What Infection-Control Setup Is Required for Intravitreal Injections in a Canadian Retina Clinic?
An intravitreal injection takes a few minutes and consumes a tray of items that were all opened for that one patient. The drug gets the attention; the tray decides whether the visit is defensible. A retina clinic running an injection list is repeating the same short sequence dozens of times in a day, and the parts that fail are rarely the drug itself. They are the sterile 12-ply gauze sponge, 200 per box left open from the previous patient, the prep bottle decanted into an unlabelled dropper, the sharps container past its fill line, and the record that names the drug but not the lot.
This page sets out what the tray holds, which Canadian rules govern each item, what the label has to prove, and what the clinic should be able to produce months later when a lot number comes up in a recall notice.
What has to be on the tray before an intravitreal injection?
The tray is built from the sequence, not from a stock list. Anaesthetic first, then the speculum, then the prep with its contact time, then the needle and the drug, then the gauze that blots the surface afterwards. Each step consumes something that was opened for that patient, and each item has a label the clinic is expected to be able to read back.
| Tray item | Used once per patient | What the label has to show |
|---|---|---|
| Eyelid speculum | Yes | Sterile statement, lot, expiry |
| Iodine-based ophthalmic prep | Per bottle policy | Drug identification number, lot, expiry |
| Sterile gauze | Yes | Sterile statement, lot, expiry |
| 30-gauge or finer needle and syringe | Yes | Sterile statement, lot, expiry, gauge |
| Examination gloves | Yes, per pair | Size, lot |
| Procedure mask | Yes | ASTM level, lot |
| Sharps container | Per fill line | Model, fill line, closure |
Two things about that table are worth pausing on. The first is that the sharps container is the only item on the list whose useful life is measured in fill rather than in time. The second is that the prep is the only drug on the tray, and it is the item most often moved out of its original packaging.
Why is the prep iodine-based, and what stays off the eye?
The ocular surface tolerates an iodine-based prep and does not tolerate the alternatives that are routine on skin. Alcohol is used to degrease the lids and lashes in some clinics, and it is kept away from the cornea. Chlorhexidine is a skin and instrument antiseptic, not an ocular one, and it does not belong on this tray at all. The practical consequence is that the injection room needs two clearly separated prep stations, because the one that is right for the eyelid is wrong for the eye.
Contact time is the other half of the prep. Iodine needs to sit on the tissue before the needle goes in, and the wait has to be written into the procedure so that a busy list does not compress it. A clinic that measures the wait has a number to show; a clinic that relies on the injector's sense of timing has a habit.

Is the iodine prep a drug or a device in Canada?
It is a drug, and the definition explains why. Section 2 of the Food and Drugs Act defines a drug to include a substance represented for use in the diagnosis, treatment, mitigation or prevention of a disease, disorder or abnormal physical state. A solution applied to the conjunctiva before a needle enters the eye is represented for exactly that purpose, so it is authorised as a drug and carries a drug identification number rather than a device licence.
That distinction is not academic at the buying desk. A clinic that treats the prep as a consumable and orders it beside the gauze is mixing two regulatory routes in one purchase order, and the invoice will not tell it so. The device lines on the tray carry a licence and a manufacturer's quality system; the prep carries a drug identification number and a lot that has to be recorded against the patient.
Which items must be single-use, and what does the label have to prove?
Everything that enters the eye or touches the ocular surface is single-use: the needle, the syringe, the speculum, the gauze and, in most clinics, the prep applicator. Section 17 of the Medical Devices Regulations requires a device sold sterile to be manufactured and put through a validated method under controlled conditions, and section 21 requires the pack to carry the word Sterile where that is intended, an expiry date set by the component with the shortest projected useful life, and an indication of what the package contains. ISO 11607-1:2019 is the standard behind the pouch that keeps the sterile barrier intact, and ISO 15223-1:2021 is the standard behind the symbols printed on it.
