How to Inspect Biological Indicators on Receipt: A Receiving Checklist for Clinics
A spore test is the smallest purchase in the reprocessing room and the only one that can void a week of release decisions. Nothing on the box reveals a night on a loading dock at -20 °C, so the first honest look happens at the door.
The 24-hour self-contained biological indicator, 50 per box is the format this checklist covers: Geobacillus stearothermophilus spores in a sealed vial, with a glass ampoule of recovery medium crushed before incubation.
What should I check on the carton before I sign for it?
Five points, ninety seconds: a dry square carton, a count that matches the slip, lot and expiry on the inner box rather than the sleeve, nothing rattling, and somewhere dry for the stock to live.
| CliniEco 24-hour indicator pack | Accept when | Flag when | Recorded in |
|---|---|---|---|
| Outer carton | Dry, corners square | Damp, crushed tape | Slip, dated |
| Count | Matches slip, e.g. 50 | Short, or bag opened | Slip with lot |
| Lot and expiry | Legible on inner box | Sleeve only | Log, same day |
| Vials | Caps seated, no rattle | Cracked plastic, media visible | Quarantine note |
| Storage fit | Dry cupboard, room temperature | Only spot is a fridge | SOP location |
Two rows carry the weight. Lot and expiry decide whether a result three weeks out can be defended; storage fit decides whether the spores survive the wait. A frozen vial separates and will not grow, which reads on the log as a cycle failure that never happened.
How do I verify a new lot before it goes into service?

Incubate one unexposed indicator from the lot beside the first test vial and label it the control. That step is what makes a positive result on day one a sterilizer problem rather than a bad shipment. The CDC dental guidelines treat biological monitoring as the confirmation a cycle worked; the control is what makes it interpretable. Write the control beside the test result, on the same line, so a reviewer sees that the spores were alive and the load still passed.
What does a 24-hour format need that a rapid-readout format does not?
An incubator holding 56 to 60 °C for a full day — a dry-block incubator for 24-hour indicators covers a single-unit practice — and willingness to hold the load until the answer exists. Read by colour change in the recovery medium, it needs no reader warm-up and no software version; the trade is time.
Ontario expects one indicator per sterilizer on every day the unit is used. A clinic that releases at close loses nothing to a 24-hour format if the indicator rides with the first load; one that releases within the hour belongs on a 3-hour rapid readout indicator.
How should the indicators be stored?
In the closed original box, dry, 15 to 30 °C, away from the wall above the sterilizer. Never in a refrigerator: condensation inside a cap changes what grows. One box open at a time, oldest lot first, so no more than a week of vials sits loose on a shelf.

A practice running a steam unit and a dry-heat unit keeps separate cartons, because a shared tray invites the wrong vial into the wrong cycle.
What goes into the receiving record?
Five fields: date received, lot, expiry, quantity, and who checked. Add the storage location where there is more than one stockroom. Those five make a positive result traceable in minutes rather than reconstructed from memory, and they are what gets asked for when a lot is questioned.
Dental practices, walk-in clinics and care homes receive the same carton; what differs is the handling after the door.Ordering for a clinic, a lab or a care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.
TSA:
Printable companion: the spore test log sheet carries a receiving block, so the lot number is captured at the door.
TSB:
See where a lot sits in your own record with the sterilization self-check, which names the missing field.
STERILE: 1
References
- CDC — Guidelines for Infection Control in Dental Health-Care Settings, MMWR 52(RR-17) (checked 18 November 2026)
- Medical Devices Regulations (SOR/98-282), section 44 — establishment licence (checked 18 November 2026)
Not sure how this applies to your own setup? Ask a sterilization compliance specialist — you’ll get a written answer specific to your equipment and province, with the regulation or standard cited.
Related Reading
- How Do You Read a Biological Indicator Result Before Releasing a Load?
- Storage, Shelf Life and Lot Tracking of Biological Indicators
Frequently Asked Questions
Can I sign for a spore test delivery and open it a week later?
Open it the day it arrives. A crushed carton or a loose cap is a delivery problem the carrier still owns that week; later it becomes yours.
Do I incubate a control from every new lot?
Yes. An unexposed indicator from the lot goes in beside the test vial, so a positive reads as a failed cycle and not as a bad shipment.
What temperature does a 24-hour indicator need?
56 to 60 °C for the full 24 hours in a dry-block incubator. The ledge above the unit is not an incubator.
Last updated: November 2026. CliniEco Medical is a licensed medical device establishment (MDEL #35334).