C-Reactive Protein Tests in Canada vs the US: Which Class Applies?

Graduated plastic specimen container with a screw cap on a dark background

Quick facts

  • A C-reactive protein (CRP) test measures an acute-phase protein that rises with inflammation, infection and tissue injury.
  • In the United States it files under 21 CFR 866.5270, C-reactive protein immunological test system, at Class II (performance standards).
  • The openFDA classification API lists the CRP assay family under codes that include NQD (Cardiac C-Reactive Protein, Antigen, Antiserum, And Control).
  • In Canada a CRP kit is an in vitro diagnostic device, and Part 2 of Schedule 1 classifies an IVDD used for diagnosis or patient management at Class II unless a listed life-threatening, genetic or staging category applies.
  • An assay is an IVD device in Canada even when the instrument that runs it is a general-purpose analyser - the reagent carries the intended use.
  • Entries, classes and product codes below were checked against the eCFR, the Regulations and the openFDA classification database on 2026-10-12.

C-reactive protein is one of the oldest inflammation markers a laboratory still runs every day. It rises within hours of an infection or an inflammatory flare, falls as treatment takes hold, and is cheap enough to sit on a routine panel. Laboratories run it as a standard sensitivity assay, as a high-sensitivity assay for cardiovascular risk, and increasingly on a near-patient platform. All of those are one device family in class terms - a Class II IVDD in both countries - but the route to that class is different. This article sets out the US entry, the Canadian rules and the quality standards behind the result. The graduated specimen containers on the same bench are ordered with the same consumables list.

What is a C-reactive protein test system?

A CRP test is a quantitative immunoassay for C-reactive protein in serum or plasma. The protein itself is produced by the liver and climbs in response to interleukin-6 during inflammation, so the result is used as a non-specific marker - it says inflammation is present, not where it comes from. Laboratories use it to follow an infection's response to treatment, to monitor inflammatory conditions, and in its high-sensitivity form to estimate cardiovascular risk.

The kit is the regulated article. A CRP test system is a reagent pack plus calibrators and controls, sold for a named analyser or as a manual method. The class attaches to that kit and to its stated claim; the chemistry analyser it sits on is a separate, general-purpose device. That split matters when a laboratory buys a new platform and re-validates a menu against it - the kit's class and claim do not move just because the analyser changed.

Graduated plastic specimen container with a screw cap on a dark background

Is a CRP test system Class II in the United States?

Yes. 21 CFR 866.5270 is titled C-reactive protein immunological test system and the entry sits at Class II (performance standards). The openFDA device classification API lists the CRP assay family under product codes that include NQD, Cardiac C-Reactive Protein, Antigen, Antiserum, And Control. A device with performance standards reaches the market under general and special controls, which for an assay means the analytical performance claims it is allowed to make.

Performance standards are the useful part for a laboratory buyer, because they point at the data a supplier should be able to show: precision, linearity, interference and the traceability of the calibrator value. The class is one number; the evidence behind it is what decides whether the assay fits a given clinical use.

Item Detail
US regulation 21 CFR 866.5270 - C-reactive protein immunological test system
US class Class II (performance standards)
Product code NQD (Cardiac C-Reactive Protein, Antigen, Antiserum, And Control)
Device type In vitro diagnostic reagent system
Route to market 510(k) with performance data; registration and listing

How does Canada classify a CRP test?

Canada treats a CRP kit as an in vitro diagnostic device and applies Part 2 of Schedule 1 of the Medical Devices Regulations. The first question is whether the assay detects a transmissible agent - CRP does not, so Part 2 Rules 1 and 2 are not engaged. The assay is used in diagnosis and to follow patients, so Rule 4 applies: an IVDD not subject to Rules 1 to 3 that is intended to be used in diagnosis or patient management is Class II, unless it falls into a listed category such as cancer screening, genetic testing or disease staging.

A routine CRP assay is none of those, so the class is Class II - the same number as the US entry, reached by a completely different route. The practical consequence is that a Canadian distributor of a CRP kit holds a Class II device licence under section 32 with an ISO 13485 quality certificate behind it, whereas the US side points to a 510(k).

Item Canada United States
Named entry None; classified by rule 866.5270, Class II
Class Class II (Part 2, Rule 4) Class II (performance standards)
Governing text Schedule 1, Part 2, Rule 4 21 CFR 866.5270
Route to market Device licence s.32 with ISO 13485 510(k) and registration
Transmissible-agent rules Not engaged; CRP is not a pathogen marker Not applicable

Which CRP formats share the same class?

Standard-sensitivity CRP, high-sensitivity CRP, a manual latex method and a near-patient cartridge are all immunological CRP test systems, so they sit in the same class. What changes between them is the measuring range and the setting - a laboratory analyser, a small clinic reader or a hand-held device - not the class.