The label is therefore doing more than decorating the pack. It is the clinic's proof that the item in the drawer is the item that was validated. Materials in contact with the ocular surface are evaluated under ISO 10993-1:2025, with cytotoxicity tested under ISO 10993-5:2009 and sensitization under ISO 10993-10:2021, so a pouch from an unlisted source is not simply a substitute.
| Label element | Why it matters on this tray |
|---|---|
| Word Sterile | Confirms the item was supplied in a sterile condition |
| Lot or control number | The number the recall notice will quote |
| Expiry date | Set by the component that expires first |
| Single-use symbol | Removes the question of reuse from the room |
| Contents or size | Gauge, ply and count, so the wrong item is caught early |
A clinic that opens a single-use item and puts it back on the shelf has stepped outside the arrangement the label describes. Section 17 places the validated method on the manufacturer; the second use has no method, no validation and no label behind it.

Does the injection need a dedicated procedure room?
It needs a controlled sequence, and a room is one way to get one. What the sequence requires is a defined clean surface, hand hygiene before gloves, gloves that are changed between patients, a mask on the injector, separation between the clean tray and the used sharps, and a light the injector can work under without leaning across the field.
Retina clinics often run injections in a shared minor procedure room, which works when the tray is set up per patient and the surface is cleaned between cases. What does not work is a room where the clean tray and the used sharps share a counter, because the reach across becomes the error. The same principle is why hand hygiene belongs at the door and not at the sink across the corridor: CCOHS notes that hand washing is the single measure that most reduces the spread of common infections, and a sink in the wrong place is a step that gets skipped.
How is the used needle handled, and what leaves the clinic?
The needle goes into the sharps container at the point of use, without recapping, and the container is closed at the fill line rather than at the brim. A one-quart container is the format most minor procedure rooms use because it fills within a reasonable cycle and can be sealed with one hand. Reusable sharps containers are specified under ISO 23907-2:2019, and the fill line on the container is the limit that makes the closure work.
What leaves the clinic is the sealed container plus the record of where it went. In Ontario, biomedical waste is tracked by the generator through the Hazardous Waste Information Network, and the carrier's manifest is the document that closes the loop. Where waste crosses a provincial or national border, the Transportation of Dangerous Goods Regulations apply to the shipment and the consignor holds the paperwork, the same way a laboratory does when it sends specimens out.
What belongs in the injection record?
Date and time, which eye, the drug and its lot, the dose, the needle gauge and lot, who performed the injection, the consent, and anything unexpected during or after the visit. The record should also be able to answer a question nobody plans for: if a lot number is published in a recall notice, can the clinic list the patients who received it before the end of the day? A record that stores the drug lot but not the device lot answers half of that question.
Recalls are published on the Government of Canada recall and safety alert service, and the useful routine is not to read it daily but to be able to run the query when a supplier writes. That takes a record format where the lot is a field, not a sentence inside a note.
Is the Canadian setup different from the American one? Less than the paperwork suggests. The aseptic sequence is the same, but the routes differ: devices reach the Canadian market through a licence and an establishment licence held by the importer, while the same product in the United States goes through a premarket notification, and the antiseptic on the tray is a drug in both countries with its own identifier. A clinic buying cross-border should ask which of the two routes each item came through, because the answer decides who holds the documentation.
Running an ophthalmic or multi-specialty clinic? The clinic consumables page groups the tray, glove and sharps lines that an injection list consumes together, and the 3.5 mil nitrile examination glove, 1,000 gloves per case is the format most rooms standardise on for a full list. A one-quart puncture-resistant sharps container, 20 per case keeps the fill cycle predictable, and a three-ply ASTM Level 2 procedure mask covers the mask worn at the slit lamp.
Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.
