Quick facts
- A whole slide imaging system creates, views and manages digital images of pathology slides as an aid to the pathologist reviewing surgical pathology material.
- In the United States it is a named entry. 21 CFR 864.3700, Whole slide imaging system, is Class II (special controls), with the product codes PSY, QKQ, PZZ and SIX.
- Canada keeps no named entry. An instrument that supplies energy to image a specimen is an active diagnostic device at Class II under Rule 10(1); used for diagnosis it is also an IVDD at Class II under Part 2 Rule 4.
- Part 2 Rule 7 raises both the scanner and its companion IVDD to the higher-risk class, so a workflow built from parts takes the class of the riskiest part.
- The US entry is framed as an aid to the pathologist, not as an autonomous reader; the intended use the maker states is what the class turns on.
- Entries, classes and product codes below were checked against the eCFR, the Regulations and the openFDA classification database on 2026-10-15.
A whole slide imaging system digitises a glass pathology slide and presents the result for review, interpretation and management. The United States names the system and places it in the special-controls class; Canada keeps no entry and reaches the same class by two different rules, one for active diagnostic imaging and one for in vitro diagnostic use. This article sets out both routes and the record a digital pathology service keeps, with the 90 mL graduated specimen cups used upstream of the scanner.
What is a whole slide imaging system?
It is the instrument and software that turn a physical slide into a digital image a pathologist can navigate at magnification. The scanner captures the slide, the viewing software assembles and presents the image, and the management software stores the case and tracks the images. Some systems add algorithms, such as an image artifact detector, that flag a scan for review.
The purpose named in the US identification is worth keeping in view: the system is intended as an aid to the pathologist to review and interpret digital images of surgical pathology slides, and it generates images that would otherwise be viewed by conventional light microscopy. The device supports a reading; it does not replace the reader.
What class is a whole slide imaging system in the United States?
It is a named entry. 21 CFR 864.3700 is titled Whole slide imaging system, and the identification reads: \"the whole slide imaging system is an automated digital slide creation, viewing, and management system intended as an aid to the pathologist to review and interpret digital images of surgical pathology slides.\" The regulation sets Class II (special controls), which means the general controls plus the special controls identified for the device.
The openFDA device classification API returns four product codes against the section, and they show how the system is broken up: PSY for the whole slide imaging system itself, QKQ for digital pathology image viewing and management software, PZZ for a digital pathology display and SIX for an image artifact detection software device. A purchaser reading only the scanner code will miss the software and display codes that apply to the same workflow.
| Item | Detail |
|---|---|
| US regulation | 21 CFR 864.3700 - Whole slide imaging system |
| US class | Class II (special controls) |
| Product codes | PSY (system), QKQ (viewing and management software), PZZ (display), SIX (artifact detection software) |
| Stated intended use | Aid to the pathologist to review and interpret digital images of surgical pathology slides |
| Reference in the entry | Images that would otherwise be viewed by conventional light microscopy |
| Premarket route | Class II route, with the special controls identified for the device |
| Read on | eCFR identification text and openFDA classification, both read 2026-10-15 |
How does Canada classify a whole slide imaging system?
Canada keeps no named entry, so the class comes from Schedule 1 of the Medical Devices Regulations. Two separate parts of the schedule reach the device. Under Part 1, Rule 10(1), an instrument that supplies energy to image or monitor a physiological process is an active diagnostic device: \"Subject to subrule (2), an active diagnostic device, including any dedicated software, that supplies energy for the purpose of imaging or monitoring physiological processes is classified as Class II.\" A scanner that illuminates a slide and captures an image supplies that energy, so Rule 10(1) applies.
Under Part 2, Rule 4, an in vitro diagnostic device used in diagnosis or patient management is classified as Class II: \"An IVDD that is not subject to rules 1 to 3 and that is intended to be used in diagnosis or patient management is classified as Class II, unless it falls into one of the following categories, in which case it is classified as Class III.\" The categories that lift an IVDD to Class III are cancer screening or diagnosis, genetic testing, screening for congenital disorders in the fetus, a risk that an erroneous diagnostic result would cause death or severe disability, disease staging, and monitoring of drug or substance levels where an erroneous result leads to an imminent life-threatening decision. A maker whose intended use reaches one of those categories has to answer for it.
Part 2, Rule 7 is the rule to read last, because it governs a system built from parts: \"In cases where an IVDD, including its analyzers, reagents and software, is intended to be used with another IVDD, the class of both IVDDs will be that of the IVDD in the class representing the higher risk.\" A scanner, its software and its display are therefore taken together at the class of the riskiest member of the set.
| Item | Canada | United States |
|---|---|---|
| Named entry | None; classified by rule | 864.3700 - Whole slide imaging system |
| Class | Class II (Part 1 Rule 10(1) as an active diagnostic device; Part 2 Rule 4 as an IVDD used in diagnosis) | Class II (special controls) |
| Governing text | Schedule 1, Part 1 Rule 10(1) and Part 2 Rules 4 and 7 | 21 CFR 864.3700 |
| Higher-class fork | Part 2 Rule 4(a)-(f) to Class III where the stated use reaches cancer screening, genetic testing, fetal screening, staging or drug-level monitoring | Special controls set for the device |
| System with parts | Part 2 Rule 7: scanner and software take the higher class of the set | Separate product codes for system, software and display |
| Licence route | Device licence for the Class II system | Class II premarket route |
Why do the two countries reach the same class here?
