Where should the otoscope speculum reprocessing log be filed at a Quebec family practice, and how many years must it stay retrievable?
A family practice owns six otoscope heads, one bottle of surface wipes and a notebook by the sink. The notebook has dates and ticks. It does not have a cycle number, a device list or a name attached to any of the ticks.
Is an otoscope speculum critical or semi-critical?
Semi-critical. It touches intact mucous membrane or non-intact skin without entering a body cavity, and the provincial guidance is explicit that semi-critical devices need, at minimum, high-level disinfection after cleaning. A speculum of that class covers more than the ear: the same box is usually stocked for nasal examinations, and the same rules follow it. Single-use specula remove the record entirely, which is why outreach kits and locum bags carry them.
Which records does the log actually consist of?
Three, and only one of them is the notebook. Provincial guidance groups reprocessing records into quality records, maintenance records and use records. The quality record holds the cycle data — parameters verified, no alarm code, indicator and biological results. The maintenance record belongs to the sterilizer. The use record is the one that links a device to a patient, so a failed load can be traced forward.

| Record | What it holds | Where it is filed |
|---|---|---|
| Quality (load) | Cycle number, load contents, indicator and BI result, unload initials | At the reprocessing station, in the cycle register |
| Maintenance | Service dates, parts, verification load | In the sterilizer's own file, by serial number |
| Use | Device and cycle date against the patient seen | In the patient chart |
The Class 4 self-sealing pouch with dual indicators is what makes the load record readable: the pouch in the tray carries a number the register can name.
Where should each part be filed?
The load record stays at the station, in the order the cycles ran, with the pouch count and the unload initials — not rewritten at month end. The maintenance file lives with the sterilizer, keyed by serial number. The patient link goes where the visit is documented: a cycle date and the site treated, short enough that a busy clinician writes it.

How many years must it stay retrievable?
Ten years after the date of the last entry is the provincial benchmark for reprocessing registers, structured as one year on site and nine in archives. Clinical records kept by physicians follow their own ten-year rule from the last entry, with paper-only records that existed before 11 July 2024 held at five. The waste regulation's five-year floor (Q-2, r. 12, s. 16) covers waste shipment documents, including the bag that a used single-use speculum leaves in.
What happens when a locum runs the station for a week?
Someone who has not unloaded a load at that practice before needs two things on the record: an attestation that they were oriented to the station, and initials on the load line that match the staff roster. A Class 5 integrator in the load, read by whoever unloads it, is what turns that week into an entry an inspector can follow. Surface wipes, such as a tub of peroxide disinfecting wipes, are logged on the cleaning side and never stand in for the semi-critical entry.
Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.Running a family practice? See our family practice supplies and reprocessing lines.
References
- Institut national de santé publique du Québec, Retraitement des dispositifs médicaux : les essentiels du contexte réglementaire et des bonnes pratiques (2026), device classes and register types (checked 22 November 2026)
- CSA Z314:23, Canadian medical device reprocessing in all health care settings (checked 22 November 2026)
- Collège des médecins du Québec, Conservation et destruction des dossiers (checked 22 November 2026)
Printable companion: the free load record sheet carries the cycle number, the device list and the unload initials, so a speculum tray stays traceable without a second notebook.
Not sure the station would pass a walk-through? The sterilization self-check walks the room in the order an inspector does.
Related Reading
- How are ENT examination instruments reprocessed in Canadian physician clinics?
- Sterilization log retention in Canada vs the US: what inspectors expect
- What instrument and waste rules apply to cerumen management in Canadian audiology?
Frequently Asked Questions
Is an otoscope speculum critical or semi-critical?
Semi-critical: it touches intact mucous membrane without penetrating it, so the minimum is high-level disinfection after cleaning. Single-use specula avoid the question entirely, which is why many practices keep them for outreach kits.
Where does the patient link go?
In the chart, as the cycle date and the site treated. The load record at the station carries the cycle number and the tray contents, and the two are joined by date when a load result turns out positive.
How long does the whole record have to be kept?
Ten years after the last entry is the benchmark for reprocessing registers, held one year on site and nine in archives. A clinical record that refers to the cycle follows clinical record retention, which is also ten years for physicians.
Last updated: November 2026. CliniEco Medical is a licensed medical device establishment (MDEL #35334).