Quick facts
- A membrane filter unit is a holder, funnel or vacuum assembly that drives a liquid sample through a membrane disc so particles or organisms are caught on the surface of the disc.
- In the United States, general purpose laboratory equipment that is labelled or promoted for a specific medical use is filed under 21 CFR 862.2050 at Class I, with the membrane filter unit under product code JRL.
- In Canada the unit is a non-invasive device and falls to Class I under Rule 7(1) of the Medical Devices Regulations.
- Rule 7(2) lifts it to Class II in two cases: it acts as a calibrator, tester or quality control support to another medical device, or it is connected to an active device already classified II, III or IV.
- So the class of a filter unit is set by what it is connected to and what it is claimed to do, not by the membrane it holds.
- Entries and rule text were checked against the current eCFR text, the Regulations and the openFDA classification database on 2026-10-11.
A membrane filter unit holds a membrane disc and pushes a liquid sample through it under vacuum or pressure. In the United States a general purpose filter unit promoted for a medical use is a named Class I entry; in Canada the same unit is a non-invasive device whose class depends on what it is attached to. This article covers both routes, the two situations that raise the Canadian class to Class II, and the standards a laboratory reads when it specifies a unit.
What is a membrane filter unit?
A filter unit is the hardware that holds a membrane disc and forces a sample through it. The common bench format is a funnel clamped over a membrane on a fritted base, connected to a vacuum line so that the liquid is pulled through and the particles are caught on the disc. A syringe-format unit drives the sample through by hand. In both cases the unit itself does not enter the sample path as an ingredient; it is a piece of equipment that produces a filter with a countable surface.
That makes it a laboratory device rather than a reagent. The result of the test lives on the membrane, not in the unit, and the unit is reused after cleaning or is a single-use item. Because the unit is equipment, the regulator's interest is in what the unit is claimed to do and what it is joined to on the bench.
Is a membrane filter unit a Class I device in the United States?
Yes, in the ordinary case. 21 CFR 862.2050 names General purpose laboratory equipment labeled or promoted for a specific medical use and classifies it at Class I (general controls). The openFDA device classification API files the membrane filter unit under product code JRL within that regulation number.
The important words in the US title are labelled or promoted for a specific medical use. A plain laboratory funnel sold as glassware is not the subject of this entry. A filter unit that the maker presents as equipment for a medical laboratory test is. Class I means the unit reaches the market under general controls and registration rather than a premarket review of the individual item.
How does Canada classify a membrane filter unit?
Canada looks at the device rather than the entry. A filter unit does not enter the body and does not contact the patient, so it is a non-invasive device and Rule 7(1) of the Medical Devices Regulations applies: "Subject to subrule (2), all other non-invasive devices are classified as Class I."
The subrule is the part that matters. Rule 7(2) raises the unit to Class II if it is intended to act as a calibrator, tester or quality control support to another medical device, or if it is intended to be connected to an active device that is itself classified as Class II, III or IV. A passive funnel on a vacuum line is Class I; the same funnel plumbed into an instrumented sampling system may not be.
| Item | Canada | United States |
|---|---|---|
| Named entry | None; classified by rule | 21 CFR 862.2050, general purpose lab equipment |
| Product code | Not applicable | JRL |
| Default class | Class I (Rule 7(1)) | Class I (general controls) |
| Lift to Class II | Rule 7(2)(a) or 7(2)(b) | Not by class; entry stays Class I |
| Route to market, Class I | Establishment licence; no device licence | General controls; registration and listing |
| Route to market, Class II | Device licence under section 32 | Not applicable |
When does a filter unit become Class II in Canada?
There are two triggers, and both are about the bench rather than the funnel. The first is the calibrator, tester and quality control support limb in Rule 7(2)(a). If the unit is sold to verify, check or support another medical device, the rule treats it as more than equipment. The second is the connection limb in Rule 7(2)(b): a unit intended to be joined to an active device already classified II, III or IV is drawn up with that device.
| How the unit is used | Canadian rule | Canadian class | US entry |
|---|---|---|---|
| Passive funnel on a vacuum line | Rule 7(1) | Class I | 862.2050, Class I |
| Holds a disc for a manual colony count | Rule 7(1) | Class I | 862.2050, Class I |
| Sold to test or verify another device | Rule 7(2)(a) | Class II | 862.2050, Class I |
| Plumbed into a Class II sampling instrument | Rule 7(2)(b) | Class II | 862.2050, Class I |
The practical message is that a laboratory should not assume a filter unit is Class I in Canada just because it is Class I in the United States. The class follows the intended connection and use, and that intended use is recorded in the maker's instructions and the bench's own procedure.
Which standards apply to membrane filtration?