Format Typical use Class, Canada Class, US
Standard CRP, analyser Infection and inflammation follow-up II (Part 2, Rule 4) II (866.5270)
High-sensitivity CRP Cardiovascular risk estimation II (Part 2, Rule 4) II (866.5270)
Manual latex CRP Low-volume laboratory II (Part 2, Rule 4) II (866.5270)
Near-patient CRP cartridge Clinic-side decision support II (Part 2, Rule 4) II (866.5270)

The one detail worth carrying into a validation plan is the claim. A high-sensitivity method has a lower measuring range and a different clinical interpretation from a standard method, and a laboratory that swaps one for the other has changed the assay, not just the part number.

Pair of blue nitrile examination gloves laid out flat for a clinic tray

What does a Class II IVDD require of the manufacturer?

A Class II IVD device needs a device licence under section 32 of the Regulations in Canada, and the application is built on ISO 13485 quality management and ISO 14971 risk management. In the United States the equivalent gate is a 510(k) with the performance data the regulation's standards imply. On both sides the calibrator value has to be traceable, which is where ISO 17511 sits.

Traceability is the part end users feel. When a laboratory changes a reagent lot and the controls drift, the manufacturer's traceability chain is what explains the shift. A laboratory accredited to ISO 15189 is expected to hold that thread and to document lot-to-lot verification, not just run the control.

Which standards sit behind a CRP result?

Three layers. The clinical-laboratory layer is ISO 15189, the quality and competence standard a medical laboratory is accredited against. The method layer is the CLSI family - the Clinical and Laboratory Standards Institute publishes the evaluation protocols for precision, linearity, interference and comparison that a laboratory applies to a new assay. The traceability layer is ISO 17511 for the assigned value on the calibrator.

None of these changes the class, and that is the point. Class II is a low bar in regulatory terms; the quality of the result is carried by the standards the laboratory chooses to work to, not by the class number on the box.

What should the order record say?

Record the assay name and its claim (standard or high-sensitivity), the platform it is validated on, the measuring range, the calibrator and control lot numbers, and the storage temperature. Add the class reference - 21 CFR 866.5270 in the United States, Part 2 Rule 4 of Schedule 1 in Canada - so a reviewer can see why the item needed a Class II licence and nothing higher.

A laboratory orders reagents and bench consumables on one requisition. Clinics that stock for a whole practice can order through the wholesale programme and request a bulk quote for a standing or annual order; the cotton rolls, 2 medium sits in the same catalogue. Reselling? become a distributor.

Sources

  1. openFDA device classification API
  2. Medical Devices Regulations, SOR/98-282, full text and Schedule 1
  3. US FDA, overview of device regulation
  4. US FDA, medical device databases
  5. Health Canada, medical devices
  6. Health Canada, establishment licensing
  7. Clinical and Laboratory Standards Institute
  8. Public Health Ontario, infection prevention and control
  9. ISO 15189:2022, medical laboratories — requirements for quality and competence
  10. ISO 17511:2020, metrological traceability of values assigned to calibrators and control materials
  11. US CDC, laboratory quality
  12. World Health Organization
  13. ISO 13485:2016, medical devices quality management
  14. ISO 14971:2019, risk management of medical devices

Printable record sheet. A laboratory bench accumulates its own compliance paper: reagent lots, control charts, temperature logs. Open the free sterilization log sheet library — the downloadable pack holds a steam-cycle log, a reprocessing-room checklist and an equipment-maintenance sheet. No account and no sign-up; print as many copies as your autoclave log needs.

Log generator. Because a quality log for reagents and instruments is the same discipline as a cycle log. Run the sterilization log generator to build a dated, facility-labelled cycle sheet you can print or file straight into your quality manual.

Ask a compliance specialist. If your laboratory validates a new assay and wants the record to line up with the standards an assessor will ask for. Send your question — you get a written answer specific to your equipment and province, with the regulation or standard it is based on cited.

Related reading

CliniEco Medical supplies eye care, laboratory and dental consumables of the kind described in this article. MDEL #35334. This article is written for optometry, laboratory and dental procurement professionals and is not legal advice; the class of a device and the records a practice keeps belong to the manufacturer and to the requirements that apply to it.

Frequently Asked Questions

What does a CRP test measure?

It measures C-reactive protein, an acute-phase protein that rises with inflammation, infection and tissue injury. It is a non-specific marker - it shows inflammation is present, not its source.

What class is a CRP test system in the United States?

Class II (performance standards), under 21 CFR 866.5270, C-reactive protein immunological test system. Product codes include NQD.

How is a CRP kit classified in Canada?

As an in vitro diagnostic device at Class II, under Part 2 Rule 4 of Schedule 1 of the Medical Devices Regulations - an IVDD used in diagnosis or patient management is Class II unless a listed higher-risk category applies.

Does the analyser change the class of the assay?

No. The kit carries the class and the claim; the chemistry analyser is a separate general-purpose device. Re-validating the assay on a new platform does not change the regulatory class of the kit.

What else is this test called?

CRP, hs-CRP, high-sensitivity CRP, acute-phase protein and C-reactive protein immunoassay all refer to the same family of measurement.

Which standard governs the calibrator value?

ISO 17511:2020 sets the requirements for metrological traceability of the values assigned to calibrators and control materials used in a laboratory assay.