- An automated FISH enumeration system is a scanning microscope, an image-analysis system and software that finds and counts fluorescent signals in interphase nuclei of tissue sections.
- In the United States it is a named entry. 21 CFR 866.4700, Automated fluorescence in situ hybridization (FISH) enumeration system, is Class II (special controls), product code NTH read from the openFDA classification database.
- Canada keeps no named entry. A system intended to examine specimens in vitro to aid a diagnosis is an in vitro diagnostic device, and Part 2 Rule 4 of Schedule 1 places it at Class II.
- Part 2 Rule 4(a) lifts an IVDD to Class III where it is intended to be used in screening for or in the diagnosis of cancer - the provision that matters most for this family.
- Part 2 Rule 7 provides that where an IVDD is intended to be used with another IVDD, the class of both is the class of the higher-risk device, so the probe kit and the scanner cannot be classified in isolation from each other.
- Sections, classes and product codes below were checked against the eCFR, the Regulations and the openFDA classification database on 2026-10-16.
An automated FISH enumeration system finds fluorescent signals in tissue sections and counts them, which is the kind of work that used to be done by eye at a microscope. The United States names the system and keeps it at Class II with special controls; Canada derives the class and lifts it to Class III as soon as the stated use reaches cancer. This article sets out both routes and the records behind the images, with the 120 mL sterile specimen container that starts the workflow.
What is an automated FISH enumeration system?
It is a complete workflow in a box. A fluorescence microscope with a motorised stage scans a slide, a camera captures images at several wavelengths, and software locates the nuclei, identifies the fluorescent probes by colour, size and shape, and counts the signals. The output is a number per cell, and a trained operator confirms it before it is reported.
The device is a hybrid. Physically it is a microscope and a camera; functionally it is an in vitro diagnostic instrument, because it is bought to examine patient specimens and to aid a diagnosis. Both the US entry and the Canadian regime focus on that functional description rather than on the optics, which is why a scanner sold for research use and the same scanner sold for diagnostic use are two different regulatory objects.
The sample matters as much as the machine. Formalin-fixed, paraffin-embedded tissue has to be sectioned, mounted, hybridised and counterstained before the scanner sees it, and each of those steps introduces variation. The US entry names formalin-fixed, paraffin-embedded human tissue specimens explicitly, which tells the buyer that the validated workflow stops there.
What class is an automated FISH enumeration system in the United States?
It is a named entry. 21 CFR 866.4700 describes a device consisting of an automated scanning microscope, an image-analysis system and customised software applications for FISH assays, intended for in vitro diagnostic use as an aid in the detection, counting and classification of cells and in the detection and enumeration of FISH signals in interphase nuclei of formalin-fixed, paraffin-embedded human tissue specimens. It is set at Class II (special controls), with FDA guidance on FISH enumeration systems as the special control.
The openFDA device classification API returns product code NTH against that section, and OYU for a DNA probe kit for a human chromosome on the same regulation number. The neighbouring immunofluorescence family sits at the same class: 21 CFR 866.4750 covers an automated indirect immunofluorescence microscope and software-assisted system at Class II with product code PIV.
| Item | Detail |
|---|---|
| US regulation | 21 CFR 866.4700 - Automated FISH enumeration system |
| US class | Class II (special controls) |
| Product code | NTH (system); OYU (DNA probe kit on the same section) |
| Stated intended use | In vitro diagnostic aid in detecting, counting and classifying cells and enumerating FISH signals |
| Specimen named in the entry | Formalin-fixed, paraffin-embedded human tissue |
| Operator requirement | A trained operator confirms the results generated by the device |
| Neighbouring section | 21 CFR 866.4750 - automated indirect immunofluorescence microscope, Class II, product code PIV |
| Read on | eCFR identification text and openFDA classification, both read 2026-10-16 |
How does Canada classify an automated FISH enumeration system?
Canada keeps no named entry, so the class is derived. The system examines specimens in vitro to aid a diagnosis, which makes it an in vitro diagnostic device, and the class then comes from Part 2 of Schedule 1 of the Medical Devices Regulations.
Part 2 Rule 4 places an IVDD intended to be used in diagnosis or patient management at Class II, unless one of six categories applies. The first of those categories is the one this family meets most often: \\"it is intended to be used in screening for or in the diagnosis of cancer.\\" Where the stated use is cancer, the class is Class III, one class above the US entry, and the route to market is a device licence rather than general controls with a special control document.
Part 2 Rule 7 then closes the loop on the rest of the order. If the scanner is intended to be used with another IVDD - a probe kit, a reagent, a control - the class of both is the class of the higher-risk device. A laboratory that buys the Class III probe and the Class II scanner on separate purchase orders has still bought one system in the eyes of the rule.
| Item | Canada | United States |
|---|---|---|
| Named entry | None; classified by rule | 866.4700 - Automated FISH enumeration system |
| Class, general diagnostic use | Class II (Part 2 Rule 4) | Class II (special controls) |
| Class, cancer use | Class III (Part 2 Rule 4(a)) | Class II |
| Internal control or paired device | Part 2 Rule 7 takes the higher class of the two | Not provided for in the entry |
| Near-patient use | Part 2 Rule 6 places a near patient IVDD at Class III | Not provided for in the entry |
| Licence route | Device licence for the applicable class | Class II with the special control guidance document |
Why does cancer use raise the Canadian class?
