Buying an ultrasonic cleaner is easy. The unit arrives, the tank fills, the timer counts down and the instruments come out visibly clean. The difficult question arrives later, when someone asks how the practice knows that a particular load in that particular tank was cleaned effectively, and what evidence exists if the answer is challenged.

That question turns a purchase into a validation file. This article sets out the three data items worth demanding from any supplier before the order is placed, and maps them onto the clauses a Canadian clinic is actually audited against. It is a companion to our specification comparison of units sold in Canada, not a repeat of it: here the subject is evidence, not the datasheet.

First, what an ultrasonic cleaner is not

An ultrasonic cleaner is a cleaning device. A washer-disinfector is a machine that cleans and then disinfects within a controlled cycle. The distinction matters because the standard most often invoked in supplier brochures, ISO 15883-1, is the general standard for washer-disinfectors: its scope covers performance requirements for cleaning and disinfection, the accessories needed to achieve that performance, and the methods for validation, routine control, monitoring and re-validation periodically and after essential repairs. Its 2024 edition replaced the 2006 edition and its 2014 amendment.

The companion part, ISO 15883-5, specifies the procedures and test methods used to demonstrate cleaning efficacy of washer-disinfectors and their accessories. Read together, the two parts describe a test regime built around a defined load, a defined test soil and a defined acceptance criterion. A supplier who says a benchtop ultrasonic unit is compliant with ISO 15883 has said something that needs two follow-up questions: which part, and which load was tested under which soil.

None of this makes an ultrasonic cleaner unusable, and none of it means a clinic needs a washer-disinfector instead. It means the claim has to be converted into a document a clinic can file. A unit such as a 4 L dual-frequency benchtop cleaner is a legitimate cleaning step; what makes it auditable is the evidence attached to it.

CliniEco CS6.2 ultrasonic cleaner, 6 litre benchtop unit with dual frequency display and temperature controls
A cycle record is only as good as the parameters the machine can display and hold: frequency, temperature and time.
CliniEco CS20 ultrasonic cleaner, 20 litre bench unit with heater, drain valve and digital control panel
A drain valve and a temperature setpoint turn a batch rinse into a documented step rather than a habit.

What do the six claims on the datasheet each prove?

Claim you will see What it can prove What it does not prove
Complies with ISO 15883-1 That a machine family was assessed against washer-disinfector requirements, if the certificate names the part and the tested configuration That your load, your water and your cycle were validated; a cleaning-only unit has no disinfection stage to validate against the thermal parts of the series
Dual frequency, 32 or 48 kHz Two cavitation regimes are available for different load types Which regime was used for the load in your record, or that cavitation reached the lumen of a hinged instrument
Degassing mode The machine can drive dissolved gas out of a fresh fill That degassing actually happened before the load went in; a default setting is not a record
Validated cleaning Nothing, until the test soil, the method, the load and the acceptance criterion are named Anything at all about your instruments
Cycle data export Time, temperature and frequency can be captured as machine data That the data reaches the clinic file, or that the values were inside limits for that load
Electrical certification The electrical build was tested to a recognised safety scheme Cleaning performance, which is a separate question from electrical safety

Data item one: efficacy evidence with a named soil and method

The first item is a cleaning efficacy report that names four things: the test soil, the method used to apply and detect it, the load configuration, and the acceptance criterion. A report that says the unit was tested and performed well, without those four, cannot be compared with anything and cannot be re-run after a repair.

This is also where the Canadian requirement becomes concrete. The review criteria used by Alberta Health Services for medical device reprocessing, which map each expectation to a clause of CSA Z314:23, record that the ultrasonic manufacturer's instructions should indicate devices or materials validated for ultrasonic cleaning, and should cover installation requirements including water quality, operating instructions, daily and routine and preventive maintenance, and daily sonication testing when in use. The same criteria state that foil testing is not acceptable, and require a commercially prepared test method instead. That single sentence retires the foil test that many clinics still run as their daily check.

Efficacy evidence and the daily test are different instruments. The efficacy report belongs to the supplier and is produced once, against a defined soil. The daily test belongs to the clinic and proves the machine was still working that morning. A supplier who offers only one of the two has offered half of the file.

Data item two: control of the parameters your record will claim

The second item is evidence that the parameters in the clinic's record are actually controlled by the machine or by the procedure. An ultrasonic cycle is not a single variable. Detergent concentration, water temperature, fill level, degassing, frequency and cycle time all move the result, and load geometry decides whether the sound field reaches the hinge of a box lock or the lumen of a cannulated item.

CSA Z314:23 addresses several of these directly as they are mapped in the Alberta criteria. Gross soil must be removed before immersion. The ultrasonic solution must be discarded at least daily and whenever visibly soiled, and a fresh solution must be degassed before use following the manufacturer's instructions. Devices must be rinsed following ultrasonic cleaning as the manufacturer's instructions require. Each of those clauses is a record field waiting to happen: if the clause defines the condition, the file should show the condition was met.

This is where the choice of unit becomes a documentation decision rather than a capacity decision. A 6 L dual-frequency unit with a drain and a temperature setpoint supports a record with named values; a 20 L bench unit with a drain valve supports batch records for cassette loads. The question to ask is not which tank is bigger, but which parameters the machine can display, hold and export so that the log entry is a reading rather than a recollection.

Data item three: change control after repair, and a plan for the day it stops

The third item is the least requested and the one an auditor is most likely to probe after an equipment fault: what happens to validation when the machine is repaired, and what the clinic does while it is out of service.

