Quick facts
- An atomic absorption spectrophotometer for clinical use identifies and measures elements and metals, such as lead and mercury, in human specimens by the wavelength and intensity of light absorbed when the specimen is converted to the atomic vapour phase.
- In the United States it is a named entry. 21 CFR 862.2850 is Class I, with product code JXR.
- Canada keeps no named entry. An instrument intended for use in diagnosis is an in vitro diagnostic device, and Part 2, Rule 4 of Schedule 1 places it in Class II.
- The instrument is one of the few routine clinical instruments that reports an element rather than a molecule, which is why the entry names lead and mercury.
- Part 2 Rule 7 applies where the instrument is used with another IVDD: both take the class of the higher-risk device.
- Entries, classes and product codes below were checked against the eCFR, the Regulations and the openFDA classification database on 2026-10-15.
An atomic absorption spectrophotometer for clinical use measures elements and metals in human specimens — lead and mercury are the two the regulation names — by the light absorbed when the specimen is turned into an atomic vapour. The United States keeps the instrument at Class I; Canada treats it as an in vitro diagnostic device and reaches Class II. This article sets out both routes and the record that goes with an elemental analysis, with the sterile 15 mL centrifuge tubes used to prepare and hold the specimen.
What is an atomic absorption spectrophotometer for clinical use?
It is the bench instrument for elemental analysis. The identification at 21 CFR 862.2850 describes a device "intended to identify and measure elements and metals (e.g., lead and mercury) in human specimens", and states the principle: the elements "are identified according to the wavelength and intensity of the light that is absorbed when the specimen is converted to the atomic vapour phase".
Three practical consequences follow from the physics. The specimen has to be brought into solution before it is vaporised, so the sample-preparation step is part of the method. The instrument reads one element at a time for most configurations, so a panel is a sequence of runs rather than one pass. And the blank and the calibration standards carry as much weight as the sample, because the measurement is a difference rather than an absolute reading.
What class is it in the United States?
It is a named entry at Class I (general controls), exempt from premarket notification under subpart E of part 807 subject to section 862.9. The openFDA device classification API returns product code JXR against the section.
The entry is unusually specific about purpose, naming lead and mercury. That specificity is useful when a laboratory is documenting a method: the instrument sits in a named section that describes the measurement rather than a general-purpose section that describes a bench tool.
| Item | Detail |
|---|---|
| US regulation | 21 CFR 862.2850 - Atomic absorption spectrophotometer for clinical use |
| US class | Class I (general controls) |
| Product code | JXR |
| Stated intended use | Identify and measure elements and metals in human specimens |
| Examples stated in the entry | Lead and mercury |
| Measurement principle | Wavelength and intensity of light absorbed in the atomic vapour phase |
| Premarket route | Exempt from premarket notification under subpart E of part 807, subject to section 862.9 |
How does Canada classify it?
Canada keeps no named entry. The instrument examines specimens taken from the human body and its readings are used in diagnosis, so it is an in vitro diagnostic device under Schedule 1, Part 2. Rule 4 is the governing text: "An IVDD that is not subject to rules 1 to 3 and that is intended to be used in diagnosis or patient management is classified as Class II, unless it falls into one of the following categories, in which case it is classified as Class III."
Elemental testing sits close to the Class III categories, which is why the stated use has to be read carefully. A method used in screening for or diagnosis of cancer, in genetic testing, in fetal congenital-disorder screening, in disease staging, or to monitor drug or substance levels where an erroneous result could lead to an imminent life-threatening decision is Class III under the same rule.
Rule 7 supplies the second route upwards: "In cases where an IVDD, including its analyzers, reagents and software, is intended to be used with another IVDD, the class of both IVDDs will be that of the IVDD in the class representing the higher risk." And Rule 8 is the residual — if rules 1 to 7 do not apply, the IVDD is Class I.
| Item | Canada | United States |
|---|---|---|
| Named entry | None; classified by rule | 862.2850 - Atomic absorption spectrophotometer for clinical use |
| Class | Class II (Part 2 Rule 4, diagnosis or patient management) | Class I |
| Class III route | Rule 4 categories, or Rule 7 where used with a higher-class IVDD | Not provided for in the entry |
| Governing text | Schedule 1, Part 2, Rules 4, 7 and 8 | 21 CFR 862.2850 |
| Typical use named | Elemental and metal testing in human specimens | Lead and mercury in human specimens |
| Licence route | Device licence (Class II) with quality-system evidence | Exempt from premarket notification, subject to section 862.9 |
Why does the class differ from the atomic absorption instrument sold to industry?
