How Should a Midwifery Clinic Reprocess Reusable Instruments in Canada?

How Should a Midwifery Clinic Reprocess Reusable Instruments in Canada?

Midwifery is the one community practice where contaminated surgical instruments arrive by car. A pair of episiotomy scissors used in a client's bedroom at two in the morning is a critical device with blood on it, and it will sit in a sealed container in a vehicle for however long the midwife takes to get back to the clinic. That interval is the part of reprocessing that no dental office has to think about, and it is where midwifery reprocessing procedures most often fail.

The instruments themselves are defined by the scope of practice. Ontario's Midwifery Act, 1991 authorises midwives to manage labour and conduct spontaneous normal vaginal deliveries, to insert urinary catheters, to perform episiotomies and amniotomies and repair episiotomies and lacerations that do not involve the anus, anal sphincter, rectum, urethra or periurethral area, to take blood samples, and to put an instrument, hand or finger beyond the labia majora during pregnancy, labour and the postpartum period. Each of those authorisations implies a device, and each device has a reprocessing consequence.

The College of Midwives of Ontario sets the professional frame. Its Professional Standards for Midwives, approved by Council on 21 March 2018 and implemented on 1 June 2018, apply to every midwife in every setting, and the standards for practice owners go further: a practice owner must ensure essential operational and clinical supplies are available to the midwives in the practice and must develop and maintain quality improvement systems. A contaminated instrument with nowhere to be cleaned is a supplies and systems failure before it is a clinical one.

The consumables that make the loop work are ordinary, and buying them by the case is what makes the loop repeatable: a 4 litre benchtop ultrasonic cleaner running dual frequency at 32 and 48 kHz for the cleaning step, and a constant-heat pouch sealer producing a 12 mm seal at 200 degrees so the packaging step produces the same seam every time rather than whatever the iron produced.

Which instruments does a midwifery practice actually reprocess?

Start from the Spaulding classification, because the level of treatment is set by where the device goes, not by what it is called. The Ontario Midwives equipment guidance states the rule that resolves most arguments: where the level recommended by the manufacturer is lower than the level the intended use requires by Spaulding's criteria, the higher level must be used. Manufacturer convenience never wins over contact site.

Instrument or device Spaulding class Minimum treatment Usual midwifery choice
Episiotomy scissors, suture scissors, needle holder Critical — enters tissue Cleaning followed by sterilization Reusable, sterilized; single-use suture packs for the repair itself
Vaginal speculum Semi-critical — contacts mucous membrane Sterilization, or high-level disinfection at minimum The provincial equipment guidance lists sterilization or single-use as the standard for a speculum
Metal catheter or reusable catheter tray Critical or semi-critical depending on use Sterilization Single-use catheters are the ordinary choice and remove the question
Cord clamp, cord scissors in a single-use birth kit Critical, single-use Sterile as supplied, discarded after use Single-use; not reprocessed
Doppler and probe Non-critical Low-level disinfection between clients Remove ultrasound gel before disinfecting; the gel bottle is dated and discarded after one month
Fetoscope, stethoscope, sphygmomanometer cuff Non-critical Low-level disinfection between clients Own equipment per midwife where possible; shared earpieces are disinfected
Pinard horn Non-critical, but porous Not effectively disinfectable in wood The equipment guidance is explicit that a wooden Pinard horn cannot be effectively disinfected
Thermometer, including one used in a birth pool Non-critical Low-level disinfection every use Probe covers plus disinfection

Two rows in that table are the ones midwives argue with. The first is the speculum: it contacts mucous membrane, so it is semi-critical, and a wipe with a disinfectant cloth does not meet the level. The second is the Pinard horn, which is still in some birth bags because it works and because it is quiet. A wooden horn cannot be effectively disinfected, and the fix is not a better wipe; it is a different device.

A constant-heat pouch sealer with a pouch roll mounted on the spindle and a 12 mm sealing bar
A fixed seal width and a fixed temperature remove the variable that hand sealing leaves in.

How do instruments get from a living room back to the bench?

The transport step is where a midwifery procedure differs from every office-based reprocessing policy borrowed from dentistry. Contaminated instruments should be contained at the point of use, kept from drying where the manufacturer's instructions allow it, and transported in a closed, labelled container so that nobody in the vehicle handles them and so that cleaning starts within the interval the clinic's procedure defines. Instruments that arrive dry and crusted after twelve hours are harder to clean than instruments that arrive wet, and harder cleaning means more manual scrubbing, which is the step most associated with sharps injuries.

