How to Operate Ultrasonic Cleaners: Daily Start-Up and Shut-Down Checks
A six-chair practice in Mississauga runs its first ultrasonic cycle at 7:50 and seats the first patient at 8:20. The tank is full, the timer reads eight minutes, the heater is set to 40 degrees, and the cassette comes out with a grey film still sitting in the hinges. Nothing on the machine is broken. What went wrong happened in the four minutes before the cassette went in.
Four things were missed, and they are the same four things that go missing in most clinics: the solution was mixed and used immediately, so the water was still shedding dissolved air; the cassette was set down flat on the tank floor rather than in the basket; the heater had not finished climbing when the timer started; and the concentrate had been measured by eye, which put it at roughly double the label dilution. Any one of those will cost cleaning performance. All four together produce instruments that look clean and fail a residue test.
The unit we stock for this bench is the CliniEco CS6.2 ultrasonic dental cleaner — 6 L benchtop, dual frequency 32/48 kHz, SB01 basket included. Two features on it exist because of the routine below: a 120-second degas cycle, and a second frequency for fine work. Neither feature helps if the machine is switched on and loaded in the same minute.
What does a morning start-up check need to cover?
Five checks, each under two minutes once the bench is set up, and each tied to a failure it prevents.
- **Volume.** Fill to the working level, not to the rim. An overfilled tank spills into the heater housing when a cassette displaces water, and an underfilled tank lets the upper part of the load run dry.
- **Dilution.** Measure the concentrate with the graduations on the jug or the pump on the container. The low-foam multi-enzyme ultrasonic concentrate — 4 L, 1:400 to 1:600 dilution is mixed by that ratio, and mixing by eye is the single most common cause of residue on instruments.
- **Temperature.** Give the heater time to reach the set point before the first load. A timer that starts on cold solution is timing a cycle the load never receives.
- **Degas.** Run the degas program, or let the unit circulate unloaded until the solution settles. This is the step that gets skipped when the first patient is already in the chair.
- **Basket and drain.** Basket in, drain valve closed, unit level. A drain valve left ajar on a 20 L unit empties the tank into the cabinet below during the cycle.
The instrument side matters as much as the machine side. ISO 17664-1:2021 requires a device manufacturer to supply validated reprocessing instructions, and Health Canada's guidance for manufacturers of reusable devices follows the same principle: the instructions supplied with the instrument govern. If they specify manual cleaning only, an ultrasonic cycle is not a permitted substitute, however well the tank is set up.
| Start-up step | What to check | Pass condition | What to record |
|---|---|---|---|
| Fill | Working level, not rim | Basket submerged, no overflow when loaded | Tank volume and mix date |
| Mix | Concentrate measured, not estimated | Dilution matches the label ratio | Product and dilution |
| Heat | Heater at set point before loading | Reading matches the set temperature | Start temperature |
| Degas | Program run or unloaded circulation | No foam layer, solution clear | Degas time if fitted |
| Load | Instruments in the basket, hinges open | Nothing touching the tank floor | Cycle count |
| Close | Lid on, timer set for the load | Cycle started after the checks above | Start and finish time |
Why does a freshly mixed tank need a degassing cycle?
Tap water holds dissolved gas, and mixing concentrate into it entrains more. Under sonication that gas comes out of solution, and while it is coming out it cushions the collapse of the cavitation bubbles that actually remove soil. Research on ultrasonic degassing in Ultrasonics Sonochemistry shows the rate dissolved gas is driven off depends on frequency and power, which is why two units on the same timer can behave differently on a fresh tank.
The signs are easy to read. A tank that has not been degassed sounds dull, throws a fine mist of bubbles onto the lid, and leaves instruments with a haze that wipes off onto a cloth — soil loosened but never lifted clear. Ten minutes later the same tank cleans properly with no change to the settings.

