How Should a Dental Laboratory Reprocess Reusable Instruments in Canada?

How Should a Dental Laboratory Reprocess Reusable Instruments in Canada?

A dental laboratory is a reprocessing facility that nobody calls one. It has a sterilizer, an ultrasonic cleaner, a sink, a bench full of items that have been in someone’s mouth, and no patients of its own to remind it that any of this matters. The rule that governs the work is not written for the laboratory either: in Ontario it is written for the dentist, in a section of the college standard titled Dental Laboratory Asepsis, which describes what must happen to impressions, prostheses, articulators and in-office laboratory burs, and leaves the commercial laboratory to be the place where it happens.

That gap is where a laboratory’s compliance actually lives. Nobody audits the laboratory directly in most provinces; the clinics that use it are audited, and the audit asks the clinic to show that the two-way handover is controlled. So the practical answer to “how should a dental laboratory reprocess reusable instruments” starts with the division of labour, then moves to the bench. The consumables side is arithmetic rather than principle: a laboratory running a Class B cycle several times a day needs Class 4 self-sealing sterilization pouches, a Class 5 integrator for every package, and a 24-hour biological indicator for each load, ordered ahead of the shelf going empty rather than the morning it does.

Why is a dental laboratory different from a dental office?

Three structural differences, and every one of them changes what the reprocessing routine has to look like.

The first is that the laboratory works on devices that arrive already contaminated and leave as product. In an office, an instrument is reprocessed between two uses by the same team, in the same building, within a known interval. In a laboratory, an impression arrives in a bag with a case number on it, having been disinfected — or not — somewhere else, an unknown number of hours earlier. The laboratory cannot verify that, so it has to record what arrived and in what condition, which turns a clinical routine into a receiving process.

The second is the material mix. A laboratory bench holds stainless steel, carbide, silicone, acrylic, gypsum, alginate, vinyl polysiloxane, zirconia and printed resin, often within the same square metre. Steam is fine for one of those and ruinous for another. That is why the provincial reprocessing guidance puts so much weight on the manufacturer’s instructions for each device and material: in an office the mix is narrower and the habit can be general, while in a laboratory there is no single rule that covers the bench.

The third is what “the patient” means. A laboratory has no patient record, no chart, no chart number. It has case numbers and clinic accounts. Every recall, every trace and every record ties to a case, not a person, and that is the single fact that makes the laboratory’s paperwork look different from a clinic’s even where the underlying requirement is identical.

Schematic titled Four families, three destinations, showing four cards for dental laboratory item families: heat-tolerant items that entered the mouth (impression trays, face bow forks); impressions and prostheses (alginate, vinyl polysiloxane, occlusion rims, try-ins); bench tools (burs, polishing points, rag wheels, lathes); and items with no patient contact that are still handled (articulators, case pans, model trimmers). Each card states its destination. Footer reads clinieco.ca and MDEL 35334.
Figure 1. Four item families and the three destinations they resolve to. The destination is set by the contact category and the material, not by what the bench habitually does — which is why a bur and an impression do not end up in the same place.

How should impressions and prostheses be handled when they arrive?

The standard is explicit about timing: impressions, prostheses and appliances must be cleaned and disinfected as soon as possible after removal from the patient’s mouth, before drying of blood or other organic debris. Dried blood is harder to remove and, once a material has set, a disinfectant that would have been fine on a fresh surface may be the wrong choice for a distorted one. “Wet” impressions and appliances should travel in an impervious bag.

What the standard does not want is duplication. Its lab section says that effective communication and coordination between the office and the commercial laboratory will ensure three things: that appropriate cleaning and disinfection are performed in the office or the laboratory; that materials are not damaged or distorted because of over-exposure to disinfectants; and that disinfection procedures are not unnecessarily duplicated. Read that as a written agreement with names in it. If the office always disinfects, the laboratory verifies and records. If the office sends a wet, bagged impression for the laboratory to process, the laboratory owns the step and the record. What fails an audit is neither arrangement — it is an arrangement nobody wrote down, so both parties disinfect, or neither does.

Two practical notes that save material. First, the maker’s contact time for the specific impression material is a material limit rather than a suggestion: alginates distort with long soaking, vinyl polysiloxanes tolerate more. Second, a laboratory should not re-disinfect a prosthesis that has already been disinfected at dispatch unless something happened in transit, because the second exposure is exactly the over-exposure the standard warns about.