The Alberta criteria require preventative maintenance of automated equipment to be documented, with frequency following the manufacturer's instructions for use, and require maintenance and repair documentation, including any testing required after the work, to be available. They also require a written contingency plan for automated equipment shutdown, naming who is notified, how the fault is fixed, and what interim measures apply, with ultrasonic cleaners listed among the equipment in scope. The clause reference attached to that expectation is CSA Z314:23 18.1.1.

In practice the supplier's contribution is three documents: the preventive maintenance schedule with intervals, the post-repair verification procedure with its acceptance criterion, and a statement of which spare parts or service actions require re-verification rather than a routine check. Without the third, a clinic cannot tell a repair that resets the evidence from one that does not.

The clause map a Canadian clinic is audited against

CSA Z314:23 clause What it requires What goes in the clinic file
11.6.6.2 Gross soil removed before immersion in the ultrasonic cleaner Procedure step, and the pre-cleaning method for each instrument type
11.6.6.5 Solution discarded at least daily and when visibly soiled; fresh solution degassed before use Solution change log, concentration and batch, degas step and duration
11.6.6.6 Equipment monitored according to current standards and documented Daily test entry with the commercial test used, lot number and result
11.6.6 and Annex G Manufacturer instructions covering installation including water quality, operation, maintenance and daily sonication testing The current instructions for use, filed with the purchase documents
11.6.7.1 Devices rinsed after ultrasonic cleaning as the instructions require Rinse step and water quality used for the final rinse
18.1.1 Contingency plan for automated equipment shutdown Written plan naming notification, repair route and interim measures

Which questions produce a usable file?

Six questions, asked in writing before the order, will tell a buyer whether the supplier can support the file afterwards. Which part of ISO 15883 was the machine assessed against, and was the assessment done on a cleaning-only unit or on a machine with a disinfection stage? What test soil and method were used in the efficacy report, and can the report be re-run? Which parameters does the machine hold and display, and can they be exported as a record? What is the daily test method named in the instructions, and is it a commercial product? What re-verification follows a repair, and which repairs trigger it? What does the supplier provide for the contingency plan when the unit is out of service?

A supplier who answers all six in writing has effectively written the validation section of the clinic's procedure. A supplier who answers none has sold a tank. For a practice that wants the answers reviewed against its own workflow, the compliance question channel takes the unit model, the load and the current log, and the printable compliance log sheets cover the daily test, solution change and maintenance entries on one page. Multi-site groups that need the same unit across several locations can order through wholesale and multi-site ordering so that the instruction set matches across sites. Reselling? become a distributor.

Two related pieces go further on the operational side: our guide to tank size, degassing and solution change and the monthly inspection checklist for ultrasonic cleaners.

Where does a clinic take this next?

Four things a clinic can act on rather than read: the printable compliance log sheet, which records receiving dates, lot numbers and open-on dates and exports from the browser, and can log pickup dates and manifest numbers as well; the clinic supply planner, which turns daily counts of bags and sharps containers into a monthly case quantity; Ask your compliance question when a finding has to be documented for an inspector; and Wholesale and multi-site ordering for multi-site groups ordering bags and containers on one schedule.

Related reading

Frequently Asked Questions

Is an ultrasonic cleaner a washer-disinfector under ISO 15883?

No by default. ISO 15883-1 sets general requirements for washer-disinfectors, including cleaning and disinfection performance and the methods for validation and routine control. An ultrasonic cleaner that cleans only is a cleaning device, so a compliance claim should name the part of the series and state whether the machine has a disinfection stage to validate.

Is the foil test acceptable in Canada?

The Alberta Health Services review criteria for medical device reprocessing, which map expectations to CSA Z314:23, state that foil testing is not acceptable and require a commercially prepared test method specified by the manufacturer. A clinic using foil as its daily check should replace it with the test named in the instructions for use.

How often should the sonication test be run?

Daily when the unit is in use. The review criteria require the equipment to be monitored and documented, with daily testing using the test specified by the manufacturer and a commercially prepared product, and CSA Z314:23 clause 11.6.6.6 is the reference attached to that expectation.

Does a new solution have to be degassed?

Yes. CSA Z314:23 clause 11.6.6.5 requires the solution to be discarded at least daily and when visibly soiled, and a fresh solution to be degassed before use following the ultrasonic manufacturer's instructions. The degas step belongs in the log because the first cycle after a fresh fill will not clean to the same standard.

What changes after a repair?

The repair record has to show any testing required by the manufacturer after the work, and the clinic has to be able to show the machine returned to its documented performance. That is why the post-repair verification procedure and the list of repairs that trigger re-verification are worth obtaining at purchase rather than after a breakdown.

Can a supplier brochure replace a validation report?

No. A brochure describes a specification; a validation report names the test soil, method, load and acceptance criterion and can be re-run. Only the second can be filed against a cleaning step and compared after a repair or a change of detergent.

Sources

- ISO. ISO 15883-1:2024, Washer-disinfectors, Part 1: General requirements, terms and definitions and tests, scope. evs.ee catalogue record - ISO. ISO 15883-5:2021, Washer-disinfectors, Part 5: Performance requirements and test method criteria for demonstrating cleaning efficacy, scope. evs.ee catalogue record - Alberta Health Services. Medical Device Reprocessing of Reusable Critical and Semi-critical Medical Devices: Review Criteria and Supporting Standards, February 2025 revision, ultrasonic cleaning criteria and CSA Z314:23 clause mapping. albertahealthservices.ca - CSA Group. CAN/CSA-Z314:23, Canadian medical device reprocessing, clauses 11.6.6, 11.6.7 and 18.1 as mapped in the review criteria above. csagroup.org - AAMI. ANSI/AAMI ST79:2017/(R)2022 with amendments A1 to A4, cleaning and verification sections. aami.org