A Bowie-Dick test belongs at the start of every working day on a pre-vacuum sterilizer, on an empty chamber, before the first load. It checks one specific thing: whether the vacuum stage removed air from the chamber. A pass means steam can penetrate a dense pack; a fail means the unit should not be used for routine loads until the fault is fixed and the test passes again.
What Does the Bowie-Dick Test Check?
The test is an air-detection method, not a sterility test. A standardized sheet or pack changes colour evenly only where steam reached it uniformly. Patches of unchanged colour mean air was trapped, which typically points to a door gasket leak, a vacuum pump that cannot reach target pressure, or a non-condensable gas in the steam supply.
Because it isolates the vacuum function, the Bowie-Dick test is the fastest way to catch an air-removal fault before it affects a load. It is not a substitute for the biological indicator, which verifies the kill, and clinics should run both on their own schedules and log them separately.
How Often Should You Run a Bowie-Dick Test?
Daily, before the first load, on any day the sterilizer is used. The test is run on a cold, empty chamber, so it is naturally the first task of the day and not an end-of-day chore. Some clinics also repeat it after a major repair, a long shutdown or a move, since any of those can change vacuum performance.
Logging the result is part of the test. An air removal result that is not written down is, for audit purposes, a test that was not done. A short line with the date, the operator and a pass or fail keeps the day's assurance in one place alongside the biological indicator record.
What Should You Record in the Bowie-Dick Log?
Record the date, the sterilizer identifier, the cycle type used to run the test, the result, and the initials of the person who ran it. If the test pack is a commercial product, noting its lot number helps if a batch of test packs is ever questioned. Keep the log with the daily biological indicator results, not in a separate binder.
The most common gap we see is a log that records the result but not the machine. In a clinic with two sterilizers, a result with no unit identifier cannot show that each machine was tested, which is exactly the point of a daily air removal programme.
What Do You Do When a Bowie-Dick Test Fails?
Re-run the test on a cool chamber to rule out a procedural cause, such as a hot chamber or a badly prepared pack. If it fails again, treat the unit as faulty, stop routine use, and check the gasket, drain, vacuum pump and steam supply before escalating to service. Record the failed result immediately rather than at the end of the day.
After the repair, run the test again and record the pass before returning the unit to service. Any load processed after a failed test should be treated as suspect, which is why the timing of the failure in the log matters and why a result written hours later is far less useful to a reviewer.
| Question | Answer |
|---|---|
| What it checks | Air removal and steam penetration |
| When to run | Start of each working day, cold empty chamber |
| Record | Date, unit, cycle, result, operator initials |
| On failure | Retest cool, then service, then retest and record a pass |
Related reading
Sterilization compliance hub; daily air removal and Bowie-Dick requirements; the Ontario daily air removal requirement.
Download the printable sterilization log sheet and use the autoclave log generator. Ask a Sterilization Compliance Specialist at the compliance desk. Ordering for a clinic, lab or care home? Wholesale and multi-site ordering and the B2B wholesale collection cover case pricing and account setup. Clinics can also 24-hour biological indicator 25-pack. Reselling? become a distributor.
Frequently Asked Questions
What does a Bowie-Dick test actually check?
It checks whether the vacuum stage removed air from the chamber, so it detects air-removal faults. It does not test spore kill, which is the job of the biological indicator.
How often should a Bowie-Dick test be run?
Daily, at the start of each working day the pre-vacuum sterilizer is used, on a cold empty chamber before the first load, and again after a major repair or long shutdown.
What should be logged for a Bowie-Dick test?
The date, sterilizer identifier, cycle type used for the test, the pass or fail result, and the initials of the person who ran it, kept with the daily biological indicator results.