Quick facts
- A cell-concentrate centrifuge separates a blood draw into a plasma or cell layer that is returned to the patient, as in platelet-rich plasma preparation.
- In the United States it files under 21 CFR 862.2050, General purpose laboratory equipment labeled or promoted for a specific medical use, at Class I, product code QBV.
- Canada does not leave it at Class I. Because the device modifies blood by centrifugation for reintroduction, Rule 6(3) of Schedule 1 sets it at Class II.
- This is a genuine US-versus-Canada difference: the same bench centrifuge is Class I south of the border and Class II in Canada.
- The separating kit - the tube, the double syringe and the transfer set - is single-use and is what a clinic reorders by the procedure.
- Entries, classes and product codes below were checked against the eCFR, the Regulations and the openFDA classification database on 2026-10-13.
A cell-concentrate centrifuge spins a small blood draw so that the plasma or cell layer can be separated and prepared for return to the patient, the step at the heart of platelet-rich plasma and similar treatments. Clinics buy the centrifuge as a bench instrument and consume a separating kit with each procedure. In the United States the instrument files as general purpose laboratory equipment at Class I; in Canada the same machine is pulled up to Class II because it modifies blood for reintroduction. This article sets out both routes and what the difference means for a clinic's order, alongside the graduated specimen container that enters the same chain of custody.
What is a cell-concentrate centrifuge?
It is a benchtop centrifuge sold for a specific clinical preparation rather than as general lab equipment. The operator draws a small volume of the patient's blood, spins it under a set programme so the red cells, the buffy coat and the plasma separate into defined layers, and then draws off the layer that the treatment calls for. The plasma or cell-concentrate is prepared to be returned to the patient in the same visit.
Two features distinguish it from the general-purpose centrifuge on the next bench. The first is the programme: the speed, time and acceleration are preset for one preparation, not set freely by the operator. The second is the kit: a matched tube and transfer set designed to keep the separated fraction sterile and identified. Neither feature changes the hardware; both change how the device is described.
What class is it in the United States?
It is Class I. 21 CFR 862.2050 is titled General purpose laboratory equipment labeled or promoted for a specific medical use, and the openFDA device classification API lists product code QBV, Centrifuge For Preparation Of Cell Concentrate And/Or Plasma Concentrate, under that regulation number. A Class I device of this kind reaches the market under general controls.
The regulation number is the anchor for a purchaser. A centrifuge sold for a named clinical preparation is described by 862.2050 and its product code; a centrifuge sold as general lab equipment is not promoted for a specific medical use and falls outside that description. The intended use on the labelling is what decides which of the two a buyer is holding.
| Item | Detail |
|---|---|
| US regulation | 21 CFR 862.2050 - General purpose laboratory equipment labeled or promoted for a specific medical use |
| US class | Class I (general controls) |
| Product code | QBV (centrifuge for preparation of cell concentrate and/or plasma concentrate) |
| Route to market | Registration and listing; general controls |
| Applied part | Blood is processed in a closed, single-use kit |
How does Canada classify a cell-concentrate centrifuge?
Canada reads the device differently. It is a non-invasive device intended for modifying the biological or chemical composition of blood for the purpose of introduction into the body, which puts it under Rule 6 of Schedule 1. Subrule (1) sets that class of device at Class III; subrule (2) lifts a potentially hazardous case to Class IV; and subrule (3) provides that a device "that accomplishes the modification by centrifugation, gravity filtration or the exchange of gas or heat is classified as Class II". A separating centrifuge does its work by centrifugation, so it lands at Class II.
That is the whole difference between the two countries. The American entry keys off the device being general laboratory equipment promoted for a specific medical use, and leaves it at Class I. The Canadian rules key off what the device does to the blood — it changes the blood's composition before reintroduction — and then drop the class back to Class II because the change is made by centrifugation rather than by a chemical or thermal process.
| Item | Canada | United States |
|---|---|---|
| Named entry | None; classified by rule | 862.2050, Class I |
| Class | Class II (Rule 6(3)) | Class I |
| Governing text | Schedule 1, Rule 6(3) | 21 CFR 862.2050 |
| Basis for the class | Modifies blood composition by centrifugation for reintroduction | General purpose laboratory equipment promoted for a specific medical use |
| Route to market | Device licence (Class II) | Registration and listing |
Why do Canada and the United States give different classes?
The two systems ask different questions. The United States asks what kind of article is being sold and how it is promoted, so a centrifuge that is general laboratory equipment keeps the Class I that general laboratory equipment carries. Canada asks what the device does to the body or to the blood, so a centrifuge that prepares a fraction for reintroduction is judged as a blood-modifying device first and a centrifuge second.
The step-down inside Rule 6 is the other half of the answer. Without subrule (3) the device would sit at Class III, because modifying blood for introduction into the body is treated as a high-risk activity. Subrule (3) recognises that centrifugation is a physical separation with a lower added risk than a chemical or thermal modification, and brings the class down to Class II. That is why the Canadian answer is Class II and not Class III.