Which items in the laboratory have to be sterilized, and which only disinfected?

The division is by contact category and by material tolerance, not by how the item feels. Table 1 sets out the six families the standard and the reprocessing guidance actually address, and Table 2 works through ten specific items with the destination each one takes.

Table 1. Six item families a dental laboratory handles, the contact category each falls into, and the clause that decides what happens to it.
What the laboratory handles Contact category What the guidance requires Where it happens
Impression trays, face bow forks and any other heat-tolerant item that goes into the mouth Semicritical or critical, depending on the item Sterilized after each patient use. These are the items the standard names explicitly as heat-tolerant and mouth-contacting, so disinfection is not the answer for them. The clinic, before the item is reused on a patient. A laboratory that receives a contaminated tray has no way to know how long it sat.
Impressions, occlusion rims, bite registrations, try-in prostheses Semicritical — the surface touches mucosa and the dental office’s gloves Cleaned and disinfected as soon as possible after removal from the mouth, before blood or organic debris dries. The manufacturer’s instructions govern which disinfectant the material tolerates. Whichever party the two have agreed on — in practice the clinic, with the laboratory verifying on arrival and recording what it received.
Laboratory burs, polishing points, rag wheels, laboratory knives, lathes Contaminated by handling during adjustment rather than by direct patient contact Cleaned and sterilized, cleaned and disinfected, or discarded after use, according to the manufacturer’s recommendations. The standard lists these items by name because they are the ones people forget. The laboratory, after every adjustment session.
Articulators, case pans, model trimmers and other items that do not normally contact the patient Non-critical, but frequently contaminated Cleaned and disinfected according to the manufacturer’s instructions. Heat is usually not an option, and over-exposure damages the material. The laboratory.
Finished prostheses and appliances Semicritical when they leave the laboratory Delivered free of contamination, which the standard says may be achieved with an appropriate low-level disinfectant by either the commercial laboratory or the dental office. Either party — but the standard warns against doing it twice, because duplication damages materials and wastes staff time.
Anything the maker marks single-use Not reusable at all Single-use devices must not be reprocessed. The provincial reprocessing guidance states that rule without a laboratory exemption. Nowhere. The item is discarded.

The sentence worth memorising is the one about the in-office laboratory: burs, polishing points, rag wheels, laboratory knives and dental lathes frequently become contaminated during adjustments to prostheses and appliances, and must be cleaned and sterilized, cleaned and disinfected, or discarded after use, according to the manufacturer’s recommendations. Three destinations in one sentence is unusual in a standard, and it is there because the correct answer genuinely varies by item: a steel bur can be sterilized, a silicone point usually cannot, and a felt wheel is often costs less to replace than to document.

Table 2. Ten items, three destinations. The right-hand column is the test to apply when a new item appears on the bench.
Item Sterilize Clean and disinfect Discard The line that decides
Stainless steel lab bur used on a prosthesis in the mouth Yes Not sufficient When blunt or corroded It entered the mouth. Heat-tolerant metal, so there is no reason to stop at disinfection.
Carbide bur used only on a stone model Yes, if the maker allows Yes, if the holder is heat-sensitive When dull The device is contaminated by handling, not by the patient. Follow the maker, not the habit.
Rag wheel, felt point, bristle brush Usually not — heat destroys the wheel Yes, or discard Often the honest answer Consumable polishing media are cheap; the reprocessing record costs more than the wheel.
Silicone polishing point Per maker only Yes When worn Silicone will not survive a steam cycle designed for metal.
Articulator No Yes When broken No patient contact, but it sits on the bench next to everything that did.
Impression in alginate No Yes, with a short contact time After pouring Alginate distorts if it soaks. The maker’s contact time is a material limit, not a suggestion.
Impression in vinyl polysiloxane No Yes After pouring Dimensionally stable, so a longer contact time is usually tolerated.
Tray the clinic sent in, heat-tolerant Yes No When damaged The standard names heat-tolerant mouth-contacting items as sterilize-after-each-use.
CAD/CAM milling bur No — not a patient item Yes, or brush and change When dull It never enters a mouth; the hazard is resin and dust, not blood.
Prosthesis being returned to the clinic Only if the maker allows Yes — the usual route Never — it is the product It must arrive free of contamination; low-level disinfection is what the standard contemplates.

