Centrifugal Chemistry Analyzers in Canada vs the US: Which Class Applies?

Graduated plastic specimen container with a screw cap on a dark background

Quick facts

  • A centrifugal chemistry analyzer for clinical use is an automatic device that centrifugally mixes a sample and a reagent and measures the concentrations of sample constituents spectrophotometrically, in conjunction with certain materials.
  • In the United States it is a named entry. 21 CFR 862.2140, Centrifugal chemistry analyzer for clinical use, is Class I (general controls), with product code JJG read from the openFDA classification database.
  • The US entry states that the device is exempt from premarket notification procedures in subpart E of part 807, subject to the limitations of the exemption section in part 862.
  • Canada keeps no named entry and classes an in vitro diagnostic device under Part 2 of Schedule 1. Part 2 Rule 8 places an IVDD at Class I where Rules 1 to 7 do not apply.
  • Part 2 Rule 7 is the clause that changes the picture: where an IVDD, including its analysers, reagents and software, is intended to be used with another IVDD, the class of both is that of the IVDD in the higher class.
  • Entries, classes and product codes below were checked against the eCFR, the Regulations and the openFDA classification database on 2026-10-16.

A centrifugal chemistry analyzer is the bench instrument that spins a sample and a reagent together and reads the result photometrically. It is unusual among laboratory devices because the United States places the analyser itself at Class I while the assays run on it sit in higher classes. Canada reaches the same starting point by a different road, and then adds one clause that ties the analyser to whatever it is used with. This article sets out both routes and the record a bench keeps, with the 15 mL sterile centrifuge tubes that are part of the same workflow.

What is a centrifugal chemistry analyzer?

It is an automatic analyser that uses centrifugal force to mix a sample with a reagent and then measures the concentration of a constituent spectrophotometrically. The US identification text limits the intended use carefully: the device is intended for use in conjunction with certain materials to measure a variety of analytes. In other words, the analyser on its own does not measure anything; it measures what someone else's reagent makes measurable.

That sentence is the whole classification story for this device. An instrument that is general purpose and depends on the assay supplied for it is treated as an instrument, and an instrument that is sold as part of a specific assay system is treated as part of that system. The two readings lead to different classes in Canada, and the same instrument can be sold both ways.

Graduated plastic specimen container with a screw cap on a dark background

What class is it in the United States?

It is a named entry at 21 CFR 862.2140, titled Centrifugal chemistry analyzer for clinical use. The identification reads: "A centrifugal chemistry analyzer for clinical use is an automatic device intended to centrifugally mix a sample and a reagent and spectrophotometrically measure concentrations of the sample constituents. This device is intended for use in conjunction with certain materials to measure a variety of analytes." The regulation sets Class I (general controls).

The openFDA device classification API returns a single product code against the section: JJG, Class I. The entry also states that the device is exempt from the premarket notification procedures in subpart E of part 807, subject to the limitations in the exemption section of part 862 - which is why a buyer who expects a clearance number for the instrument will not find one.

Item Detail
US regulation 21 CFR 862.2140 - Centrifugal chemistry analyzer for clinical use
US class Class I (general controls)
Product code JJG
Stated intended use Mix sample and reagent centrifugally; measure sample constituents spectrophotometrically, with certain materials, for a variety of analytes
Premarket route Exempt from premarket notification, subject to the limitations of the exemption
Read on eCFR identification text and openFDA classification, both read 2026-10-16

How does Canada classify it?

There is no named entry, and the applicable text is Part 2 of Schedule 1 rather than Part 1. The starting point is Part 2 Rule 8: "If rules 1 to 7 do not apply, all other IVDDs are classified as Class I." A general-purpose analyser with no transmissible-agent claim and no diagnostic claim of its own reaches that rule and sits at Class I, the same class the US entry gives it.

The clause that changes the picture is Part 2 Rule 7: "In cases where an IVDD, including its analyzers, reagents and software, is intended to be used with another IVDD, the class of both IVDDs will be that of the IVDD in the class representing the higher risk." The word analysers is in the rule. Where the instrument and an assay are supplied as one system, the analyser does not keep its own class: it takes the class of the assay, which can be Class II or Class III.

There is also a Canadian rule for instruments used close to the patient. Part 2 Rule 6 states that a near patient IVDD is classified as Class III, and Part 2 Rule 9 then sets out a table that overrides Rules 1 to 8 for the items listed in it. A buyer who reads only Rule 6 will over-class the instrument; a buyer who reads only Rule 8 will under-class a system that is sold with its own assay.

Foldable examination table covered with a light blue disposable protective sheet

Why does the analyser's class depend on the assay?

Because that is what Part 2 Rule 7 says, and because the US entry says something adjacent in the opposite direction. The US identification text limits the analyser to use "in conjunction with certain materials", which is what keeps it at Class I: the instrument is not the thing that produces a diagnostic result. Canada states the same relationship as a classification rule, and the consequence is that the analyser inherits a higher class rather than keeping its own.