References
- ISO 10993-1:2025 — Biological evaluation of medical devices, evaluation and testing within a risk management process (checked 5 November 2026)
- ISO 10993-5:2009 — Biological evaluation of medical devices, tests for in vitro cytotoxicity (checked 5 November 2026)
- ISO 10993-10:2021 — Biological evaluation of medical devices, tests for skin sensitization (checked 5 November 2026)
- ISO 11607-1:2019 — Packaging for terminally sterilized medical devices, requirements for materials and sterile barrier systems (checked 5 November 2026)
- ISO 15223-1:2021 — Medical devices, symbols to be used with information to be supplied by the manufacturer (checked 5 November 2026)
- ISO 13485:2016 — Medical devices, quality management systems, requirements for regulatory purposes (checked 5 November 2026)
- ISO 17665:2024 — Moist heat sterilization of health care products, requirements for the development, validation and routine control (checked 5 November 2026)
- ISO 11135:2014 — Ethylene oxide sterilization of health care products, requirements for development, validation and routine control (checked 5 November 2026)
- ISO 14971:2019 — Medical devices, application of risk management to medical devices (checked 5 November 2026)
- ISO 23907-2:2019 — Sharps injury protection, requirements and test methods, reusable sharps containers (checked 5 November 2026)
- Food and Drugs Act, R.S.C. 1985, c. F-27, section 2 definitions (checked 5 November 2026)
- Medical Devices Regulations, SOR/98-282 (checked 5 November 2026)
- Transportation of Dangerous Goods Regulations, SOR/2001-286 (checked 5 November 2026)
- CCOHS — Hand washing: reducing the risk of common infections (checked 5 November 2026)
- CCOHS — Chemical protective clothing, glove selection (checked 5 November 2026)
- Ontario Ministry of the Environment — Hazardous Waste Information Network (checked 5 November 2026)
- Recalls and safety alerts, Government of Canada (checked 5 November 2026)
Printable companion: the injection tray and lot log templates carry the same columns as the tables below, so the tray is checked against a sheet that can be filed with the visit.
Sizing the tray lines against an injection list? The clinic supply planner converts injections per clinic day into case quantities per item.
Related Reading
- Ophthalmic Procedure Kits in Canada vs the US: Which Class Applies?
- Eye Shields for Canadian Eye Clinics: Post-Mydriatic, Post-Operative and UV Types
Frequently Asked Questions
What has to be on the tray for an intravitreal injection?
An eyelid speculum, an iodine-based ophthalmic prep, sterile gauze, a 30-gauge or finer needle with its syringe or a prefilled syringe, gloves, a procedure mask and the drug itself. Everything that touches the ocular surface is opened from its own pack in front of the patient.
Is alcohol ever used on the eye before an injection?
No. Alcohol is used on the lids and lashes in some clinics to degrease the skin, and it is kept off the ocular surface because it irritates the cornea. The conjunctival prep is iodine-based, and the two are stored separately so the wrong bottle is not picked up in a dim room.
Why do clinics wait after applying the iodine prep?
The prep needs contact time on the tissue before the needle goes in, and the wait is part of the written procedure rather than a matter of habit. A tray set up in advance with the injection drawn up shortens that window, which is why the sequence is documented and timed.
Can the iodine prep be poured into a smaller bottle for the room?
Not without losing the label. A secondary container carries no lot, no expiry and no drug identification number, and once the original bottle is out of sight the record cannot show what was applied. Keep the prep in the container the manufacturer labelled.
Does the clinic need a licensed device to do intravitreal injections?
The procedure is a medical act and sits with the physician's licence. What the clinic buys as consumables are devices, and the importer that supplies them holds a medical device establishment licence, which is the number a buyer asks for when setting up an account.
How often should the injection tray be reviewed?
At the start of every injection list and again when a new lot is opened. A tray check is a two-minute task that produces a dated line in the record, and it is the line an auditor reads first when a lot number has to be traced.
Are prefilled syringes treated differently from vials?
Yes at the label level. A prefilled syringe is a device supplied with the drug, so the pack carries both the device labelling and the drug labelling, and the expiry is whichever of the two comes first. A separate vial and syringe carry two lots that both go in the record.
What happens to the used needle and the empty drug container?
The needle goes straight into the sharps container at the point of use, without recapping. The empty container follows the same sharps or drug waste route the clinic's carrier accepts, and the manifest records it, which is what closes the loop at the end of the quarter.
Last updated: October 2026. CliniEco Medical is a licensed medical device establishment (MDEL #35334).