Failed spore test: the load recall order a clinic follows, step by step
The tube comes out of the incubator with growth in it, and every pack processed since the last clean test is now in question. What follows is a sequence, not a discussion.
What gets quarantined after a positive spore test?
Everything sterilized since the last negative biological indicator, plus the test load itself. In practice that means clearing the sterile shelf into a labelled tote, noting each pack's autoclave indicator tape lot beside it in the cycle log, and telling the front desk that no pack from the tote goes to a chair until release is signed. Implantable and critical items get a separate label, because they stay held until the investigation closes rather than until the next clean test.
The 24-hour biological indicators now matter twice: the positive tube is evidence, and the repeat test is the thing that ends the hold.
How far back does the hold run in Ontario and Quebec?

| Province | Monitoring rhythm | Typical hold window |
|---|---|---|
| Ontario | biological indicator each day the unit is used | back to yesterday's clean test |
| Quebec | weekly test | back to last week's clean test |
That single difference decides how many packs sit in the tote. It is also why a clinic that changed provinces, or bought a used unit from another province, should check which rhythm its own college asks for rather than inheriting the previous owner's habit.
What do you re-test before blaming the sterilizer?
Four things, in this order. Read the control tube from the same lot — a control that fails to grow invalidates the run. Check the incubator: a dry block that has drifted will read a failing tube as passing, or the reverse, so confirm the set point and that the wells were loaded correctly. Open a fresh indicator lot. Then repeat the test with a normal load in the chamber.

Only after a repeat failure do you move to the machine: door gasket, chamber drain strainer, air removal on a pre-vacuum unit, and load density. Over-packed trays and heavy instrument sets are the two causes a clinic can fix the same afternoon.
Who has to be told, and what do they get?
The clinical lead decides what happens to patients who received held instruments; your job is to give them the list fast and in writing — load date, items, cycle number, patient list if the practice tracks it. Office management gets the hold notice and the release signature. A service engineer gets the repeat result, not the first one.
How do you close the record?
One page: what failed, what was held, when it was released, what the repeat showed, what changed, who signed. Add the service report if an engineer attended. Then re-run the biological indicator incubator check on the weekly sheet, because a hold that ends without a machine check tends to come back.
Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.
References
- CSA Z314:23 — Reprocessing of medical devices in health care settings (checked 21 November 2026)
- Royal College of Dental Surgeons of Ontario — infection prevention and control standard (checked 21 November 2026)
- Ordre des dentistes du Québec — prévention et contrôle des infections, retraitement (checked 21 November 2026)
- ISO 11138-1:2017 — Biological indicators for sterilization processes (checked 21 November 2026)
Printable companion: the free sterilization log library includes a non-conforming load sheet that records the hold, the release and who signed it.
Rebuilding stock after a hold burns through indicators? The clinic supply planner works out how many tests a month of cycles needs.
Not sure how this applies to your own setup? Ask a sterilization compliance specialist — you’ll get a written answer specific to your equipment and province, with the regulation or standard cited.
Related Reading
- After a Failed Spore Test: Quarantine, Retest and Release in a Canadian Clinic
- Spore Test Positive: What a Clinic Must Do Before Using the Load Again
- Spore Test Frequency After a Major Sterilizer Repair
Frequently Asked Questions
Does one positive tube stop the whole practice?
It stops the loads it covers, not the practice. Hold everything processed since the last negative test, run the investigation, and release only after a clean repeat, a passing chemical indicator check and a decision recorded by name.
Can we re-sterilize the held packs instead of discarding them?
Depends on the packaging, and this is where clinics get it wrong. Single-use pouches are not designed for repeat cycles unless the instructions say so; instruments can usually be reprocessed from cleaning. Decide per item and write the decision down.
How do we know the test itself was valid?
Read the control. A control that fails to grow, a tube incubated in a block outside 56 to 60 C, or an expired indicator lot makes the result invalid rather than positive. That is a different investigation.