Which Standards Apply to Medical Bowie-Dick Test Pack in the US vs Canada? Sizes and Sourcing
Which standards apply to a Bowie-Dick pack in Canada and the US?
The Bowie-Dick test answers a narrower question than most people assume: it shows whether air was removed from the chamber before steam arrived. That single purpose is why the pack is a Class 2 chemical indicator rather than a biological one, and why the standards that govern it are about the test's construction rather than about sterility. CliniEco lists the CliniEco Medical Bowie-Dick test pack — ISO 11140-5 Type 2, 6-pack, and the notes below are what a Canadian buyer checks against what a US buyer checks instead.
What standard is the pack rated to?
| Item | Canada | United States |
|---|---|---|
| Indicator class | Class 2 chemical indicator, ISO 11140-1 | Class 2 chemical indicator, ISO 11140-1 |
| Pack specification | ISO 11140-5 Type 2 designation | ISO 11140-5 Type 2 designation |
| Reprocessing standard cited | CSA Z314 | ANSI/AAMI ST79 |
| Frequency for pre-vacuum units | Every day the unit is used, per CSA Z314 and the manufacturer | At the start of the day, per AAMI ST79 |
| Failure action | Stop, retest, quarantine | Stop, retest, quarantine |
Two rows are identical because the product question is identical in both markets. The rows that differ are procedural, and they matter because a clinic audited against the wrong country's baseline will be told its record is either excessive or incomplete.

Which sterilizers need one, and which do not?
The test belongs to pre-vacuum sterilizers, where a vacuum stage pulls air out before steam enters. A gravity displacement unit has no such stage, so the pack cannot tell you anything about it, and running one there is a habit rather than a control. That distinction is the first thing to settle before ordering: count pre-vacuum units only, because the count sets the annual quantity and a mixed fleet is where over-ordering starts.
What record does each jurisdiction expect?
The date, the sterilizer identifier, the cycle number, the pack lot, the result, and the action taken if the result was a fail. Both jurisdictions expect the record to be producible on demand, and both expect the failed pack to be kept rather than discarded, since a failure that leaves no artefact cannot be investigated. Write the record at the machine, not at the end of the week.

What happens after a failed pack?
Stop releasing loads from that unit, run the test again after checking the door seal and the chamber for obvious faults, and if it fails a second time, take the sterilizer out of service and call service. The loads processed since the last passing test are the ones to quarantine, and the decision on whether to recall them belongs to the person named in the procedure, which is why the procedure should name one.
What sizes and pack counts are sold?
The consumer unit is a 6-pack and the purchasing unit is a case of ten 6-packs, so the arithmetic a buyer needs is packs per week multiplied by the weeks the unit is in service. A two-unit practice running every weekday needs about ten packs a week, which makes a case a quarter of supply rather than a month. Packs are dated, so buying a year at once is a shelf-life decision, not a saving.
Where does a clinic buy Bowie-Dick packs?
On the same account as the Class 5 steam chemical integrator, 100-pack that goes inside the load and the autoclave indicator tape, 3-roll set that goes outside it, so one order covers the release evidence for a cycle from three directions.
Running a dental practice or a hospital sterile processing department? The dental practice supplies page groups the monitoring lines by cycle type.
Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.
References
- ISO 11140-3:2007 (checked 18 October 2026)
- ANSI/AAMI ST79:2017/(R)2022 w/ AMDs A1:2020, A2:2020, A3:2020, A4:2020 (checked 18 October 2026)
Not sure how this applies to your own setup? Ask a sterilization compliance specialist — you’ll get a written answer specific to your equipment and province, with the regulation or standard cited.
Related Reading
- What Is a Bowie-Dick Test and Which Cycles Need It?
- ISO 11140-3 class 2 chemical indicators for Bowie-Dick tests: What Canadian Clinics Should Verify Before Ordering
Frequently Asked Questions
What standard is a Bowie-Dick test pack made to?
It is a Class 2 chemical indicator under ISO 11140-1, and the air-removal packs are covered by the ISO 11140 parts written for that test. The CliniEco pack is supplied to the ISO 11140-5 Type 2 designation, which is the specification to quote when a purchaser compares it to another pack.
Which sterilizers need a Bowie-Dick test?
Pre-vacuum and forced-air removal sterilizers, where the test proves the air was removed before steam entered. A gravity displacement unit has no air-removal cycle to verify, so the test does not apply and a pack bought for one is money spent on the wrong question.
Is the test daily in both countries?
For a pre-vacuum unit in routine use, yes. The Canadian expectation comes from CSA Z314 and the sterilizer manufacturer's instructions, and the US baseline comes from ANSI/AAMI ST79, which places the test at the start of the day before the first load. Both treat a failed pack as a stop.
How many packs does a clinic use?
One per pre-vacuum sterilizer per day it is used, plus repeats. A two-unit practice on a five-day week uses about ten packs a week, so a 6-pack is a few days of supply and a case of ten 6-packs is roughly a quarter.
Last updated: October 2026. CliniEco Medical is a licensed medical device establishment (MDEL #35334).