Two rules sit underneath both tables. The first is that single-use means single-use: the provincial reprocessing guidance states plainly that critical and semi-critical devices labelled single-use must not be reprocessed, and that rule has no laboratory exemption. The second is that heat-tolerant items which enter the mouth belong in the sterilizer after every patient, which in a laboratory context usually means the clinic’s trays and face bow forks rather than the laboratory’s own tools.

What are the steps in a laboratory reprocessing cycle?

The provincial guidance describes reprocessing as a chain rather than an act, and it lists the components in order: collection at point of use, containment and transport, disassembly where required, inspection, cleaning, disinfection or sterilization at the level the item’s risk class requires, rinsing after disinfection, drying or aeration, reassembly and functional testing, clean and dry storage, and clean transportation. For a laboratory that chain starts at the receiving bench, not at the sink.

Schematic band titled What the laboratory actually controls, showing four steps the laboratory owns: receive and record, clean, package and release, each with the evidence it leaves, from the case number on arrival to the internal indicator and the printed cycle parameters. A closing line notes that the handover at each end is a record, not a conversation, and that a case number on the load sheet makes a recall possible. Footer reads clinieco.ca and MDEL 35334.
Figure 2. The four steps the laboratory actually controls: receive and record, clean, package, release. Each one leaves a line, and the handover at each end is a record rather than a conversation.

Three steps in that chain are where laboratories most often improvise. Cleaning is the first: manual scrubbing, an ultrasonic bath and a washer-disinfector are all acceptable, but the ultrasonic bath has to be tested and the solution changed on the maker’s schedule, not on the day it looks cloudy. A laboratory cleaning burs and handpieces in a bath that has not been tested is cleaning in a liquid of unknown performance. Inspection is the second: the guidance requires visual inspection after cleaning and before terminal sterilization, with repeat cleaning for anything that is not clean, and it says not to reassemble before sterilization unless the maker requires it. Drying is the third: items are dried before packaging, stainless steel is dried immediately after rinsing to prevent spotting, and lumens — where a laboratory has any — are dried with filtered air rather than left wet.

Packaging comes next. The guidance requires an internal chemical indicator inside each package, container or bundle, positioned in the area least accessible to steam rather than automatically in the centre, and an externally visible Class I indicator on every package, examined immediately after the cycle to confirm the item was exposed to the process. A laboratory that tapes a pouch shut and reads only the tape has done one of the two required checks.

How should a laboratory monitor a sterilizer it runs all day?

By three separate means, and the guidance is specific about all three. Chemical monitoring: the internal indicator per package and the external Class I indicator per package, plus an air-removal test with a Class II indicator every day the sterilizer is used if it is a dynamic air removal type. Biological monitoring: a biological indicator each day the sterilizer is used and for each type of cycle used that day, citing CSA Z314.3, plus a BI in every load containing implantable devices. Physical monitoring: the cycle printout, signed to verify that the time, temperature and pressure required were actually achieved.

Release is the point most laboratories get wrong without noticing. The guidance says items in a processed load should not be released until the biological indicator result is available, and then gives the escape route: where quarantine pending BI results is not possible, evaluation of a Class 5 or 6 chemical indicator together with the specific cycle’s physical parameters may justify releasing routine loads. Implantable devices must be quarantined until the BI result is in. A laboratory running a same-day turnaround is living on that escape route every day, which is exactly why the integrator has to be in the package and the printout has to be legible. A Class 5 integrator plus indicator tape covers the routine; 24-hour biological indicator covers the day the cycle type changes, because the BI requirement follows the cycle, not the calendar.

Two monitoring questions come up in every laboratory that runs more than one sterilizer. The first is whether a BI from one unit can cover another: it cannot, because the requirement is per sterilizer, per day, per cycle type. The second is what a new or repaired unit needs: the guidance requires at least three consecutive cycles with the appropriate biological and chemical challenges in an empty sterilizer plus at least one cycle challenged with a full test load before routine service, with three consecutive air-removal tests for a dynamic air removal unit, and a positive BI means the unit is not approved for use.

What does the laboratory owe the clinic, and what does the clinic owe the laboratory?

The standard’s own answer is a written understanding between the two parties, and it lists the three things that understanding has to secure: cleaning and disinfection happen on one side or the other, materials are not damaged by over-exposure, and the work is not duplicated. In practice a laboratory’s side of that agreement should be four sentences long. Which items the laboratory disinfects on arrival and which it verifies as already done. Which items the laboratory sterilizes and returns in a pouch with an indicator. What the laboratory records and for how long. Who is called when something goes wrong with a case that has already left the building.