Because a wrong result in cancer work changes treatment in a way that cannot be undone. Part 2 Rule 4 separates the six categories that carry that consequence - cancer screening and diagnosis, genetic testing, fetal congenital disorder screening, a risk of death or severe disability from an erroneous result, disease staging, and drug monitoring where a wrong result leads to an imminent life-threatening decision - and assigns them Class III. A microscopy system that only assists a morphological description sits at Class II; the same hardware with an intended use that names cancer sits at Class III.
That is why the intended-use statement is the document a laboratory manager should read before the specification sheet. The two can look almost identical, and they carry different classes in Canada. The US route does not make the same split; the entry keeps the class and regulates the residual risk through the special control guidance, which is why a team that plans its licences from the US database will underestimate the Canadian requirement.
What should a laboratory check before ordering?
Four checks. First, confirm the intended-use statement and whether it names cancer screening, cancer diagnosis or disease staging, because Part 2 Rule 4(a) and 4(e) turn on those words. Second, confirm the specimen types the maker validated, since the US entry names formalin-fixed paraffin-embedded tissue and a scan of anything else is a scan outside the stated use. Third, confirm the probe kits the system is intended to be used with and their classes, because Part 2 Rule 7 applies the higher class to both. Fourth, confirm the operator confirmation step, since the software suggests results and the trained operator confirms them; a workflow without that step does not match the device.
| Check | What to ask for | Why it matters |
|---|---|---|
| Intended-use statement | Label text naming or excluding cancer and staging | Determines Class II or Class III in Canada |
| Validated specimen types | The specimen list the maker verified | Outside the stated use there is no validated claim |
| Paired devices | Probe kits and reagents and their classes | Part 2 Rule 7 applies the higher class to both |
| Operator step | Who confirms the suggested result and how | The device is an aid, not the final report |
| Image records | Retention period and format | Images are the evidence behind the count |
| Consumables | Specimen containers, cassettes and slides | A mislabelled block is a wrong result |
Slides and blocks travel in the same containers whether the workflow ends at a microscope or at a scanner. The 120 mL sterile screw-cap container and the 15 mL sterile centrifuge tubes are the same catalogue items the histology bench already orders.
A histology and molecular laboratory also reorders the consumables used upstream of the scanner. Clinics that stock for a whole practice can order through the wholesale programme and request a bulk quote for a standing or annual order; the 120 mL sterile screw-cap specimen container sits in the same catalogue. Consumables for this line are listed under diagnostic laboratories. Reselling? become a distributor.
Sources
- openFDA device classification API
- Medical Devices Regulations, SOR/98-282, full text and Schedule 1
- US FDA, medical device databases
- US FDA, overview of device regulation
- Health Canada, medical devices
- Health Canada, establishment licensing
- ISO 15189:2022, medical laboratories, requirements for quality and competence
- ISO 17511:2020, metrological traceability of values assigned to calibrators
- ISO 13485:2016, medical devices quality management
- ISO 14971:2019, application of risk management to medical devices
- ISO 11137-1:2006, radiation sterilization of health care products
- ISO 20776-1:2019, susceptibility testing of infectious agents
- Clinical and Laboratory Standards Institute
- World Health Organization
- US CDC, laboratory quality
- Public Health Ontario, infection prevention and control
- Standards Council of Canada
Printable record sheet. Scanning systems generate more images in a week than a laboratory can hold in a drawer, so the retention rule has to be written down before the first scan. collect the imaging and retention record sheets — the downloadable pack contains a scan record, an image-retention schedule and a reviewer sign-off line, printable on plain paper with no account and no sign-up
Log generator. Choose the system, the specimen type and the province, and the page produces a dated scan record with the confirmation step built in. build a dated scan record sheet — every generated sheet carries the block number, the specimen type and the reviewer line, so two technologists document the same scan the same way
Ask a compliance specialist. A classification question usually turns on how the device is used, not only on its name. Ask a compliance specialist — you get a written answer specific to your equipment and province, with the regulation or standard it is based on cited.
Related reading
- Whole slide imaging systems: class in Canada and the US
- Automated tissue processors: which class applies
- Flow cytometers in Canada vs the US: which class
CliniEco Medical supplies eye care, laboratory and dental consumables of the kind described in this article. MDEL #35334. This article is written for optometry, laboratory and dental procurement professionals and is not legal advice; the class of a device and the records a practice keeps belong to the manufacturer and to the requirements that apply to it.
Frequently Asked Questions
What class is an automated FISH enumeration system in the United States?
It is 21 CFR 866.4700 at Class II (special controls) with product code NTH. The special control is FDA guidance on automated FISH enumeration systems.
What class is an automated FISH enumeration system in Canada?
As an in vitro diagnostic device used in diagnosis it is Class II under Part 2 Rule 4 of Schedule 1, and it rises to Class III where the intended use is screening for or diagnosis of cancer.
Does the probe kit change the class of the scanner?
It can. Part 2 Rule 7 provides that where an IVDD, including its analyzers and reagents, is intended to be used with another IVDD, the class of both is the class of the higher-risk device.
What specimens does the US entry cover?
The identification names interphase nuclei of formalin-fixed, paraffin-embedded human tissue specimens, and the system is intended for in vitro diagnostic use as an aid in detecting, counting and classifying cells.
What else is an automated FISH enumeration system called?
FISH scanner, scanning fluorescence microscope, automated image analysis system, and enumeration workstation. The class follows the intended-use statement, not the trade name.
Who confirms the result the software produces?
A trained operator. The US identification states that a trained operator must confirm results generated with the device, which is why the sign-off step belongs in the workflow records.