Autoclave & Ultrasonic Equipment

The sterilization room is three jobs on one bench: cleaning an instrument before anything is packed, sealing the pack, and proving afterwards that the cycle it went through actually worked. This page covers the units that do those three jobs - the ultrasonic tank, the constant-heat sealer, the incubator and the indicator reader - plus the consumables each one depends on. Packaging sits under sterilization pouches and the tests sit under autoclave monitoring and spore testing. The sterilizer itself is not sold here: the Class B benchtop unit on our site is a pre-launch listing, not a purchasable model, so nothing on this page assumes you are buying an autoclave from us.

10 products
More about Autoclave & Ultrasonic Equipment

Short answer: a single-surgery dental clinic, veterinary practice or med-spa running one autoclave needs a 6 L ultrasonic tank that takes a full instrument cassette, a 12 mm constant-heat pouch sealer, and a dry-block incubator for 24-hour spore tests. Add a 4-well fluorescence reader only if the daily test is a 20-minute, 1-hour or 3-hour rapid-readout indicator. Cycle volume decides the size, not the catalogue.

How many cycles a week does the room actually run?

Count it before ordering anything, because the count sets the tank, the block and the number of spare wells. The arithmetic that survives an inspection is simple: cycles per day, multiplied by days the autoclave runs, gives the number of packs that have to be sealed and the number of indicator wells that get filled.

Take a four-chair dental clinic that runs the autoclave four times a day, five days a week. That is twenty cycles and twenty packs to seal, and at one biological indicator per sterilizer per monitoring day it is twenty wells filled in a week. Run six days and the figure climbs to twenty-four. A 20-well dry block - the HB105-S1 dry-bath incubator - is then full with no room left for the re-test after a failed indicator, which is the seat you only need on the week you need it badly. Two sterilizers, or a second suite in the same building, is where a block with spare capacity stops being optional.

The cleaning side counts differently. A 6 L ultrasonic tank holds one full instrument cassette per cycle, because the tank floor is 31.1 by 17.2 cm - the footprint a standard cassette needs. If your instruments arrive loose instead of in a cassette, you are either unpacking them onto the basket or running two cycles, and that is the point at which a second tank or a longer working day becomes the cheaper decision. The CS6.2 6 L dual-frequency cleaner is sized around that single-cassette cycle: 32 kHz for heavier soil, 48 kHz to reach lumens and threaded joints, with a 120-second degas at the start of the run so the first load is not fighting a head of air.

Which configuration fits a one-chair practice, a four-chair clinic, or a building with two sterilizers?

The table below is the same decision, stated three ways. Read across the row that matches your cycle count rather than the row that matches your ambition - the units are not upgradable after the fact, and buying a size up usually means a tank that sits half loaded.

Room profile Cycles per day Cleaning Sealing Incubation and readout Why this row
Single surgery (solo dental, vet, med-spa, tattoo studio) 1-4 6 L tank, multi-enzyme concentrate at 1:400 12 mm constant-heat sealer, or self-sealing pouches if every pack fits a stocked size 20-well dry block plus 24-hour spore tests One cassette per cycle; one indicator well per operating day leaves seats to spare
Multi-chair clinic (4-8 chairs, one autoclave) 6-10 Same 6 L tank run back to back; solution bought by the larger pack Sealer in continuous operation after a 2-3 minute pre-heat 24-hour block with USB data export, or a second 20-well block Wells run out before the tank does; export replaces a hand-copied temperature
Building with two sterilizers, or a suite per floor 10+ One tank per reprocessing room, not one tank shared between floors Second sealer only if both rooms pack separately 4-well fluorescence reader if the daily test is rapid-readout BIs are counted per unit per day, so the block and the reader both double

What are the specifications, compatibility limits and shipping format of each unit?

Every figure below is the one an order turns on: the seal width that decides whether a pack peels cleanly, the well diameter that decides whether an ampoule seats flush, and whether the unit leaves as a single bench machine or a case pack. Consumables are separate lines throughout - the tank does not ship with solution, the reader does not ship with ampoules.

