Preventing cross-contamination with 96-well PCR plates in a water and wastewater laboratory

Aluminium PCR sealing films fanned out on a white background

Preventing cross-contamination with 96-well PCR plates in a water and wastewater laboratory

A water laboratory runs a 96-well plate for coliform confirmation, and one well in the middle of the run comes up positive beside a no-template control. The instrument did nothing wrong. Something moved.

Plate testing concentrates that risk: 96 wells on 9.0 mm centres inside a 127.76 by 85.48 mm footprint, all open at once during setup. Control starts before the plate is unwrapped.

Where does a 96-well plate carry contamination?

Four places, and only one of them is the plate. The closure is first: a film lifted and re-pressed, or caps set down on the bench, transfers whatever was on the bench into the well. The second is the plate edge, where a gloved fingertip lands when the plate is carried. The third is the tip, since an unfiltered tip draws aerosol into the barrel and the next dispense puts it in another well. The fourth is the stack, when a plate that has been through the cycler is set down beside a fresh one.

Our filtered 200 and 300 microlitre pipette tips are racked and sterile for this step, because the filter is the only part of the setup standing between the sample and the pipette barrel. Eluates that wait for a repeat run go into 2.0 mL screw-cap microtubes, and dilutions sit in 0.5 mL ClearSeal microtubes with o-rings.

Full-skirted 96-well PCR plate with a 8 by 12 well grid

Which plate and closure combination keeps wells apart?

Format Closure that suits it Contamination risk Where it is used
Full-skirted 96-well Film or foil seal Low if the seal is single-use and pressed flat Instruments that need a rigid plate
Semi-skirted 96-well Film seal Medium; the plate flexes when carried Thermal cyclers with a plate adapter
Non-skirted 96-well Strip caps or film Higher; no frame to hold Short runs, manual handling

A skirted plate is easier to keep clean, because the frame gives you somewhere to hold that is not the well surface. The trade-off is volume per well.

What does a Canadian water laboratory have to be able to show?

Aluminium PCR sealing films fanned out on a white background

The standard for the analyte, and the traceability of the batch behind the result. In Ontario, Schedule 1 of O. Reg. 169/03 sets Escherichia coli and total coliforms at not detectable, so an amplicon in the wrong well is not a rerun. It is a reported result that has to be corrected before the certificate leaves the building.

That is why the plate record matters as much as the plate. ISO/IEC 17025:2017 requires technical records that let a result be repeated, which in practice means plate lot, seal lot, extraction batch and control positions on one page.

How should the bench be set up to keep wells independent?

Work in one direction: extraction, setup, amplification, reading, and never back. Unwrap a plate only when the master mix is ready, seal it as soon as the last well is loaded, and label it before the first dispense rather than after.

Then write the lot numbers down while the plate is still on the bench. A plate that has been read cannot be identified from outside, and a bag discarded at the end of the run cannot be identified at all.

Water laboratories that also run membrane filtration keep hold times on the same sheet. Bench consumables for a municipal or provincial laboratory sit in the diagnostic laboratory supply line, and the wider plastics range is in laboratory consumables.

Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.

References

  1. Ontario Regulation 169/03, Ontario Drinking Water Quality Standards, Schedule 1 (checked 25 November 2026)
  2. ISO/IEC 17025:2017, general requirements for the competence of testing and calibration laboratories (checked 25 November 2026)
Printable companion: the laboratory log sheet library includes a plate run sheet with plate lot, seal lot, extraction batch and control well positions. Converting samples per week into plates per quarter? The clinic supply planner does the arithmetic, including the wells lost to controls.

Related Reading

Frequently Asked Questions

Do 96-well PCR plates arrive sterile?

Not by default. Sterility is a separate claim printed on the bag, and PCR grade covers the reaction, not freedom from DNA.

Should a plate be sealed or capped?

Both are in use. A pressure-sensitive film gives a flat optical surface for reading and a clean edge for the instrument; strip caps suit plates that are opened repeatedly. What matters is that the closure is single-use and sits flat across the whole row.

How do you stop amplicon reaching the next well?

Physically separate the setup bench from the amplification and reading area, change gloves between them, and use filtered tips so aerosol does not travel back into the tip. Plate geometry does not prevent carry-over; bench movement does.

Does a positive well next to a negative control invalidate the batch?

It does until the cause is established. A drinking-water result is a recorded measurement, so the run is held and the affected wells re-extracted before the report is issued.

Last updated: November 2026. CliniEco Medical is a licensed medical device establishment (MDEL #35334).

Vous travaillez déjà avec des cliniques ? Devenez représentant régional.