On the CS6.2 the degas programme runs for 120 seconds, which takes the guesswork out of the morning. On a unit without one, run it unloaded while it heats and load nothing until the surface stops releasing bubbles. Degas again after a refill with cold water, and after a full drain and refill, because both put gas back into solution.
Which frequency, temperature and time should a load run at?
Frequency is the setting most clinics set once and never touch, and it decides what the tank can reach. Lower frequency makes larger bubbles with more energy at the surface, suiting heavier soil and solid instruments; higher frequency makes smaller bubbles that penetrate narrow gaps, suiting hinges and fine work.
| Load | Frequency | Solution temperature | Time in tank | Notes |
|---|---|---|---|---|
| Orthodontic pliers, forceps, extraction instruments | 32 kHz | 40 to 50 degrees Celsius | 8 to 10 minutes | Hinges open, instruments separated in the basket |
| Cassettes of hand instruments with textured grips | 48 kHz | 40 to 55 degrees Celsius | 6 to 8 minutes | Better reach into knurling and ratchets |
| Mixed general cassettes on a single-frequency unit | 49 kHz | 40 to 55 degrees Celsius | 8 to 10 minutes | Typical of the larger 8 L bench units |
| Heavily soiled first load of the day | 32 kHz | At the detergent label temperature | 10 minutes | Change solution before the next load if foam persists |
Temperature sits inside the detergent's stated band rather than at whatever the heater can reach. Warm solution works faster; above the label range protein can set onto the instrument surface, and below it the enzyme works too slowly for a six-minute cycle. The CS6.2 heater covers 40 to 60 degrees Celsius, and a bench running more than one cassette per cycle usually moves up to the 8 L CS8.0 benchtop unit — 49 kHz with heater rather than running the same load twice.
Time is the weakest lever on the machine: doubling the cycle does not fix a tank loaded cold, dosed heavy or carrying a second cassette. Fix the setup, then adjust the timer.
How do you confirm the tank is actually cavitating?
A benchtop cleaner can heat, count down and move the water without delivering cavitation, because the heater, timer and generator board are separate circuits. A failed transducer or a delaminated bonding pad leaves a machine that looks normal in operation.
Two checks separate the machine that is working from the machine that is merely on. The first is sound: a working tank gives a sharp hiss with a fine buzz through the tank wall, not a low hum or a coarse shake. The second is the foil test — suspend a strip of kitchen foil in the basket, run the tank for one to three minutes and look at it. Cavitation pits and perforates the foil where it is working, and the pattern should be even across the strip. A 2026 study in Molecules used a modified aluminium foil test to quantify cavitation dose and showed how strongly it shifts with the operating parameters.
Run a foil test after any service visit, after the unit is moved, when a load comes out worse than usual, and as part of the monthly inspection. A bright, unmarked patch marks a zone where the load is not being cleaned — exactly the failure a cycle count cannot detect.

What should never go into an ultrasonic tank?
The list of things that should stay out of a stainless tank is short and expensive to learn the hard way.
- **Soft metals.** Aluminium, anodised aluminium, carbon steel and plated items can darken, pit or lose their coating. Instrument trays made of anodised aluminium are the usual casualty.
- **Handpieces and turbines.** Follow the manufacturer's instructions; many specify manual cleaning and a lubrication step a tank cycle does not replace.
- **Optics and cemented assemblies.** Cavitation works on the cement line as readily as on soil.
- **Chlorine bleach and flammable solvents.** Bleach attacks stainless steel and produces fumes at working temperature; solvents in an open heated tank are an exposure risk with no upside.
- **Loose sharps.** Needles, scalpel blades and burs go into a sharps container or a covered basket handled with tongs. Bloodborne pathogen rules apply to the decontamination sink exactly as they do at the chair.
- **Anything that touches the tank floor, or a second cassette stacked on the first.** Both shield the load from cavitation and both show up as a clean-looking instrument with soil left in the box joint.
What does an end-of-day shut-down look like?
