What Documentation Applies to Dialysis Water and Dialysate Testing in Canada?
A renal unit in Hamilton checks for chlorine at the start of every shift, and its monthly culture comes back from a laboratory a week later. Both readings land in one folder, and that folder is what an inspector reads first: not the treatment, the record of the water. Gloves and wipes come off the shelf at the chair, and a 1,000-glove case of 4.0 mil nitrile exam gloves scales with treatment volume. The water record is what gets audited.
What records does a dialysate testing programme have to produce?
| Record | Written when | What it must show |
|---|---|---|
| Chlorine and chloramine check | Start of each shift | Reading, method, point and initials |
| Water hardness and softener | Daily | Result and regeneration note |
| Conductivity and temperature | Each shift | Value and meter used |
| Microbiological sample | Per programme interval | Sample point, count, endotoxin result |
| Corrective action | After an exceedance | Retest result and disinfection performed |
The interval belongs to the unit's programme. The fields belong to good practice: a result with no sample point and no time is a number, not a record.

Which results belong in the log?
Water and dialysis fluid are sampled at different points, and the current standards treat them as separate qualities: ISO 23500-3 sets the water the unit produces, ISO 23500-5 sets the fluid delivered to the patient. A log that files both under "water" cannot show which limit a result was compared against, which is the first thing a reviewer asks when a count looks high.
Who reviews and signs the record?

Two roles, not one. The person who took the reading writes it down and initials it; a named reviewer signs the review, and that signature is what turns a sheet into a record. When a result goes past an action level, the file needs the retest, the disinfection performed and the date it was cleared, so the sequence is visible without asking anyone to remember it.
How long should water records be kept?
The retention period is set by the facility's records policy and the renal programme it reports to, and it is usually longer than the equipment's service life. The practical test is whether a record from three years ago can be produced with its sample points intact. Keep water logs beside maintenance records for the same system: a disinfection date with no maintenance note, or the reverse, is harder to defend than either alone.
Which ISO 23500 editions are current in 2026?
Both parts were reissued in 2024. ISO 23500-3:2024 covers water for haemodialysis and related therapies, ISO 23500-5:2024 covers the quality of dialysis fluid, and the 2019 editions of both are withdrawn. A unit manual, a validation report or a supplier sheet still citing the 2019 parts is quoting superseded text, which is worth catching before an inspector does.
Between treatments, the surface and the waste stream follow the same discipline. A 160-wipe canister of biodegradable disinfecting wipes covers the station, and a 30-gallon leak-proof biohazard waste bag matches the tote the hauler lifts. Neither fixes a water system, and both show up in the same inspection walk.
Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.
References
- ISO 23500-3:2024 — Water for haemodialysis and related therapies (checked 4 November 2026)
- ISO 23500-5:2024 — Quality of dialysis fluid for haemodialysis and related therapies (checked 4 November 2026)
- Public Health Ontario — Infection Prevention and Control (checked 4 November 2026)
Printable companion: the free test and review log templates carry a sample point and a reviewer line, which is what makes a reading auditable later.
Turning treatment numbers into stock levels? The clinic supply planner converts treatment bays and shifts into gloves, wipes and waste bag quantities.
Related Reading
- Dialysis Water and Dialysate Culture Monitoring Frequency for Canadian Units
- Dialysis Machine and Dialyzer Disinfection SOP: What the Log Must Record
Frequently Asked Questions
What has to be recorded after a chlorine check?
The reading, the method or meter used, the point it was taken from, the time, and initials. A number without a point and a time cannot be compared against a limit later.
How are water and dialysate samples treated differently?
ISO 23500-3 covers the water the unit produces and ISO 23500-5 covers the dialysis fluid delivered to the patient, so they are different qualities with different sample points. A log that calls both "water" loses the comparison.
Who signs a dialysis water log?
The person who took the reading records it, and a named reviewer signs the review. A correction is struck through once, initialled and dated, never overwritten.
Which edition of ISO 23500 applies now?
ISO 23500-3:2024 and ISO 23500-5:2024 are the published editions; the 2019 editions of both parts are withdrawn, so a unit manual citing 2019 is quoting withdrawn text.
Last updated: October 2026. CliniEco Medical is a licensed medical device establishment (MDEL #35334).