A carton of filtered tips arrives with sealed racks inside a bag, a lot number on the outside, and a certificate that nobody reads until a run fails. The failure it prevents is quiet: a tip that sits a fraction off the barrel draws air, under-delivers, and produces a result that looks like a chemistry problem.

Two questions carry the dock check: does this rack fit the pipettor it will be used on, and can the sterility claim be traced?

What do you check on a box of filtered tips before you accept it?

Open one rack, not the whole case. Seat a tip on the pipettor, press and release the ejector, and watch three things: whether the tip seats square, whether it wobbles when the barrel is tapped, and whether the filter sits below the collar rather than flush with it. Then check the rack itself for cracked wells and the bag for a lost seal. A rack of sterile filtered 200 and 300 microlitre tips is inexpensive enough that a rejected rack is not an argument, but the lot behind it is.

Check What a failure looks like Disposition
Tip seating Tip tilts or drops when the barrel is tapped Hold the rack, test a second
Filter position Filter flush with the collar or pushed in Quarantine the rack
Ejector fit Tip will not release, or releases with a jump Note the pipettor model on the record
Rack wells Cracked or missing well dividers Accept the rack, reject the case count
Bag seal and indicator Bag open, or indicator absent Quarantine and raise with the supplier
Lot and certificate Lot illegible, or certificate missing Hold until the document arrives

A rack of filtered pipette tips with the filter visible inside each tip collar

Does a filter make the pipette sterile?

No, and the distinction matters at the bench. The filter is an aerosol barrier: it stops droplets and aerosols travelling between the sample and the barrel, which protects the next sample and the pipettor. Sterility is a separate property of the tip and rack, established when the manufacturer validated a radiation dose under ISO 11137-2 and released the lot to an assurance level.

That leaves one honest rule for the dock: physical checks cover fit, and documents cover sterility.

How do you test tip fit without contaminating a rack?

Wide-bore filtered pipette tips standing in a rack beside a single tip held upright

Take the test tip from the rack that will be used, do the seat-and-wobble check dry, then discard it. If a functional check is needed, run it with water on a balance at the volume the method uses, and record the pipettor, the tip lot and the result together.

A 200 microlitre wide-bore filtered tip and its 1000 microlitre version seat on the same barrel but draw through a larger opening, so a fit test that passes on a standard tip has to be repeated on the wide-bore rack.

What has to be on the certificate for a sterile rack?

Three things: the lot, the sterilization method and dose, and the assurance level the lot was released against. ISO 8655-2 belongs on the pipette file rather than the tip file, because the standard describes piston-operated pipettes and the tip they are specified with, not tips as measuring instruments.

Everything else is arithmetic: rack count against case count, lot against carton, expiry against shelf. If a diagnostic laboratory runs accredited work, the receiving record and the certificate belong in the supplier file, and the containment side is covered in the Canadian Biosafety Standard wherever a rack holds material that has to be inventoried.

Racks, reservoirs and transfer formats for a working bench sit in laboratory consumables, and the intake route for a testing service starts on the laboratory supply line.

Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.

References

  1. ISO 8655-2:2022, piston-operated volumetric apparatus, Part 2: pipettes (checked 26 November 2026)
  2. ISO 11137-2:2012, sterilization of health care products, radiation, Part 2: establishing the sterilization dose (checked 26 November 2026)
  3. Canadian Biosafety Standard, third edition (checked 26 November 2026)
Printable companion: the laboratory receiving and quarantine log library has a tip-rack acceptance sheet with fit, filter and lot columns. Matching a rack to a pipettor? The clinic supply planner sets out rack formats and reorder points per bench.

Related Reading

Frequently Asked Questions

Does a filtered tip make the pipette sterile?

No. The filter is an aerosol barrier that keeps the barrel and the sample apart. Sterility is a property of the tip and the rack, set by the dose the manufacturer released the lot against.

What does a tip label citing ISO 8655 actually state?

It states fit with a pipette built to that standard. ISO 8655-2 covers the pipette and its specified tip as a system, not the tip as a stand-alone measuring device.

How many tips should be tested for fit?

One from each rack opened, plus one from the first rack of each lot. Fit failures cluster by mould, so one confirmed failure is a lot conversation.