The evening routine is shorter than the morning one, and it protects the machine rather than the load.
| Shut-down step | What to do | Why it matters |
|---|---|---|
| Remove the load | Basket out, instruments rinsed and dried | Residue left to dry on instruments is harder to remove than residue rinsed while wet |
| Drain the basket | Invert it over the sink, check for trapped items | A bur left in the basket turns up in the next load |
| Tank | Wipe the rim and the inside walls above the solution line | Splash above the line dries to a mineral film that stains and holds soil |
| Solution | Drain if the reuse period is up, or if the unit will sit for a weekend | Standing warm solution grows biofilm and etches the tank walls |
| Lid | Leave it off or ajar on a drained tank | Condensation under a closed lid on an empty tank keeps the interior damp |
| Power | Switch off at the wall, not just at the panel | A unit left in standby still holds the heater circuit live |
Draining is where clinics differ, and the detergent's stated reuse period should decide it rather than habit. A tank drained every evening runs clean but costs more concentrate; a tank on a three-day change needs a mid-week check for film on the walls. Either routine works if it is written down and followed, which is the part most benches are missing.
What should the daily log hold, and why does it carry weight in Canada?
CSA Z314:23 covers medical device reprocessing in all health care settings, not hospitals alone, and documentation is part of what the standard expects a reprocessing program to hold. Public Health Ontario's clinical office guidance adds that equipment used to clean, disinfect or sterilize — ultrasonic washers included — must meet standards established by Health Canada, the Public Health Agency of Canada, CSA and provincial guidance. Neither document prescribes a form. Both make "where is that written down" a fair question in an inspection.
The record does not need to be elaborate. Date and shift, operator initials, solution mix date and dilution, temperature at the start of the first cycle, number of cycles run, any foam or film observed, whether the cavitation check was performed and what it showed, and what was done about anything that failed. Six lines a day, written at the bench, in the sheet the practice already keeps for the sterilizer.
What that record buys is a chain. The CDC guideline is explicit that cleaning must precede all disinfection and sterilization, and that visible soil left in place interferes with microbial inactivation later. An ultrasonic cycle nobody recorded cannot show the cleaning step ran on the day a load failed. A line in a log can.
Ordering for several chairs, a second site or a shared reprocessing room? Wholesale and multi-site ordering covers case quantities and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.
Running ultrasonic cleaning in a dental practice? Dental practice sterilization and hygiene supplies collects the tanks, solutions and monitoring consumables that sit on the same bench.
Not sure how this applies to your own setup? Ask a sterilization compliance specialist — you will get a written answer specific to your equipment and province, with the standard or regulation cited.
References
- CSA Z314:23, Reprocessing of medical devices in all health care settings (checked 19 November 2026)
- Public Health Ontario, cleaning, disinfection and sterilization practices in all health care settings, 3rd edition (2013) (checked 19 November 2026)
- Public Health Ontario, Infection Prevention and Control for Clinical Office Practice (checked 19 November 2026)
- Public Health Ontario, Routine Practices and Additional Precautions reference sheet (checked 19 November 2026)
- CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities — Executive Summary (checked 19 November 2026)
- CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities (checked 19 November 2026)
- CDC, Guidelines for Infection Control in Dental Health-Care Settings, MMWR 2003;52(RR-17) (checked 19 November 2026)
- Royal College of Dental Surgeons of Ontario, Infection Prevention and Control (checked 19 November 2026)
- Health Canada, Guidance Document: Information to Be Provided by Manufacturers for the Reprocessing and Sterilization of Reusable Medical Devices (checked 19 November 2026)
- US FDA, Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling (checked 19 November 2026)
- ISO 17664-1:2021, Processing of health care products — Information to be provided by the medical device manufacturer for the processing of medical devices (checked 19 November 2026)
- ISO 17665:2024, Sterilization of health care products — Moist heat (checked 19 November 2026)
- British Columbia Ministry of Health, provincial guidance on cleaning, disinfection and sterilization of medical equipment and devices (checked 19 November 2026)
- US Occupational Safety and Health Administration, 29 CFR 1910.1030 Bloodborne Pathogens (checked 19 November 2026)
- Frequency and power dependence of ultrasonic degassing, Ultrasonics Sonochemistry, 2021 (checked 19 November 2026)
- Effect of Experimental Parameters on Cavitation Dose in Ultrasonic Baths via Modified Aluminum Foil Test, Molecules, 2026 (checked 19 November 2026)
- Reprocessing of dental instruments in washer-disinfectors: does a representative test soil exist in dentistry?, GMS Krankenhaushygiene interdisziplinär, 2012 (checked 19 November 2026)
Printable companion: the printable sterilization and reprocessing log sheets carry the columns described below, so the start-up and shut-down checks get written down at the bench instead of reconstructed from memory a month later.