Because the intended use makes the device. An atomic absorption instrument sold to an environmental or industrial laboratory is not a medical device at all: it examines water, soil or product samples, not human specimens, and it makes no clinical claim. The moment the maker states that the instrument is for identifying and measuring elements in human specimens, it enters the in vitro diagnostic chain in Canada and a named section in the United States.
That is why the same box can carry different paperwork from two suppliers. The regulation number follows the stated intended use, so a laboratory buying an instrument for clinical work should ask for the labelling that carries the clinical claim rather than assume the hardware spec sheet settles it.
What should a laboratory check before buying one?
Five checks. First, confirm the stated intended use names human specimens, because that is what puts the instrument in the diagnostic chain. Second, confirm the elements and the concentration ranges the instrument is validated for, and whether the laboratory's own menu fits inside them. Third, confirm the sample-preparation and matrix requirements the maker states, since the digestion or dilution step is part of the method. Fourth, confirm the reference materials and the quality-control scheme the laboratory will use, because an elemental result is judged against a traceable standard. Fifth, confirm the exhaust and venting requirement, since the vapour phase is produced at high temperature.
Laboratories running an elemental method also reorder the specimen-preparation consumables used at the same bench. Clinics that stock for a whole practice can order through the wholesale programme and request a bulk quote for a standing or annual order; the sterile 15 mL centrifuge tubes sits in the same catalogue. Consumables for this line are listed under diagnostic laboratories. Reselling? become a distributor.
How does the class affect the laboratory record?
It does not. The class sets the market-entry path and the licence the supplier holds. The record is set by the laboratory quality system: sample identifier, preparation step, instrument identifier, wavelength, the calibration standards and their traceability, the blank, the control results that bracketed the run, and the operator.
Because an elemental result is a difference measurement rather than an absolute one, the blank and the standards belong in the same record as the sample. A result without them cannot be reconstructed, and reconstruction is what an audit asks for.
Sources
- openFDA device classification API
- Medical Devices Regulations, SOR/98-282, full text and Schedule 1
- US FDA, medical device databases
- US FDA, overview of device regulation
- Health Canada, medical devices
- Health Canada, establishment licensing
- ISO 15189:2022, medical laboratories, requirements for quality and competence
- ISO 17511:2020, metrological traceability of values assigned to calibrators
- ISO 13485:2016, medical devices quality management
- ISO 14971:2019, application of risk management to medical devices
- ISO 20776-1:2019, susceptibility testing of infectious agents
- Clinical and Laboratory Standards Institute
- World Health Organization
- US CDC, laboratory quality
- Public Health Ontario, infection prevention and control
- Standards Council of Canada
- CCOHS, OSH answers
A printable elemental-run sheet. The fields an atomic absorption log needs — sample, preparation, wavelength, standards, blank, controls and operator — fit on one page you can print and file. Open the compliance log centre and take the blank sheet: no account and no sign-up, and it can be copied for the whole bench.
Build the run calendar at once. Enter the laboratory name, the instrument serial and the quality-control scheme once, and the sterilization log generator produces a dated set of rows you can print or file with the analytical records.
Ask a compliance specialist. A classification question usually turns on how the device is used, not only on its name. Ask a compliance specialist — you get a written answer specific to your equipment and province, with the regulation or standard it is based on cited.
Related reading
- Flame emission photometers in Canada vs the US
- Clinical mass spectrometers in Canada vs the US
- Laboratory thermometers in Canada vs the US
CliniEco Medical supplies eye care, laboratory and dental consumables of the kind described in this article. MDEL #35334. This article is written for optometry, laboratory and dental procurement professionals and is not legal advice; the class of a device and the records a practice keeps belong to the manufacturer and to the requirements that apply to it.
Frequently Asked Questions
What class is an atomic absorption spectrophotometer in the United States?
Class I. It is a named entry at 21 CFR 862.2850 with product code JXR, exempt from premarket notification under subpart E of part 807 subject to section 862.9.
What class is it in Canada?
Class II. It is an in vitro diagnostic device intended for use in diagnosis, which Part 2 Rule 4 of Schedule 1 classifies as Class II.
Which elements does the regulation name?
The identification gives lead and mercury as examples of the elements and metals the device is intended to identify and measure in human specimens.
How does the measurement work?
The specimen is converted to the atomic vapour phase and the element is identified by the wavelength and intensity of the light it absorbs.
When would it be Class III in Canada?
Where the use falls into a Rule 4 category — cancer screening or diagnosis, genetic testing, fetal screening, disease staging, or drug-level monitoring with an imminent life-threatening consequence — or where Part 2 Rule 7 makes it share the class of a higher-risk IVDD it is used with.
Is the same instrument used in industry a medical device?
Not unless it is intended for human specimens and carries a clinical claim. The intended use is what brings an instrument into the in vitro diagnostic chain, not the hardware.