An autoclave spore test, or biological indicator, answers one question: did this cycle actually kill the toughest organism we can measure? Self-contained indicators are read in about 24 hours, conventional paper strips take up to seven days, and a positive result stops load release until the cause is found. Most questions Canadian clinics ask are about incubation time, warm starts, controls and failures.

Quick facts
- Ontario expects a biological indicator for each sterilizer on every day it is used.
- Self-contained spore tests are read in about 24 hours; conventional paper strips take up to seven days.
- Match the indicator to the cycle: steam, ethylene oxide and dry heat each need their own type.
- A failed result suspends load release until the cause is corrected and the sterilizer re-tests clean.
How Long Does a Spore Test Take, and Why Timing Matters
Incubation time is set by the indicator maker, not by the clinic. Self-contained products are read at their stated point, commonly around 24 hours for a colour-change vial, while conventional paper spore strips need up to seven days of incubation before a final negative can be recorded. The practical rule is to follow the instructions for use and never shorten the window because the morning is busy.
What is a warm start, and why does it matter?
A warm start means the sterilizer begins a cycle with the chamber already hot from a previous load. It changes air removal and condensate behaviour, so the cycle should be validated for that condition. A cold start, from an idle chamber in the morning, is the harder test of air removal. Log which condition applies to each test.
Do you need a positive control with every run?
A positive control is an unexposed indicator from the same lot, incubated beside the test unit to prove that the indicator and the incubator actually work. Without it, a negative result only proves that something did not grow. Incubate the control for the same window as the test, and record the lot number alongside the result.
Spore Test Decision Points at a Glance
| Situation | Action | Timing |
|---|---|---|
| Routine monitoring in Ontario | Indicator placed in a process challenge device, run with a normal cycle | Every day the sterilizer is used |
| Weekly baseline reference | One indicator per sterilizer per week, the CDC and AAMI ST79 reference point | Contrast only, not the Ontario rule |
| Positive control | Incubate an unexposed control from the same lot | Same incubation window |
| Failed result | Stop releasing loads, quarantine the load, investigate | Immediately on readout |
| Record keeping | Log load, cycle, indicator lot, result and operator | Retain per provincial guidance |
How Often to Run a Spore Test in Ontario
In Ontario, the RCDSO requires a biological indicator for each sterilizer on every day it is used, daily monitoring, not weekly. The weekly baseline is the CDC and AAMI ST79 reference used in most American states. Five days a week for fifty weeks works out to roughly 250 indicators per sterilizer per year, which is why annual quantities surprise clinics that budget for a monthly test.
The requirement is set out in our answer on whether you need a biological indicator for every load in Ontario, and the device in the note on why Ontario dental offices use a process challenge device. See also AAMI ST79 versus CSA Z314, compared line by line.
What to Do After a Failed Spore Test
A failed indicator is a stop signal, not a paperwork problem. Remove the sterilizer from service, quarantine every load processed since the last passing test, and reprocess those loads before they reach a patient. Then check the usual causes: loading pattern, air removal, cycle parameters, the indicator lot and the incubator. Do not return the sterilizer to service until a repeat test passes.

Choosing Indicators and Readout Equipment
Buy the format that matches your sterilizer and workflow. A 24-hour self-contained biological indicator 25-pack suits clinics that need a same-day answer, and a rapid-reader seed trial lets a clinic test the readout workflow before committing. Monitoring questions sit under the sterilization and compliance hub.
Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers.
Related reading
- Do you need a biological indicator for every load? Ontario answers
- PCD explained: the process challenge device Ontario offices use
- AAMI ST79 vs CSA Z314 monitoring requirements compared
- Rapid Readout Biological Indicator Readers: Every Major Brand Compared (2026 Canada Price Guide)
- BI vs Chemical Indicators: What's the Difference in Sterilization Monitoring?
- The Real Annual Cost of Biological Monitoring: Locked Ecosystems vs Open-System BIs in Canadian Dental Clinics
- free Ontario sterilization compliance log
- printable autoclave sterilization log sheet (free)
Frequently Asked Questions
How long does a spore test take to incubate?
Self-contained indicators are read at their stated point, commonly around 24 hours, while conventional paper spore strips need up to seven days of incubation before a final negative can be recorded. Always follow the indicator maker's instructions for use.
Do I need a positive control with every run?
Yes. An unexposed control from the same lot proves the indicator and the incubator are working, so a negative test result actually means something. Incubate it for the same window and log the lot number.
What should a clinic do after a failed spore test?
Stop releasing loads, quarantine everything processed since the last passing test, and reprocess those loads. Investigate loading, air removal, cycle parameters, the indicator lot and the incubator before returning the sterilizer to service.
how often to do spore test in dental office rcdso
In Ontario, the RCDSO requires a biological indicator for each sterilizer on every day it is used, which is daily monitoring, not weekly. The weekly baseline belongs to the CDC and AAMI ST79 reference used in most American states. At five days a week for fifty weeks that is roughly 250 indicators per sterilizer per year.