A 4 litre benchtop ultrasonic cleaner with a stainless steel tank, lid handle and a single front control knob
A benchtop ultrasonic cleaner handles the cleaning step that hand scrubbing cannot reach, and its log is one of the three reprocessing records an inspector asks for.

The second half of the transport question is the vehicle. A sealed container in a car that sits in a summer parking lot reaches temperatures that will cook biological soil onto the instruments and, if the container holds anything packaged, degrade the packaging. The written procedure should say that instruments travel in the passenger compartment and come into the clinic on arrival rather than staying in the vehicle until the end of the day.

What does the reprocessing loop look like step by step?

Six steps, each with a failure mode worth writing down, because the record only protects the clinic if it identifies which step failed when something goes wrong.

Step What happens Failure mode Record
Point of use Device contained, kept moist where permitted, sharps separated Instrument dropped loose into a birth bag Nothing at this step; the exposure incident file if a sharp injures someone
Transport Closed labelled container, passenger compartment, arrives with the midwife Container left in the vehicle overnight Procedure document; no per-trip log
Cleaning Manual wash plus ultrasonic, or a washer-disinfector, per the manufacturer Instruments go straight into the sterilizer Ultrasonic or washer log per day of use
Inspection and packaging Dry, inspected, hinged instruments opened, pouched or wrapped with a chemical indicator Wet instruments sealed into a pouch Packaging lot and indicator on the pack label
Sterilization Cycle appropriate to the load, monitored with chemical and biological indicators Cycle type chosen for speed rather than for the load Sterilization log with contents, parameters, indicators, person responsible
Storage and release Dated pack into clean storage; quarantine for any implant or device held for a biological result Pack released before the biological indicator read negative Storage location and daily temperature record; release note on the load line

The cleaning step deserves a sentence of its own because it is the step midwives most often compress. An ultrasonic cleaner does not replace manual cleaning of hinged instruments; it removes soil from surfaces and from the box joints that a brush cannot reach, provided the instruments are opened and the solution is changed on the interval the manufacturer specifies. A washer-disinfector does both cleaning and thermal disinfection in one pass and produces its own record, which is why practices with enough volume move to one.

What packaging does a midwifery practice need?

Packaging has to do three things: keep the contents sterile until the pack is opened, allow the sterilant in and out, and carry an indicator that shows the pack was exposed to the process. For a small practice, pouches and rolls are the practical answer, and the standard behind them is ISO 11607-1 for packaging systems, with the European pouch standard EN 868-5 describing the paper-film construction that most Canadian clinics actually buy.

The packaging decision connects to storage. A pack that is sealed with a 12 mm continuous seal and stored in a closed cabinet in a room whose temperature and humidity are recorded will hold its claim until the pack is damaged or the seal fails. A pack sealed with a seal that has channels in it will fail the seal integrity check, and the failure will not be visible until someone opens it in front of a client. That is why the sealer, not the pouch, is the device worth standardising: a constant-heat sealer with a fixed temperature and a fixed seal width removes a variable that a hand-held impulse sealer leaves in.

Which reprocessing records does the practice keep?

Three, and they match the three that the Public Health Ontario IPAC checklist grades for clinical office practice. The sterilization log is item 1.1 and is a legislated requirement: load control label with sterilizer number, load number and date; the chart or printout of physical parameters; the load contents; the person responsible; the chemical indicator results; and the biological indicator results. The high-level disinfection log is item 1.2 and is graded high risk: date and time, contact time, and the person performing it. The ultrasonic cleaner and washer-disinfector logs are item 1.3 and follow the manufacturer's instructions for use.

A midwifery practice that uses single-use devices for most procedures still needs at least two of those three. If a speculum is single-use and a suture pack is single-use, the remaining reusable items are the episiotomy scissors and any metal instrument in the birth bag, and those require the sterilization log every time the sterilizer runs. The high-level disinfection log appears the moment a semi-critical reusable item stays in service — a metal speculum, for instance, or a reusable probe cover.