There is a second consequence that a clinic running both a separating centrifuge and a routine lab centrifuge meets directly. The two machines sit in different classes in Canada even though they look alike and share a bench: one is a blood-modifying device and the other is not. A procurement list that treats them as interchangeable will mis-state the class of the separating unit, which is exactly the kind of error an inspection is designed to find.
What should a clinic check before ordering one?
Three checks. First, confirm the intended use on the labelling: a unit promoted for a named clinical preparation is described differently from a general lab centrifuge. Second, confirm that the separating kit — the tube, the transfer set and the double syringe — is available on the same order, because that kit is the consumable the procedure consumes. Third, confirm the centrifuge programme can be locked, since a preparation that returns a fraction to the patient needs a repeatable, documented spin.
Regenerative-medicine and sports clinics also reorder the sample-handling consumables that feed the same bench. Clinics that stock for a whole practice can order through the wholesale programme and request a bulk quote for a standing or annual order; the graduated specimen container sits in the same catalogue. Reselling? become a distributor.
How does the class affect a separation clinic's routine?
Class II in Canada means the separating centrifuge carries a device licence, so the clinic's supplier documents the device under the higher of the two classes the unit is sold at in North America. For the clinic the routine is a per-procedure record: the programme used, the separation time and the kit lot. The centrifuge's own schedule covers the rotor, the seals and the calibration of the speed and timer.
Because the separated fraction is returned to the patient, the record is the part that matters most in a review. A clinic that can show which programme ran, with which kit lot, on which date, answers the question a reviewer would ask; the class is the reason the record is expected at all.
Sources
- openFDA device classification API
- Medical Devices Regulations, SOR/98-282, full text and Schedule 1
- US FDA, medical device databases
- US FDA, overview of device regulation
- Health Canada, medical devices
- Health Canada, establishment licensing
- Clinical and Laboratory Standards Institute
- Public Health Ontario, infection prevention and control
- ISO 15189:2022, medical laboratories - requirements for quality and competence
- ISO 13485:2016, medical devices quality management
- ISO 14971:2019, application of risk management to medical devices
- US CDC, laboratory quality
- World Health Organization
- Standards Council of Canada
- CSA Group
- CCOHS, OSH answers
- ISO 15197:2013, in vitro diagnostic test systems
- ISO 11135:2014, ethylene oxide sterilization of medical devices
- Health Canada, blood and plasma establishments
- Public Health Ontario, laboratory specimen collection
Printable record sheet. A separation service documents each spin, and the printed sheet is the copy the clinic keeps with the patient record. Open the sterilization log sheet library — the downloadable pack holds a steam-cycle log, a reprocessing-room checklist and an equipment-maintenance sheet. No account and no sign-up; print as many copies as your log needs.
Log generator. The generator builds the centrifuge log columns so each programme run is recorded in one pass. Run the sterilization log generator to build a dated, facility-labelled cycle sheet you can print or file straight into your quality manual.
Ask a compliance specialist. Whether a centrifuge is Class I or Class II turns on the intended use; a written answer keyed to your unit and province settles it. Ask a compliance specialist — you get a written answer specific to your equipment and province, with the regulation or standard it is based on cited.
Related reading
- Laboratory centrifuges in Canada vs the US
- Cytocentrifuges in Canada vs the US
- Blood collection tube systems in Canada vs the US
CliniEco Medical supplies eye care, laboratory and dental consumables of the kind described in this article. MDEL #35334. This article is written for optometry, laboratory and dental procurement professionals and is not legal advice; the class of a device and the records a practice keeps belong to the manufacturer and to the requirements that apply to it.
Frequently Asked Questions
What class is a cell-concentrate centrifuge in the United States?
Class I, under 21 CFR 862.2050, general purpose laboratory equipment labeled or promoted for a specific medical use, product code QBV.
What class is it in Canada?
Class II, through Rule 6(3) of Schedule 1 of the Medical Devices Regulations, as a non-invasive device that modifies blood by centrifugation for reintroduction.
Why is the class different between the two countries?
The United States keys off how the article is promoted, and leaves general laboratory equipment at Class I; Canada keys off what the device does to the blood, and classifies a centrifugal blood-modifying device at Class II.
Would a general-purpose lab centrifuge be Class II in Canada?
Rule 6 applies to a device that modifies blood for introduction into the body; a centrifuge used only to spin routine specimens is not doing that work, so the rule does not reach it.
What else is a cell-concentrate centrifuge called?
It is also written as platelet-rich-plasma centrifuge, PRP centrifuge, cell separator centrifuge, and plasma-concentrate centrifuge.
Is the separating kit part of the device?
It is a single-use accessory of the same device family and is what the clinic reorders per procedure.