. (mean per strip)
D-value ≥ 1.5 min at 121 °C (steam)
Test conditions 121 °C / 132 °C / 134 °C saturated steam — vacuum and gravity cycles
Standards ISO 11138 and ISO 11140 compliant (manufacturer declaration)
Pack size 50 vials per pack
Frequency of use One test per sterilization cycle
Kit contents BI ampoules and culture media tubes

Parameters as published by the manufacturer. Batch certificate of analysis is traceable from the lot number on the box.

INTCO Nitrile Gloves — 1,000 Gloves per Case (10 Boxes × 100) | Touch Flex / Advancare Wholesale