What must a Canadian laboratory record when a specimen is rejected for testing?

Specimen transport bag printed 95kPa with biohazard label, peel-and-seal strip and closing instructions

What must a Canadian laboratory record when a specimen is rejected for testing?

A community laboratory in Mississauga opens 240 accessions on a Monday. By ten o'clock one EDTA tube comes back clotted and never reaches the analyser. The next twenty minutes decide whether that event is reconstructable six months later, when a clinician asks why a potassium was never reported.

The bag the tube arrived in belongs to that record. The UN3373 specimen transport bag with a rear document pouch keeps the requisition with the specimen, so nothing has to be opened to read the paperwork. When a specimen is rejected, that pairing is what lets the laboratory state which tube, from which patient, in what condition, arrived.

What has to be recorded when a specimen is rejected?

Four things, in this order: what arrived, why it was not tested, who was told, and where the container went. A log that carries only the first two answers an audit question but not a clinical one. A rejection is not a deleted result; it is a documented decision that no result will be produced, and it has to carry the same traceability as a report.

Which fields belong on the record?

Three clear specimen transport bags printed 95 kPa certified, SPECIMEN SAFETY BAG, with a four-step closing procedure

A reviewer reads it for completeness, not prose.

Field Why it is asked for
Accession number links the event to the request and to the transport document
Date and time received separates a collection failure from a transport delay
Specimen type and container shows whether the 120 mL screw-cap specimen container was the problem
Rejection reason code makes the log countable by cause rather than by anecdote
Who made the decision names the technologist or pathologist who released it
Clinician notified, how and when the one field the ordering office ever checks
Repeat requested, yes or no shows the loop was closed, not just logged
Disposition of the container retained, returned or discarded, with a date
Transport condition on arrival temperature, leakage, packaging breach
Linked nonconformity or corrective action ties the entry to the quality system

Does a rejected specimen still produce a report?

Not a result, but a notice. The ordering clinician needs to know that no result is coming, and the record has to show that the notice was sent. A cancelled accession with no notification is the version auditors write up: the chart looks identical to a test still running. The rejected specimen is not relabelled and re-entered; the record closes with a new collection request, not a corrected tube.

Specimen transport bag printed 95kPa with biohazard label, peel-and-seal strip and closing instructions

Why is a transport failure kept in a second file?

Because it is a different kind of event. Clinical rejection belongs to the laboratory quality system, where ISO 15189:2022 requires nonconforming work to be recorded and, where it can recur, acted on. Damage in transit belongs to the Transportation of Dangerous Goods Regulations, which govern how a Category B specimen is packaged and documented for movement. One file explains why a patient was re-drawn; the other explains why a package failed. Merging them loses both.

Where does the record stop being paperwork?

When someone reads it. A rejection log with a reason code per entry becomes a purchasing signal: if short fills and leaks cluster on one container size, or if every case of 90 mL graduated sample cups arrives with a loose lid, the fix is a different format, not a reminder to the collector.

Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.

Running a diagnostic laboratory? See our laboratory specimen handling and transport supplies.

References

  1. ISO 15189:2022, Medical laboratories — Requirements for quality and competence (checked 12 November 2026)
  2. Transportation of Dangerous Goods Regulations, SOR/2001-286 (consolidated text) (checked 12 November 2026)

Printable companion: the free specimen rejection log lists the same ten fields in column form, so the entry is written at the bench.

Turning rejection volume into a bag order? The clinic supply and cost planner converts accessions per day into packs per month at a stated resend rate.

Related Reading

Frequently Asked Questions

Can a mislabelled specimen be relabelled and then tested?

No. Relabelling breaks the link between container and collection and erases the evidence that a labelling failure happened. The record shows the specimen was rejected and a new collection was requested, with the reason code kept so the trend can be reviewed.

Who keeps the rejection record, the laboratory or the clinic?

The laboratory keeps it, because the laboratory made the decision not to test. The ordering clinic receives the notification and, where a repeat is needed, the collection request. Both sides hold a copy; only one holds the original.

How long must a rejection record be kept?

It is set in the laboratory quality manual and by its accreditation requirements; ISO 15189:2022 requires the retention period to be defined rather than left open.

Is a rejection a nonconformity?

Yes. Nonconforming work is recorded, evaluated and, where it can recur, followed by action. Most rejection logs stop at the first step, which is why the same short fill returns every quarter.

Last updated: November 2026. CliniEco Medical is a licensed medical device establishment (MDEL #35334).

Vous travaillez déjà avec des cliniques ? Devenez représentant régional.