Unit Key specification Compatibility that decides the order How it ships
CS6.2 ultrasonic cleaner 6 L tank, 31.1 x 17.2 x 11.6 cm; dual frequency 32/48 kHz; two ceramic heaters; digital timer; 120-second degas One full instrument cassette per cycle; certification by Intertek to CSA C22.2 No. 68 and UL 73 (cETL) Single bench unit with basket, part SB01; replacement baskets ordered separately
Multi-enzyme cleaning solution Protease, lipase, amylase and cellulase; low-foam; 1:400 to 1:600; 2-10 minutes at 20-50 degrees C A 6 L tank takes 15 mL at 1:400 (10 mL at 1:600); 24 months sealed, 90 days after opening 1 L bottle or 5 L pail; one 5 L pail makes 2,000 to 3,000 litres of working solution
Constant-heat pouch sealer 200 degrees C thermostatic head; 12 mm seal width; 300 mm maximum seal length; 100 W, 110 V / 60 Hz; IPX0 indoor use Pouches and rolls up to 300 mm wide, three rolls side by side; 2-3 minute pre-heat then continuous operation One stainless-steel bench unit with built-in cutter; no consumables included
HB105-S1 dry-bath incubator 20 wells (5 x 4) at 11 mm; room temperature +5 to 105 degrees C; +/-0.5 degrees C control; 115 V, 110 W Self-contained vials of 8 mm body and 10 mm cap; cTUVus to UL 61010-1 and CAN/CSA-C22.2 No. 61010-1-12 Single unit, 3.2 kg with block fitted; display only, no data export
24-hour incubator with USB export 56-60 degrees C held to +/-0.5 degrees C; 8.8 mm machined wells; LCD with USB data export; 100-240 V adapter Ampoules read at 56-60 degrees C, the growth band for Geobacillus stearothermophilus One bench unit; no case pack exists, and a two-year warranty applies
4-well fluorescence reader Four wells, model TR-R30MP4, held at 56-60 degrees C; reports positive or negative fluorescence; USB result storage Reads 20-minute, 1-hour and 3-hour fluorescence ampoules; none are included in the box Instrument only; ampoules reordered as separate 50-packs

Do you need a pouch sealer if you already buy self-sealing pouches?

Self-sealing pouches close on a pressure-sensitive strip: no machine, no pre-heat, one fixed size per pack, and nothing to warm up at the start of the day. A sealer covers the two cases a pre-made pouch cannot. Packs longer or wider than any stocked size, and rolls cut to the length the pack actually needs - the 76 mm by 50 m sterilization roll is the usual starting point, and a 30 cm cut gives roughly 166 pouches per roll. If every pack you make already fits the Class 4 self-sealing pouch you buy, the sealer adds a step and removes none.

The seam is the part that matters. A constant-heat head holds one temperature for the whole run instead of spiking per impulse, and the 12 mm band is the full width of the heating element, which is what lets the pack peel open at the point of use without the film tearing away from the paper. That is a packaging decision, not a sterilization one: the seal keeps the contents sterile until opening, and the indicator inside the pack is what reports whether the cycle reached it.

What has to be in place before the units go into service?

Three things, and none of them is optional. First, the electrical side: the cleaner is certified to CSA C22.2 No. 68 and UL 73, the dry block to CAN/CSA-C22.2 No. 61010-1-12, and all of it runs on 110-115 V at 60 Hz in a Canadian clinic. Second, the water and the bench: the tank needs a low-foam enzyme at the dilution above, a rinse step after the cycle, and an indoor IPX0 position that will not be splashed. Third, the monitoring routine, which has to start with the first cycle rather than the first inspection.

In Ontario, the dental regulator's infection-control standard expects a biological indicator on every day a sterilizer runs, a chemical indicator in every pack, and a written record of both. The practical version of that routine is a 24-hour self-contained spore test read by colour change, a Class 5 integrator (ISO 11140-1 Type 5) in each pack, Bowie-Dick packs (ISO 11140-5 Type 2) for the daily vacuum check on a pre-vacuum unit, and Class 1 indicator tape on the outside. Practices that read rapid results instead use the 20-minute or 3-hour fluorescence ampoules in the reader above.

What none of that does is certify a load as sterile. An indicator reports the conditions inside the pack it travelled in; it is evidence about the cycle, not a statement about every instrument in the room. The record is what turns the evidence into something an inspector can check, and the field list for it is at the compliance log centre, printable and free to use.

Batch, compatibility and ordering notes

  • Consumables never ship with instruments: no ampoules with the reader, no solution with the tank. Order both lines or nothing runs on day one.
  • Well diameter is a specification, not a detail. An 11 mm block takes an 8 mm vial body with a 10 mm cap; a wider cap will not seat, and a loose tube with an air gap around it is the usual reason one indicator in a batch reads late.
  • The incubator and the reader leave as single bench units. There is no case pack of either, and no quantity discount that changes the fact that one reprocessing bench needs one of each.
  • Solution has a clock: 24 months sealed, 90 days once opened, so a larger pack only saves money if the volume justifies it inside that window.
  • Equipment ships under CliniEco Medical's MDEL #35334 as a licensed importer and distributor; documentation, manuals and certificates are supplied as documents on request.
  • French-language manuals are available on request; the units themselves are marked in English and use North American plugs.

Where to go next

Clinics ordering a full room - tank, sealer, block, reader and the consumables behind them - can open an account with volume terms on the clinic and institutional ordering page. If the question is not which unit but which monitoring interval applies to your sterilizers, that answer belongs with a specialist: send it through the sterilization compliance consultation and it comes back in writing, with the standard behind it. Packaging and the tests themselves are one click away in the two collections linked above.