Class 5 Integrators in Gravity Displacement Cycles: What They Can and Cannot Show

A Class 5 integrator can be used in a gravity displacement cycle, and it will respond to the time, temperature and steam contact the cycle delivers; what it cannot do is test air removal, because a gravity cycle has no vacuum stage to test. So it is a useful per-pack record and a poor substitute for the two tests that actually answer the harder questions. The CliniEco Class 5 steam chemical integrator, ISO 11140-1 Type 5, 100 per boxClass 5 steam integrators, 100 per pack are the usual inside-the-pack record for both cycle types.

What does a Class 5 integrator respond to?

All the critical parameters of the steam cycle: time, temperature and the presence of saturated steam. It is calibrated to a defined response, which is why it can stand in for the pack-level question "were the conditions reached here?" better than a printed Class 4 indicator does.

CliniEco Class 5 steam chemical integrators, 100 per pack

It is defined in ISO 11140-1, and the definition is about response rather than about what the clinic does with the result. A pass is evidence, not a release decision on its own.

How is a gravity cycle different from a pre-vacuum cycle?

A gravity cycle relies on steam displacing air from the top of the chamber downwards, with the air leaving through a drain. There is no vacuum stage, so air removal depends on the load arrangement and the steam flow rather than on a pump.

CliniEco 24-hour biological indicators, 25 per pack

That makes load configuration more important, not less. In a gravity cycle a tightly packed chamber, a tray with a solid base or a pouch lying flat on another pouch can all trap air that no pump will remove.

Question Class 5 integrator Biological indicator Bowie-Dick test
Parameters reached at this point Yes Indirectly No
Spores killed No Yes No
Air removal in a vacuum cycle No No Yes
Applies to a gravity cycle Yes Yes No

What can a Class 5 integrator show in that cycle?

It shows that the conditions at its position met the response specification. If it sits in the middle of a dense set and passes, that is meaningful evidence about the hardest position in that pack.

It also shows variation between loads. A set of integrator results that usually read well and occasionally read poorly is a signal about loading practice, and that signal is hard to get from tape or from a once-a-day biological indicator.

What can it not show?

It cannot show that spores were killed, and it cannot show that air was removed from the chamber. Those are the biological indicator question and the air removal question, and neither is answered by a chemical integrator in a gravity cycle.

It also cannot show that a pack was loaded correctly. An integrator on top of a loose pouch measures the easy position; a pass there says very little about the set beneath it.

So what monitoring does a gravity cycle need?

A biological indicator on the schedule the jurisdiction requires, chemical indicators in the packs, and a physical parameter record for each cycle. In Ontario the expectation for dental sterilizers is a biological indicator for each sterilizer on every day it is used, not a weekly test; the weekly baseline is the US CDC and AAMI ST79 position.

A Bowie-Dick type test does not apply here, because there is no vacuum stage to check. Clinics that run both cycle types should make that distinction explicit in the log rather than leaving a reader to infer it.

How should the result be recorded?

Record the integrator result against the load, with its position noted. A result without a position cannot be interpreted later, and position is exactly the variable a gravity cycle is sensitive to.

Keep the cards or a photograph with the log where the format allows. The log generator prints a position column for this reason, and filling it in takes seconds at the time and cannot be reconstructed later.

Related reading

sterilization monitoring collection; what each indicator proves; what tape records must show; wholesale and multi-site ordering Reselling? become a distributor.

Take the paperwork with you: print the sterilization log sheet, pre-fill a week of cycles with the log generator, or ask a sterilization compliance specialist to review how the records line up with the load sheets.

Frequently Asked Questions

Can I use Class 5 integrators in a gravity displacement sterilizer?

Yes. They respond to time, temperature and steam contact, which a gravity cycle delivers.

Do I still need a biological indicator?

Yes. A chemical integrator cannot show that spores were killed, which is the question the biological indicator answers.

Does a Bowie-Dick test apply to a gravity cycle?

No. The test checks air removal through a vacuum stage, and a gravity cycle does not have one.