The answer you hear most often — “one spore test a week” — is neither entirely wrong nor valid everywhere. In Canada, the frequency of a biological indicator (BI) is set by the professional college or health authority of each province, not by a single national rule. A clinic that copies the instruction from another province, or from a supplier, can believe it is compliant when it is not.
This guide sets out the Quebec rule, the Ontario rule, and where the confusion comes from, so you can check your own — sterilizer by sterilizer, and cycle type by cycle type.
What frequency does Quebec require?
In Quebec, the guidelines of the Ordre des dentistes du Québec (ODQ) on infection prevention and control and the reprocessing of medical devices (revision February 2026) call for a verification of every Class N, B or S sterilizer — and for each type of cycle used — by means of a process challenge device (PCD) containing a biological indicator and a Type 5 chemical indicator.
The text is explicit: daily verification is recommended, but it is weekly verification that is required, for each type of cycle used. The two controls (BI and Type 5 chemical indicator) run in two stages; they cannot be performed in the same cycle. Where non-sterile implantable devices are involved, a PCD fitted with a BI must also accompany each affected load.
And in Ontario?
Ontario follows a stricter rule. The Royal College of Dental Surgeons of Ontario (RCDSO) requires a biological indicator each day the sterilizer is used, and for each type of cycle used that day: daily monitoring, not weekly. For Class B sterilizers, a Bowie-Dick air-removal test is also required at the first cycle of the day.
In other words, an Ontario clinic and a Quebec clinic do not share the same monitoring calendar — even though they work with the same instruments and the same sterilizer types.
Why do you so often hear “once a week”?
The weekly benchmark comes from the United States: the AAMI ST79 guide and CDC recommendations keep a weekly baseline for most American facilities. Because much of the material online mixes American and Canadian sources, that baseline is often presented as a general rule — when in Canada the frequency is fixed at the provincial level, and it may be stricter than that baseline.
The right question is therefore not “is it every week or every day?”, but “what does my province require, for each of my sterilizers and each of my cycle types?”
| Territory | Biological indicator frequency | Scope | Source |
|---|---|---|---|
| Quebec | Weekly (required) — daily recommended | For each sterilizer and each type of cycle used | ODQ, PCI and reprocessing guidelines (rev. February 2026) |
| Ontario | Each day the sterilizer is used | For each sterilizer and each type of cycle used that day | RCDSO infection prevention and control standard |
| United States (baseline) | Weekly | For most facilities | AAMI ST79 and CDC recommendations |
| Implantable devices | Each affected load | In addition to the base frequency | ODQ (Quebec) |
What does not change from one province to the next?
- An internal chemical indicator inside every package, and an external indicator (tape) on every package.
- An up-to-date sterilizer log: date, sterilizer code, load number and control result.
- A process challenge device placed at the hardest-to-sterilize location in the chamber.
- Following the sterilizer manufacturer’s instructions for loading, weight and maintenance.
For a clinic looking for a starting point, a 24-hour readout biological indicator (box of 25 tests) suits a weekly cadence, while a 1-hour fluorescence biological indicator suits a site that needs to release loads faster.
How to set your own monitoring schedule
Start by counting your sterilizers and listing your cycle types (wrapped, unwrapped, and so on). Then apply your province’s frequency to each “sterilizer × cycle type” pair, and multiply to get an annual need for indicators.
Clinics that supply several rooms usually order through an institutional wholesale account or a B2B purchasing group rather than buying pack by pack. Reselling? become a distributor.
How to read a biological indicator result
A biological indicator contains a defined population of spores — the microorganisms most resistant to the sterilization process. A negative result means those spores were inactivated. Because they survive better than the other pathogens present on instruments, a negative indicator means those other agents were destroyed too. The type of indicator must match the sterilizer type in use (Class N, B or S).
Rapid formats are read in a reader rather than by eye: a 3-hour rapid-readout biological indicator is read in a 4-well fluorescence reader, while a 24-hour self-contained indicator is read by a colour change.
- Follow the manufacturer’s instructions for incubation and reading; a rapid-readout indicator needs a dedicated reader.
- A control indicator incubated alongside helps interpret the test result.
- Record the result in the log with the date, sterilizer code and load number.
- Where your protocol requires it, do not release a risk load before you have the result.
What to do if a biological indicator test fails
A failure is not an administrative detail: it goes straight to confidence in the process. The sequence is always the same.
- Withdraw the affected load from circulation and identify the devices it contains.
- Reprocess after correcting the probable cause — loading, residual air, maintenance, wrong cycle.
- Record the failure in the sterilization failure log, with the list of recalled devices and the resolution applied.
- Check maintenance and, if necessary, calibration of the sterilizer before returning it to service.
Finally, keep in mind that a biological indicator does not replace any other layer of monitoring: the internal chemical indicator in every package, the external indicator, the air-removal test for Class B sterilizers, and the sterilizer log. It is their combination — not any one of them — that documents the compliance of the process.
Looking to go further? Download ready-to-print log templates in the compliance log centre and score your own practice with the free sterilization self-check. For guidance matched to your situation, talk to a sterilization compliance specialist.
Related Reading
- How often to run a spore test in a dental office (RCDSO)
- Biological indicator formats compared
- Biological indicator, 20-minute rapid readout (50 tests)
Related Reading
- How often to run a spore test in a dental office (RCDSO)
- Biological indicator formats compared
- Ready-to-print log templates in the compliance log centre
- Talk to a sterilization compliance specialist
Frequently Asked Questions
Is one spore test a week enough in Quebec?
For a clinic in Quebec, the baseline required by the ODQ is weekly, but it applies per sterilizer and per type of cycle used, and daily verification is recommended. The phrase “once a week” is therefore not a complete answer unless both variables are stated.
Does the same rule apply everywhere in Canada?
No. The frequency is set province by province: Ontario requires a verification on each day of use, while Quebec keeps a weekly baseline with a daily recommendation. Check the rule of your own province and professional college.
Does the Bowie-Dick test replace the biological indicator?
No. The Bowie-Dick test checks air removal in a Class B sterilizer; it does not replace the biological indicator, which is the only test that verifies spore destruction. They are two complementary controls.
How many biological indicators should you plan for in a year?
The calculation depends on the number of sterilizers, the number of cycle types used, and the frequency required in your province. In Ontario it is calculated on days of use; in Quebec, on a weekly basis per sterilizer and per cycle type.