The clinic’s side is equally short: bag wet impressions, disinfect before dispatch or say plainly that it did not, and send trays and face bow forks that have been sterilized rather than rinsed. A laboratory that sends a one-page version of this to new accounts, with a place for the clinic to initial it, has a document it can produce in an audit that asks the clinic to demonstrate control of the handover. The CliniEco guide to how cases are packed and tracked between Canadian practices and laboratories covers the physical side of the same handover.

How should reprocessed items be stored and moved inside the laboratory?

The guidance is more prescriptive here than most laboratories expect. A sterile store is a separate, enclosed, access-controlled area; ambient temperature is 18 to 23 °C with relative humidity of 30 to 60 per cent; containers must be moisture-resistant and cleanable, and corrugated cardboard is explicitly excluded; shelving is closed, clean, dry and dust-free, with items off the floor and away from windowsills and sinks, solid shelves rather than open wire, and bottom shelves at least 25 cm off the floor with 46 cm of clearance below the ceiling. Sterile stock is not shared with unrelated consumables.

For a laboratory the realistic version is a marked cupboard or a closed rack in the dry area, not a shelf over the sink, with processed items separated from unprocessed items in a way that a locum can read at a glance. The guidance makes that separation a requirement rather than good practice: once devices have been reprocessed there must be a process to ensure they can be differentiated from devices that have not, using external chemical indicators such as tape that changes colour, or labelling and colour coding for items that received high-level disinfection. In a laboratory the colour of the tape is doing the work of the whole receiving process, so a roll of tape that has been sitting in a warm drawer since last winter is a compliance problem rather than a stationery problem.

What records does a laboratory reprocessing routine produce?

Four families, and the laboratory already has most of the data even if it is not written down. The first is the receiving record for each case: what arrived, from which clinic, whether it was bagged wet, and what disinfection the laboratory performed or verified. The second is the load record: which sterilizer, which cycle, what was inside, who ran it, and the chemical and biological results attached to it. The third is the equipment record: ultrasonic test results, solution changes, sterilizer preventive maintenance. The fourth is the dispatch record, which is what makes a recall possible at all. The questions an auditor actually asks about those records are collected in what an infection control auditor checks about dental instrument reprocessing.

Where does a laboratory most often fall short?

Not usually at the sterilizer. The failures that survive a walk-through are quieter, and four of them repeat. A receiving bench with no record, so the laboratory cannot say what condition a case arrived in. An ultrasonic bath that has never been tested and a solution that is changed when it looks dirty. A pile of processed burs and wheels on the dry bench beside a pile of unprocessed ones, distinguished by nothing. And a pack of pouches stored in a warm room above the sink, where the indicator tape goes brown before it is used, so the one signal the laboratory relies on has already fired.

Each of those is a design problem, not a discipline problem. A receiving sheet on a clipboard, a test log taped to the ultrasonic lid, two trays that are physically different colours, and a closed cupboard in the dry area all cost less than the training session that tries to fix them afterwards. The laboratory’s other reprocessing-adjacent duties are covered separately: how laboratory burs and handpieces are maintained, impression disinfection on arrival, ultrasonic cross-contamination in a dental laboratory, and the waste streams that gypsum, monomer and alloy create, in biomedical waste segregation for a dental laboratory. The chain itself, from cleaning to storage, is set out once at the level of principle in the instrument reprocessing workflow guide.

One boundary belongs at the end of a page like this. Reprocessing guidance tells a laboratory how to make a device safe to handle; it does not decide whether the laboratory may manufacture or sell the device it is making. If a laboratory also places custom-made prostheses on the Canadian market, device licensing and establishment licensing are Health Canada questions with their own answers, and they are outside the scope of this page.

Primary sources. The item rules, handover requirements, monitoring requirements, storage geometry and reprocessing chain above are quoted from the college standard of practice for Ontario dentists and the provincial reprocessing best practices that sit behind it; a specific item is classified by its contact category and its maker’s instructions, not by this page.

Where does a dental laboratory take this next?

Three of the tools here are usable rather than readable. The printable compliance log carries the fields a receiving check, a lot trace and a reprocessing record actually need, so the paperwork can start without inventing a format. The clinic supply planner turns a month of case volume into pouch, integrator and indicator quantities, so the bench is never restocked mid-cycle. The sterilization monitoring calculator works out how many indicators a year of cycles consumes, and the provincial self-check lists what an inspector asks for in each province. Multi-site or purchasing-group enquiries go through the wholesale account route. Reselling? become a distributor.

