Handling Biological Indicators from Delivery to Storage: Step-by-Step
The box arrives at reception between patients, gets opened on the front desk, and the vials are put wherever there is space. That ten-minute sequence decides more about a monitoring record than the reading of any single result, because every later question about a failed test begins with the lot number printed on the carton now sitting under the counter.
The 24-hour self-contained biological indicator, 50 tests stays usable only if it is stored as the pack instructs, and the handling below is written for a clinic taking a case or two a year, not a hospital taking pallets.
What do I check when the box arrives?
Four things, before the box leaves the front desk: the outer packaging is intact and dry, the count matches the packing slip, the lot number and expiry are legible on every inner pack rather than only the carton, and the storage statement on the pack can be met by the place the clinic intends to keep it.
| Stage | Check | What fails here | What to file |
|---|---|---|---|
| Delivery | Carton intact, no water damage | Damp cartons signed for unopened | Signed packing slip with the date received |
| Unpacking | Count against the order, inner packs present | Short shipments found weeks later | Lot and expiry copied to the stock card |
| Label check | Lot, expiry, organism, intended process readable | A date unreadable at the point of use | Pack label kept with the lot until finished |
| Storage | Stated temperature, original packaging, no sun | A cupboard above a sterilizer | Storage location named in the written procedure |
| Rotation | Oldest lot in front, open date marked | Two open packs of different lots in one drawer | Open date written on the pack |

Where should the vials live?
Where the pack says, which for most self-contained indicators is a dry cupboard at room temperature. Three locations fail repeatedly in small clinics: the refrigerator, because it feels like it preserves things; the shelf above the sterilizer, because it is convenient and warm; and the treatment room drawer, where a week's supply is carried and then stays for a month.
The original packaging is part of the storage instruction, not waste paper: it carries the lot and expiry and keeps the strip out of the light.
How do I rotate stock so an expired vial is never read?

Write the open date on the pack, keep the oldest lot at the front, and check the expiry at the point of use rather than at purchase.
Never keep two open packs of different lots in the same drawer, since that is how a record ends up naming a lot the operator did not use. And stop using a lot with enough vials left to finish the current week, so the last days of a lot never become a reason to stretch a date.
What paperwork belongs with the lot?
The certificate of analysis, the pack label or a copy of it, and the stock card showing date received and date opened. The 24-hour indicator, 25 tests suits a single-sterilizer practice that would rather not hold a large lot, and the 24-well dry-block incubator is where the vial goes next. Both are stocked for dental practices and care homes ordering by the case.
Ordering for a clinic, a lab or a care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.
TSA:
Printable companion: the sterilization log sheet opens with a receiving block, so the lot number is written down when the box is opened rather than recalled later.
TSB:
Want to see where a lot sits in your own record? The sterilization self-check traces a result back to the lot and shows which field is missing.
STERILE: 1
References
- CDC — Guidelines for Infection Control in Dental Health-Care Settings, MMWR Recommendations and Reports 52(RR-17) (checked 18 November 2026)
- Medical Devices Regulations (SOR/98-282), section 44 — establishment licence requirement (checked 18 November 2026)
Not sure how this applies to your own setup? Ask a sterilization compliance specialist — you’ll get a written answer specific to your equipment and province, with the regulation or standard cited.
Related Reading
- Biological Indicator Certificate of Analysis: What Canadian Clinics Should File
- 24-Hour Self-Contained Biological Indicators: Pack Sizes, Incubation and the Record
Frequently Asked Questions
Should spore tests be kept in the fridge to stay fresh?
No. A refrigerator is where condensation forms, and moisture on the strip or in the cap changes what grows. Store at the range printed on the pack, in the original packaging, away from heat and direct sun.
What if a vial is dropped before use?
Treat it as unusable and record it. A cracked cap or a displaced strip can still look intact, and neither can be inspected from outside, so discard it and take the next one from the pack.
How long before expiry should a clinic stop using a lot?
Before the date, with enough vials left to finish the working week. A practice running daily monitoring should not reach the last days of a lot and have to decide whether a date still applies.
Last updated: November 2026. CliniEco Medical is a licensed medical device establishment (MDEL #35334).