Filing these on paper is acceptable and common in small practices. What is not acceptable is a log that lives in one midwife's bag. The practice owner standard about quality improvement systems is the hook: the record has to belong to the practice, be readable by whoever is on call, and survive a midwife leaving. A printable compliance log sheet is enough to start, provided the completed sheets go into a file the practice owns.

Where does a clinic take this next?

Four things a clinic can act on rather than read: a printable compliance log sheet that records receiving dates, lot numbers and open-on dates and exports from the browser; the clinic supply planner that turns daily bench counts into cases per month; and, where a finding has to be documented for an inspector, Ask your compliance question which returns a written answer citing the regulation or standard it rests on.

Facility and multi-site buyers can also ask about wholesale and multi-site ordering for case pricing and account setup. Reselling? become a distributor.

Related reading

Frequently Asked Questions

Can instruments from a home birth be cleaned at the client's home?

Manual cleaning at the point of use is limited to what the manufacturer's instructions allow and to what can be done without turning a client's kitchen into a reprocessing area. The realistic option in a home is containment: keep the instrument moist where permitted, seal it in a labelled container, and get it to a proper sink, ultrasonic cleaner or washer-disinfector within the interval your procedure defines. Cleaning that happens in a domestic sink without the right detergent, water temperature and drying is not a reprocessing step that can be documented, and a device that cannot be documented cannot be defended.

Does a midwifery practice need a washer-disinfector, or is an ultrasonic enough?

An ultrasonic cleaner plus manual cleaning of hinges and lumens is enough for low volume, provided the solution is changed on the manufacturer's interval and the instruments are dried before packaging. A washer-disinfector adds repeatability and produces its own cycle record, which is useful when several midwives share a practice. Neither replaces the inspection step: a hinged instrument that comes out of either machine with soil in the box joint has not been cleaned, whatever the log says.

Is a domestic pressure cooker acceptable for sterilizing instruments?

No. A pressure cooker is not a medical device sterilizer, it does not have a validated cycle, and it produces no physical parameter record and no way to place a process challenge device reproducibly. Canadian reprocessing guidance and the equipment guidance from Ontario Midwives assume a validated sterilizer with monitoring. The savings from a pressure cooker are small compared with the cost of a load that cannot be released.

How long can contaminated instruments wait before cleaning?

The shorter the better, and the clinic's own procedure should state a number rather than leaving it to judgement. Instruments that cannot be cleaned immediately should be kept moist using a product intended for that purpose, because dried blood and tissue require more mechanical scrubbing, and mechanical scrubbing is where sharps injuries happen. Overnight is a failure of the transport step, not a normal interval to plan around.

Do dopplers and fetoscopes need to be on a log?

They are non-critical items and need low-level disinfection between clients, not a sterilization record. What they do need is a written cleaning schedule and a supply of the correct wipes, plus the discipline to remove ultrasound gel before disinfecting because gel protects organisms from the wipe. The gel bottle itself is a documented item in the provincial equipment guidance: label it with the date opened, do not refill it, and discard it after one month.

Who is accountable if a midwife uses a single-use device twice?

The midwife, professionally, and the practice owner, operationally. Reuse of a device labelled single-use is a departure from the manufacturer's instructions and from provincial guidance, and it does not become acceptable because the device looks intact. The practice-level fix is availability: the practice owner standard requires essential supplies to be available to midwives, and most single-use reuse in small practices traces back to a bag that was not restocked rather than to a clinical decision.

Sources

Midwifery Act, 1991, S.O. 1991, c. 31, sections 3 and 4, scope of practice and authorized acts.

College of Midwives of Ontario. Professional Standards for Midwives, approved 21 March 2018, implemented 1 June 2018, including practice owner standards.

College of Midwives of Ontario. Practice advisory: communicable illness and transmission prevention, standard precautions and professional standard references.

Ontario Midwives. Midwifery equipment disinfecting guidance, 2019, including the higher-level rule where manufacturer and intended use disagree.

Public Health Ontario. IPAC Checklist for Clinical Office Practice: Reprocessing of Medical Equipment/Devices, 2nd revision, July 2019, record keeping items 1.1 to 1.3.

CSA Group. CAN/CSA-Z314:23, Canadian medical device reprocessing; ISO 11607-1:2019 for packaging systems; EN 868-5:2018 for paper-film pouches and reels.