The class says what the device is; the standards say how the work must be done. ISO 15189:2022 sets out what a medical laboratory must do to show competence, and it is the frame that makes an equipment record worth keeping. A manufacturer that builds laboratory equipment to a management system does so under ISO 13485:2016, and the risk file follows ISO 14971:2019.
In Canada the laboratory's own quality frame is supported by Public Health Ontario and by the Clinical and Laboratory Standards Institute, whose method documents are the ones a technologist reads when a filtration step has to be written down. The US CDC laboratory quality pages carry the equivalent US material.
How does a laboratory specify and record a filter unit?
Specify by the membrane the unit is built for, by the funnel volume and by the joint that connects it. A unit that takes a 47 mm disc and a 100 mL funnel is chosen for the volumes the bench actually runs, and the joint is chosen so the unit fits the manifold already in place. Standardising to one disc size keeps the restock simple and keeps the counts comparable between technologists.
Record by what leaves the unit. The consumables that travel with a filtration bench - graduated leak-proof sample cups for the sample as it arrives, cotton swabs for bench work and individually wrapped swabs where a sterile single swab is needed - are restocked as a set, and the bench keeps its equipment log beside them. Laboratories that buy at case level use the B2B wholesale collection and the wholesale and multi-site ordering page. Reselling? become a distributor.
Sources
- openFDA device classification API
- Medical Devices Regulations, SOR/98-282, full text and Schedule 1
- US FDA, medical devices
- US FDA, medical device databases
- US FDA, overview of device regulation
- Health Canada, medical devices
- Health Canada, establishment licensing
- Standards Council of Canada
- CSA Group
- ISO 13485:2016, medical devices quality management
- ISO 14971:2019, risk management of medical devices
- CCOHS, OSH answers
- Clinical and Laboratory Standards Institute
- Public Health Ontario, infection prevention and control
- ISO 15189:2022, medical laboratories, requirements for quality and competence
- US CDC, laboratory quality
- World Health Organization
- 21 CFR 862.2050, General purpose laboratory equipment labeled or promoted for a specific medical use
- Public Health Ontario, infection prevention and control
- Clinical and Laboratory Standards Institute
A filtration bench is run to a written method, and the method has to be printable and kept. The printable blank equipment and method sheet at the compliance log centre gives a lab columns for the unit, the membrane lot, the operator and the result, and it opens without an account and no sign-up.
Where a laboratory records the same step many times a day, the log generator on that page builds a sheet sized to the number of benches and shifts the lab runs, so a single-bench laboratory is not handed a form built for a hospital.
Have a question about your own facility? Send it in and you will get a written answer specific to your setup, with the regulation or standard it is based on cited. Ask a compliance specialist about your equipment and province.
Related reading
- Laboratory glassware in Canada vs the US: which standards apply
- Multiwell cell culture plates in Canada vs the US: which class
- Phase contrast microscopes in Canada vs the US: which rules
CliniEco Medical supplies eye care, laboratory and dental consumables of the kind described in this article. MDEL #35334. This article is written for optometry, laboratory and dental procurement professionals and is not legal advice; the class of a device and the records kept by a practice belong to the manufacturer and to the requirements that apply to it.
Frequently Asked Questions
Is a membrane filter unit a regulated device?
In the United States, general purpose laboratory equipment promoted for a specific medical use is a device at 21 CFR 862.2050, Class I. In Canada, a filter unit is a non-invasive device classified under the rules in Schedule 1 of the Medical Devices Regulations.
What class is a filter unit in Canada?
Class I under Rule 7(1), unless Rule 7(2) applies. Rule 7(2)(a) lifts it to Class II if it is a calibrator, tester or quality control support to another device, and Rule 7(2)(b) lifts it to Class II if it is intended to be connected to an active device classified II, III or IV.
What is the US product code for a membrane filter unit?
The openFDA classification database lists Unit, Filter, Membrane under product code JRL, within regulation number 862.2050 at Class I.
Does a plain glass funnel fall under 21 CFR 862.2050?
Not on its own. The entry covers general purpose laboratory equipment that is labelled or promoted for a specific medical use. A funnel sold simply as glassware is not presented for a medical use and is outside the entry.
Does a Class II filter unit need a licence in Canada?
Yes. A Class II device needs a device licence under section 32, supported by a CAN/CSA-ISO 13485 quality management certificate. A Class I device needs no device licence of its own.
Which standard frames a laboratory's equipment records?
ISO 15189:2022 sets the requirements for quality and competence in a medical laboratory and is the frame for the equipment and method records. ISO 13485:2016 covers a manufacturer's quality system and ISO 14971:2019 the risk file.
How should a filtration step be recorded?
As a written method with the unit, membrane lot, operator and result. The compliance log centre provides a printable blank method and equipment sheet and a log generator sized to the number of benches a laboratory runs.