Questions that go past a template — a record format a particular inspector asked for, or a retention period your college is challenging — belong with the compliance consultation desk rather than a generic contact form.

If you would rather test the routine than read about it, the 24-hour biological indicator sample request is the simplest way to confirm that the incubator, the records and the staff habit all work together. The sterilization compliance hub collects the rest of the guides for this setting.

Related Reading

Frequently Asked Questions

Does a dental laboratory have to follow the dental college’s infection control standard?

The standard binds the dentist, not the laboratory, and that is precisely why it reaches the laboratory. It requires the dental office and the commercial laboratory to coordinate so that cleaning and disinfection happen on one side or the other, materials are not damaged by over-exposure, and the work is not duplicated. In practice a laboratory complies with the standard because the clinics it serves have to demonstrate that the handover is controlled, and the practical instrument is a written understanding between the two parties.

Do impressions need to be sterilized before a laboratory handles them?

No. Impressions and prostheses are cleaned and disinfected rather than sterilized, and the standard is explicit that this happens as soon as possible after removal from the mouth, before blood or organic debris dries. What is sterilized is heat-tolerant equipment that entered the mouth, such as impression trays and face bow forks, after each patient use. Impression materials would be damaged by a steam cycle, which is why the standard instead warns against over-exposure to disinfectants.

What is the difference between the burs a laboratory sterilizes and the ones it discards?

Material tolerance decides it. The standard says in-office laboratory burs, polishing points, rag wheels, laboratory knives and lathes must be cleaned and sterilized, cleaned and disinfected, or discarded after use, according to the manufacturer’s recommendations. A stainless steel bur tolerates steam; a silicone point or felt wheel usually does not. Where an item is cheap and its maker gives no validated reprocessing instruction, replacement is the simpler and defensible answer.

How often does a laboratory need to run a biological indicator?

Each day the sterilizer is used and for each type of cycle used that day, plus a biological indicator in every load that contains implantable devices. The requirement follows the sterilizer and the cycle, not the calendar: a second sterilizer needs its own test, and a day on which two cycle types run is a day with two tests. A positive result means the unit is not approved for use until the cause is corrected and requalification is complete.

Can a laboratory release a load before the biological indicator result is back?

The provincial guidance says items in a processed load should not be released until the BI result is available, and then provides a defined alternative: where quarantine pending results is not possible, evaluation of a Class 5 or 6 chemical indicator together with the cycle’s physical parameters may justify releasing routine loads. Implantable devices must be quarantined until the result is in. A same-day turnaround therefore depends on the integrator and the printed parameters being present and legible for every load.

What has to be inside a sterilization pouch a laboratory sends back to a clinic?

An internal chemical indicator inside each package, container or bundle, positioned in the area judged least accessible to steam rather than automatically in the centre, plus an externally visible Class I indicator on every package that is examined immediately after the cycle. The class of internal indicator depends on the parameters being measured, which is why a Class 5 integrator is the usual choice where the load must be released on chemical evidence.

How should a laboratory store sterile items waiting to go back to a clinic?

In a separate enclosed area with access control, at 18 to 23 °C and 30 to 60 per cent relative humidity, in closed, clean, dry and dust-free shelving, not on the floor, not near a windowsill or a sink, with corrugated cardboard excluded and unrelated consumables kept out. Processed items also have to be distinguishable from unprocessed items, which is what external indicator tape or a colour code is for.

Is a laboratory responsible if a clinic sent a contaminated impression?

The laboratory is responsible for what it does with the case once it arrives, including recording the condition it received it in and disinfecting it if that task is on the laboratory’s side of the agreement. Responsibility for the clinical step stays with the clinic. That is why a receiving record with the date, the case number, the condition on arrival and the disinfection performed matters: without it, neither party can show where a step was missed, and in a recall neither party can identify the case.

CliniEco Medical holds Health Canada MDEL #35334. The item rules and handover duties described here follow the college standard and the provincial reprocessing guidance that set them; an individual item is classified by its contact category and its maker’s instructions, not by this page. This page is general information about reprocessing equipment and consumables, not clinical or legal advice. Follow the instrument and sterilizer manufacturers' validated instructions; where this page and a maker's instructions differ, the maker's instructions govern. Health Canada MDEL #35334 covers our establishment licence; it is not a statement about any individual product.

Already working with clinics? Become a regional representative.