Because both systems treat digital pathology as a device that supports a diagnosis rather than as ordinary software. The United States wrote a named Class II entry with special controls; Canada wrote a general rule for active diagnostic imaging and a separate rule for in vitro diagnostic devices used in diagnosis, and a whole slide imaging system answers to both.
The routes still differ in an important way. The US entry fixes a class for the named system, and the software and display carry their own product codes. Canada applies two independent rules and a third rule that resolves a system of parts, so the class of a Canadian submission follows the riskiest member of the workflow rather than the name on the scanner.
What should a laboratory check before ordering one?
Four checks. First, confirm whether the quote covers the scanner alone or the full workflow of scanner, viewing software, management software and display, because the class attaches to the set. Second, read the intended-use statement carefully, since the Part 2 Rule 4 categories turn on whether the stated use reaches cancer screening, staging or drug-level monitoring. Third, confirm the image format and the display specification the maker states, because a viewer is only as good as the panel behind it. Fourth, confirm the validation and reconciliation routine the maker describes for the transition from glass to digital reading.
A histology and pathology laboratory also reorders the consumables used upstream of the scanner. Clinics that stock for a whole practice can order through the wholesale programme and request a bulk quote for a standing or annual order; the 90 mL graduated specimen cups sits in the same catalogue. Consumables for this line are listed under diagnostic laboratories. Reselling? become a distributor.
How does the class affect the laboratory record?
The class sets the market-entry path; the laboratory's working routine is set by its own quality system. ISO 15189 is the framework medical laboratories use to state requirements for quality and competence, and it is where the record of a digital reading sits: which slide, which scan, which reader and which result.
A digital workflow also adds its own record: the scan settings, the version of the viewing software and the disposition of any artifact flagged by the system. Keeping those beside the case record is what lets a later review reconstruct how an image was read, which is the same discipline the laboratory already applies to a glass slide.
Sources
- openFDA device classification API
- Medical Devices Regulations, SOR/98-282, full text and Schedule 1
- US FDA, medical device databases
- US FDA, overview of device regulation
- Health Canada, medical devices
- Health Canada, establishment licensing
- ISO 15189:2022, medical laboratories, requirements for quality and competence
- ISO 17511:2020, metrological traceability of values assigned to calibrators
- ISO 13485:2016, medical devices quality management
- ISO 14971:2019, application of risk management to medical devices
- Clinical and Laboratory Standards Institute
- World Health Organization
- US CDC, laboratory quality
- Public Health Ontario, infection prevention and control
- Standards Council of Canada
- CCOHS, OSH answers
A printable pathology accession sheet. The case identifier, the block and slide count, the scanner used and the reader fit on one printed page, so every digital case starts with a record that can be filed. Open the compliance log centre for the blank form - no account and no sign-up - and print a run for the whole bench.
Turn the accession book into dated sheets. Set the laboratory name, the scanner and the bench once, and the sterilization log generator lays out dated sheets you can print and file with the case records.
Ask a compliance specialist. A classification question usually turns on how the device is used, not only on its name. Ask a compliance specialist — you get a written answer specific to your equipment and province, with the regulation or standard it is based on cited.
Related reading
- Automated zone readers in Canada vs the US
- Cytocentrifuges in Canada vs the US
- Microtomes in Canada vs the US
CliniEco Medical supplies eye care, laboratory and dental consumables of the kind described in this article. MDEL #35334. This article is written for optometry, laboratory and dental procurement professionals and is not legal advice; the class of a device and the records a practice keeps belong to the manufacturer and to the requirements that apply to it.
Frequently Asked Questions
What class is a whole slide imaging system in the United States?
Class II (special controls). It is a named entry at 21 CFR 864.3700, with product codes for the system, the viewing and management software, the display and an artifact detection software device.
What class is it in Canada?
Class II. A scanner that supplies energy to image a specimen is an active diagnostic device under Part 1 Rule 10(1), and an in vitro diagnostic device used in diagnosis is Class II under Part 2 Rule 4.
When would a whole slide imaging system reach Class III in Canada?
Where the stated use falls into one of the Part 2 Rule 4 categories: cancer screening or diagnosis, genetic testing, screening for congenital disorders in the fetus, a risk that an erroneous result would cause death or severe disability, disease staging, or drug-level monitoring where an erroneous result leads to an imminent life-threatening decision.
How is a system made of a scanner, software and a display classified?
Under Part 2 Rule 7, where an IVDD is intended to be used with another IVDD, the class of both is the class of the higher-risk device. The workflow therefore takes the class of its riskiest member.
Does the US entry cover the software as well as the scanner?
The named entry covers the whole slide imaging system, and the classification database shows separate product codes for viewing and management software, a display and an artifact detection software device. All are read against the same section.
Does the class set how a digital slide is recorded?
No. The class sets the market-entry path. The record is set by the laboratory quality system, for which ISO 15189 is the usual framework, and covers the slide, the scan, the reader and the result.