The practical difference for a laboratory is what the supplier is asked to produce. A general-purpose analyser sold with no assay and no diagnostic claim is a Class I instrument in both countries. The same cabinet sold as part of an assay system answers a different question in Canada, and the licence, the labelling and the change-control obligations travel with the system rather than with the box.

Item Canada United States
Named entry None; classified under Part 2 of Schedule 1 862.2140 - Centrifugal chemistry analyzer for clinical use
Class, general-purpose analyser Class I (Part 2 Rule 8) Class I
Class, analyser sold with an assay Class of the higher-risk IVDD in the system (Part 2 Rule 7) Not provided for; the entry keeps one class
Near-patient placement Part 2 Rule 6 sets Class III, subject to the Rule 9 table Not addressed in the entry
Governing text Schedule 1, Part 2, Rules 6, 7, 8 and 9 21 CFR 862.2140
Premarket route Device licence where the class requires it Exempt from premarket notification, subject to the exemption limits

What should a laboratory check before ordering?

Four checks. First, confirm whether the analyser is quoted alone or as one system with an assay, a reagent pack and software, because Part 2 Rule 7 turns on that. Second, confirm the placement - laboratory bench or near patient - because Rule 6 and the Rule 9 table answer differently from Rule 8.

Third, confirm the tube and sample volumes the instrument expects, since the analyser geometry and the consumable have to agree before a validation run means anything. Fourth, confirm the traceability chain the maker states for the calibrator, because a concentration reported from a photometric reading is only comparable across instruments when the calibration is traceable to a stated reference.

Tube stock sits in the same workflow as the analyser. Laboratories that buy across several benches order through the wholesale programme. Clinics that stock for a whole practice can order through the wholesale programme for case pricing and account setup; resellers can become a distributor. The 15 mL sterile centrifuge tubes sits in the same catalogue. Consumables for this line are listed under diagnostic laboratories.

How does the class affect a validation run?

The class sets the licence path and the conditions attached to it. It does not set the validation. What a bench records is set by the maker's instructions and the laboratory's own quality system: the analyser identifier, the reagent lot, the calibrator lot, the control result, the specimen and the operator.

Where several benches in one laboratory run the same assay, the record is also what makes them comparable, because a result is only transferable when the instrument, the calibration and the control are all named. The class answers who reviews the system before it is sold; the record answers whether the run can be repeated.

Sources

  1. openFDA device classification API
  2. Medical Devices Regulations, SOR/98-282, full text and Schedule 1
  3. US FDA, medical device databases
  4. US FDA, overview of device regulation
  5. Health Canada, medical devices
  6. Health Canada, establishment licensing
  7. ISO 15189:2022, medical laboratories, requirements for quality and competence
  8. ISO 17511:2020, metrological traceability of values assigned to calibrators
  9. ISO 13485:2016, medical devices quality management
  10. ISO 14971:2019, application of risk management to medical devices
  11. ISO 11137-1:2006, radiation sterilization of health care products
  12. ISO 20776-1:2019, susceptibility testing of infectious agents
  13. Clinical and Laboratory Standards Institute
  14. World Health Organization
  15. US CDC, laboratory quality
  16. Public Health Ontario, infection prevention and control
  17. Standards Council of Canada

Printable record sheet. A validation sheet for an analyser records the reagent lot, the calibrator, the control result and the operator on one printable page. Download the analyser validation sheet — and print one per run - no account and no sign-up.

Log generator. Type the laboratory and the analyser once and the log generator at the compliance log centre — lays out a dated, facility-labelled sheet for the quality manual.

Ask a compliance specialist. A classification question usually turns on how the device is used, not only on its name. Ask a compliance specialist — you get a written answer specific to your equipment and province, with the regulation or standard it is based on cited.

Related reading

CliniEco Medical supplies eye care, laboratory and dental consumables of the kind described in this article. MDEL #35334. This article is written for optometry, laboratory and dental procurement professionals and is not legal advice; the class of a device and the records a practice keeps belong to the manufacturer and to the requirements that apply to it.

Frequently Asked Questions

What class is a centrifugal chemistry analyzer in the United States?

Class I, under the named entry at 21 CFR 862.2140, product code JJG. The entry also makes the device exempt from premarket notification, subject to the limitations of the exemption.

What class is it in Canada?

A general-purpose analyser reaches Class I under Part 2 Rule 8. Where the analyser is intended to be used with another IVDD, Part 2 Rule 7 gives both the class of the higher-risk device.

Why is the analyser not classified under the Part 1 rules in Canada?

Part 1 governs medical devices other than in vitro diagnostic devices. An analyser used in vitro is an IVDD and is classified under Part 2 of Schedule 1.

Does the Canadian rule reach the reagents and software?

Yes. Part 2 Rule 7 expressly names analyzers, reagents and software in the clause that gives both devices the class of the higher-risk one.

What about a near-patient version?

Part 2 Rule 6 classes a near patient IVDD as Class III, and Part 2 Rule 9 then applies the table to the items listed in it. Both have to be read.

Does the class change the validation record?

No. The class sets the licence path. The record - analyser, reagent lot, calibrator, control and operator - is set by the maker's instructions and the laboratory quality system.