If your Ontario practice runs a pre-vacuum autoclave, the air removal test is the daily check an inspector asks about first. For a vacuum sterilizer the test is a Bowie-Dick style pack, built around an ISO 11140-5 Type 2 indicator, and it is run at the start of each day the sterilizer is used, before the first instrument load. It answers a question the biological indicator cannot: whether the chamber actually purged its air.
That distinction matters because a sterilizer can pass a spore test and still fail to sterilise a dense load. Steam that cannot reach a surface because a pocket of air is sitting on it will not kill spores there, and the biological indicator that sits in an open basket may never see that pocket. The air removal test is the check that looks at the chamber's behaviour rather than at one sample's outcome.
What does an air removal test actually detect?
A Bowie-Dick style test measures air removal and steam penetration. A pre-vacuum cycle removes air in pulses before the steam is admitted; if the vacuum pump, the door gasket, the chamber drain or the non-condensable gas load is out of specification, some air stays behind and forms a cool pocket. The test pack puts a chemical indicator sheet in the middle of a stack of porous material, so any residual air pocket leaves a lighter, unreacted patch in the centre of the sheet after the cycle.
The result is qualitative: a uniform colour change means air removal worked, and a patchy or central light area means it did not. Because the test is about uniform steam contact, it is the fastest early warning a clinic gets that something in the vacuum system is drifting. A slow seal leak, a partially blocked drain, or a worn gasket will usually show up here before it shows up in a spore test.
Which air removal test does Ontario require on every day the sterilizer is used?
For vacuum-assisted sterilizers, Ontario practice follows the RCDSO infection prevention and control standard, which expects a daily air removal test on each day the sterilizer is used, together with daily monitoring of each cycle type. In plain terms, the sterilizer gets a Bowie-Dick style pack in an empty chamber at the start of the day, and the clinic records the result before any instruments are released.
Gravity-displacement sterilizers have no vacuum phase, so there is no air removal test to run; the daily check there is the biological indicator and the physical monitoring of the cycle. This is one of the places where a clinic that mixes both sterilizer types gets into trouble: the same log sheet is used for both machines, and the air removal row is left blank for the vacuum unit because nobody wrote down what it is for.
How do you run a Bowie-Dick style pack correctly?
Place the pack horizontally in the lowest part of the empty chamber, near the drain, where residual air and condensate collect. Never run it with an instrument load. Run the cycle type the pack is validated for, normally the pre-vacuum cycle at the temperature and exposure time you actually use for wrapped instruments.
Read the indicator sheet as soon as the cycle finishes, while the pack is still warm and the colour change is easy to see. Record the date, the sterilizer identifier, the cycle type, the result and the initials of the person who read it. If the pack is single-use, discard it; do not re-run a pack that has already been through a cycle.
| Check | Vacuum (class B / S) | Gravity (class N) |
|---|---|---|
| Air removal test | Each day the sterilizer is used | Not applicable |
| Biological indicator | Each day of use, per cycle type | Each day of use, per cycle type |
| Chemical indicator | Outside and inside each pack | Outside and inside each pack |
| Physical monitoring | Every cycle (time, temperature, pressure) | Every cycle |
| Record kept | Daily log, retained for inspection | Daily log |
How is the air removal test different from the biological indicator?
They are complementary, not substitutes. The air removal test checks whether steam reached a difficult porous load uniformly; the biological indicator checks whether the cycle actually killed a known resistant spore population. A chamber with a small air leak can pass a biological indicator in an easy load and still fail a dense one, and a chamber with a good vacuum can still under-kill if the exposure time is wrong.
This is why neither test can be dropped. Running only the spore test because it is the one the college names explicitly leaves you blind to air removal faults; running only the Bowie-Dick leaves you blind to heat and time faults. Both are recorded on the same daily log for a reason.
What do you do when an air removal test fails?
A failed test means the last load is not sterile and the sterilizer must not be used until it passes. Quarantine any load processed since the last passing test, do not release it, and re-run the affected instruments after the fault is corrected. Check the simple causes first: an over-filled chamber, a pack that was not laid flat, a drained or damaged gasket, a clogged chamber drain screen, or a vacuum pump that is not reaching its target pressure.
Then re-test on an empty chamber. A single failure that clears on a repeat, with the physical cycle parameters showing normal vacuum performance, is a different situation from two or three failures in a row, which point at a mechanical fault and need a service call. Record the failure, the corrective action and the passing retest on the same log page so the sequence is visible to an inspector.
What should your clinic do next?
Print the daily sterilization log sheet so the air removal result has its own row, and work through the sterilization self-check to see which of your machines need a weekly versus daily routine. The sterilization compliance hub collects the provincial and standard-by-standard guides in one place.
For the consumables side, the sterilization monitoring collection lists the Bowie-Dick packs, chemical indicators and 3-hour fluorescence biological indicators that match a daily schedule. If you are setting up a multi-machine clinic or a group of sites, wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Not sure which monitoring consumable fits your cycles? Start with the 24-hour biological indicator 25-pack. Still weighing whether your records would satisfy an inspection, or how a provincial requirement reads in practice? Ask a compliance specialist. Reselling? become a distributor.
Related reading
- Air Removal Testing on Pre-Vacuum Autoclaves: Daily Bowie-Dick Requirements
- Bowie-Dick Test Packs: ISO 11140-5 Type 2 and Daily Vacuum Checks
- How Often Should You Spore Test in a Dental Office? The Daily Rule
Frequently Asked Questions
Does every Ontario sterilizer need a daily air removal test?
No. The air removal test applies to vacuum-assisted (class B and class S) sterilizers. Gravity-displacement sterilizers have no vacuum phase, so there is no air removal test to run; those machines rely on the daily biological indicator together with physical cycle monitoring.
Can a sterilizer pass a spore test and still fail the air removal test?
Yes. A biological indicator in an easy, open load can pass while a small air pocket affects a dense or wrapped load. The air removal test looks at uniform steam penetration across a porous pack, so the two tests detect different faults and neither replaces the other.
What should be recorded on the daily log for the air removal test?
Record the date, the sterilizer identifier, the cycle type, the pack result and the initials of the person who read it. If the test fails, the same page should show the failure, the corrective action and the passing retest, so the sequence is visible without extra paperwork.
How many failed air removal tests mean a service call?
One failure that clears on an empty-chamber repeat with normal vacuum parameters is usually a loading or pack-handling cause. Two or three consecutive failures point at a mechanical fault such as a leaking door gasket, a blocked drain screen or a vacuum pump that cannot reach target pressure, and need a service visit.