Codyson CS4.0 ultrasonic cleaner, three-quarter view with lid open

CliniEco Dental Ultrasonic Cleaner — 4 L Benchtop, CS4.0 Dual-Frequency 32/48 kHz

SKU CLI-ULTRA-CLEAN-40 · CliniEco Medical
$449.00
In stock · ships from Toronto, ON
CLI-ULTRA-CLEAN-40sku
Sterilization & Cleaning Equipmentcategory
CliniEco Medicalbrand
In stockavailability
Pack sizeSingle configuration Single configuration
Health Canada licensed importer & distributor · MDEL #35334 · Ships from Toronto
Free shipping over $99 in Ontario · $199 rest of Canada

Specifications

SKUCLI-ULTRA-CLEAN-40
BrandCliniEco Medical
CategorySterilization & Cleaning Equipment
AvailabilityIn stock
Ships fromToronto, ON

Pricing

OrderUnit price
1 case$449.00 CAD
2–4 casesRequest quote
5+ casesRequest quote
Case pricing shown. Clinic and volume pricing is quoted per order — most quotes go out within one business day.
Read the full product guide 5 sections · specs, use, compliance

Quick answer: A 4 L tank with both frequencies on one dial and a five-step timer — enough for the loose instruments from one chair, and not enough for a full cassette.

Most benchtop ultrasonic cleaners sold online are built for jewellery and eyewear. They are small, they are cheap, and the timer is a single knob you cannot repeat. The CS4.0 is the same platform as our 6 L unit in a smaller footprint: a 4,000 ml stainless tank, 32 kHz and 48 kHz on the same generator, controlled heating with a temperature read-out in both °C and °F, a drain valve, and a timer with five fixed steps so the same load gets the same cycle next week.

Scope of use. This is a cleaning device. It removes soil — cement residue, polishing paste, dried blood in hinges and threads — as a step in instrument reprocessing. It makes no microbicidal claim, and it does not replace any later step in the workflow set out in the instrument manufacturer’s instructions for use. The Health Canada guidance on reprocessing information is why the instrument manufacturer’s instructions — not the cleaner’s specification sheet — decide what happens to a given instrument after it comes out of the tank.

What are the CS4.0 specifications?

The numbers below are the ones that decide whether the unit fits a room: how much liquid you have to mix, how long the cycle runs, and what the electrical certificate actually names.

CS4.0 — 4 L benchtop ultrasonic cleaner, specifications as supplied
Field Specification
Tank capacity 4 L (4,000 ml) stainless steel, with drainage valve
Frequency Dual — 32 kHz for heavier soil, 48 kHz for fine crevices and hinged joints
Timer Five fixed steps: 5 / 7.5 / 10 / 12.5 / 15 minutes
Program memory Remembers the last timer and temperature setting between cycles
Heating Controlled heating with a temperature display in °C and °F
Housing Moisture-resistant PCB in a hybrid aluminium and plastic housing
Supply AC 100–120 V, 50/60 Hz, North American plug
Electrical certification cETL — Intertek report 220112139GZU-001, to UL 73 and CSA C22.2 No. 68
Included Unit, one-year manufacturer warranty; stainless basket available as an accessory
List price CAD $449.00

Which tank size in the range fits your room?

The size break in this range is not a matter of taste. A standard dental instrument cassette does not fit a 4 L tank, and it does fit a 6 L tank. If your day is one chair’s hand instruments, the 4 L is the whole requirement; if a cassette goes in the tank, you are shopping one row down.

Model range as listed — tank, frequency and the load each size is built for
Model Tank Frequency Built for List price (CAD)
CS4.0 4 L 32 / 48 kHz Single chair, loose instruments $449.00
CD-4862 6 L 35 kHz Full cassette, five temperature presets $649.00
CS6.2 6 L 32 / 48 kHz Full cassette, dual frequency $699.00
CD-4875 7.5 L 35 kHz Larger trays, five temperature presets $849.00
CS8.0 8 L 49 kHz Two chairs into one cycle $999.00
CS20 20 L 35 kHz Several chairs or batch work $1,199.00

Two practical checks before you order. First, measure the longest instrument you routinely clean — a 4 L tank sets the limit sooner than most people expect. Second, count how many times a day you would run the tank: a single load in the morning and a single load at close is the pattern this size is built for, and three or four loads is the pattern that pushes a practice into the 6 L class.

Six litre benchtop ultrasonic cleaner, straight-on view of the unit with the lid closed

How do you run a repeatable cleaning cycle on the CS4.0?

The value of this unit in an inspection is that the cycle can be described in one line and repeated tomorrow by someone else. That means fixed numbers, not judgement calls.