Working out how much solution a bench actually draws in a year? The clinic supply planner converts cycles a day into concentrate and tank changes, so the order matches the workload rather than the shelf.
Related Reading
- Ultrasonic Bath Degassing: Why Fresh Solution Needs It and How Long It Takes
- What Does Dual-Frequency 32/48 kHz Change in Ultrasonic Cleaning?
- Ultrasonic Cleaner Basket Loading: Why Stacked Instruments Are Not Clean
- How Often Should an Ultrasonic Cleaner Be Drained and Cleaned?
- Setting up an ultrasonic cleaning bench in a Canadian dental clinic
Frequently Asked Questions
How long should an ultrasonic cleaner run for a cassette of dental instruments?
Five to ten minutes at the working temperature is the range most benchtop units are used at, but the timer is the least reliable variable on the machine. A short cycle in a tank that was never degassed, or that is loaded past the basket, cleans less than a longer cycle in a tank that was set up properly.
Do I have to degas a new tank of ultrasonic solution?
Yes, or the first load of the day is being cleaned in water that is still shedding dissolved air. Mixing solution entrains gas, and gas cushions the collapse of the cavitation bubbles that do the work. Units with a degas program run it for 120 seconds; units without one are usually run unloaded for a few minutes while they come up to temperature.
What temperature should an ultrasonic cleaner be set to for instruments?
The detergent label sets the band, and the heater on a benchtop unit usually covers 40 to 60 degrees Celsius. Running colder slows the chemistry, and running hotter than the label allows can fix protein onto the surface instead of lifting it. Check the label before turning the dial.
Can instruments sit directly on the bottom of the ultrasonic tank?
No. Instruments on the tank floor damp the transducer that is bonded underneath it, scratch the stainless, and shield their own undersides from cavitation. Everything goes in the basket, and the basket sits in the solution rather than on the tank bottom.
How do I know the ultrasonic cleaner is working if the water looks agitated?
Movement on the surface proves the pump, the sweep or the heater is running, not that cavitation is reaching the load. The check that answers the question is a foil test on a strip of kitchen foil, plus listening for the sharp hiss a working tank makes instead of a flat hum.
Why is there foam on my ultrasonic cleaning solution?
Foam usually means the concentrate was mixed too strong, the tank is overloaded, or protein is being pushed out of joints faster than the solution can hold it. Low-foam enzyme concentrates exist for exactly this reason; a tank that foams heavily loses cavitation energy to the foam layer.
Should the tank be drained at the end of every day?
Not every day, but standing solution has a life. Follow the detergent maker's stated reuse period, drain before a weekend or a shutdown, and rinse the tank before it dries. A tank left full and warm over a weekend comes back with a film on the walls and a smell that topping up will not fix.
Can dental handpieces go into an ultrasonic cleaner?
Only if the handpiece manufacturer's instructions say so, because that document governs how the device is cleaned and lubricated after cleaning. Many handpiece instructions specify manual cleaning and a lubrication step that a tank cycle does not replace, and provincial guidance expects the manufacturer's instructions to be followed.