Cycle set-up step by step
Step Setting Why it matters
1. Fill Working level with solution at the label dilution (1:400 to 1:600 for our enzyme concentrate) Under-filling loads the transducer unevenly; over-filling spills into the housing seam
2. Heat Set the temperature on the front panel and wait for the display to reach it The same set-point every day is what makes the log entry defensible
3. Load Instruments in the basket or a tray, never flat on the tank floor Metal on metal blocks cavitation underneath the instrument
4. Run Choose 5, 7.5, 10, 12.5 or 15 minutes — the unit remembers the last setting A fixed step is repeatable; a hand-held knob is not
5. Drain and dry Empty through the drain valve, dry the chamber, record the cycle Residue left overnight is the usual cause of a cloudy tank

Two habits make the difference. Keep the basket in — instruments resting on the tank floor leave shadows in the cavitation field and wear the stainless. And change solution on a schedule rather than when it looks dirty: the log is the evidence, and a tank that is topped up instead of changed cannot be reconstructed afterwards.

What an infection-control inspection asks for is the record, not the machine. Ontario’s provincial reprocessing guidance asks for documentation and maintenance of records for each component of the cleaning, disinfection and/or sterilization process, based on the manufacturer’s recommendations, and repeats that records must be retained according to the facility’s policy (Public Health Ontario, “Best Practices for Cleaning, Disinfection and Sterilization of Medical Equipment/Devices In All Health Care Settings”, 3rd edition, checked 2026-11-07). A unit with a drain valve, a visible timer and a stable temperature set-point is what makes that record short enough to write at the end of a shift.

What does the cETL mark cover, and what does it not?

The CS4.0 carries a cETL mark from Intertek under report 220112139GZU-001. The mark on the unit is an ETL mark issued by Intertek, tested against UL 73 and CSA C22.2 No. 68. It is an electrical safety certificate. It is not a UL listing and it is not CSA certification, and it says nothing about how well the tank cleans. Canadian reprocessing practice sits on a separate standard, CSA Z314:23, whose scope puts documentation inside the reprocessing quality system rather than inside the appliance.

What the electrical certificate covers, and what it does not
Covers Does not cover
Electrical safety of the appliance, assessed to UL 73 and CSA C22.2 No. 68 Cleaning performance on your instruments, on your water hardness, or at your solution concentration
The specific unit family named on Intertek report 220112139GZU-001 The condition of an individual unit after a drop, a repair, or a modification
Compliance of the supply, plug and cord assembly for the North American market Any clinical outcome, or any claim that the unit removes or inactivates micro-organisms
The identity of the certificate holder and the factory it was issued to Your own reprocessing records, which are yours to keep and produce

What is in the box, and what should be ordered with it?

One cleaner, 4 L, 110 V with a North American plug, plus the manufacturer warranty. The stainless basket is an accessory rather than a line item, so say so on the order if you want it in the same box. Nothing else is required to run a cycle except solution and water — no plumbing connection, no drain pump.

Multi-enzyme ultrasonic cleaning solution is what goes in the tank, sold as a 1 L bottle or a 5 L pail and diluted 1:400 to 1:600. The 6 L CS6.2 is the unit to look at if your load grows into full cassettes. Read how to choose a tank size and what to record on the cleaning log before you commit to a size.

Related reading

CliniEco Medical is a licensed medical device establishment (MDEL #35334). We are the Canadian importer of record for this unit: duties and brokerage are handled on our side, and you receive it from our Toronto warehouse with a one-year manufacturer warranty supported in Canada. Questions on which size fits your room go to our sterilization desk.

Ships in 1 week. Ships within one week of payment; we confirm the date with you before you pay.

Common questions

Will the CS4.0 take a full instrument cassette?

No. A standard dental cassette needs the 6 L class, which is why the 6 L CS6.2 exists in the same range. The 4 L tank takes the loose hand instruments from one chair, a small tray, or a load of mirrors and probes.

Does the CS4.0 replace any later step in reprocessing?

No. It is a cleaning device: it removes soil, and it makes no microbicidal claim. Everything the instrument manufacturer's instructions for use require after cleaning still applies, and the provincial requirements for the rest of the workflow are unchanged.

Which frequency should be used, 32 kHz or 48 kHz?

32 kHz drives larger bubbles and is the setting for heavier soil such as cement residue. 48 kHz drives smaller bubbles into threads, hinges and fine crevices. Because both are on the same generator, one tank covers a mixed load instead of forcing a choice at purchase.

What does the cETL mark on this unit mean?

It is an ETL mark issued by Intertek, tested against UL 73 and CSA C22.2 No. 68, under report 220112139GZU-001. It certifies the electrical safety of the appliance. It is not a UL listing, it is not CSA certification, and it makes no statement about cleaning performance.

What power supply does the CS4.0 need?

AC 100–120 V, 50/60 Hz with a North American plug, which is what a standard clinical circuit provides. The unit ships with the plug fitted; no adapter is needed.