Sterility assurance level (SAL) 10-6 is a probability, not a promise. It is the regulatory definition of "sterile" for a terminally sterilized medical device: the probability of a single viable micro-organism surviving on one item after sterilization must be no greater than one in a million. Three numbers get mixed together in clinic training - SAL, log reduction and D-value - and they answer three different questions. This article sets out what each number measures, how they relate arithmetically, and how to check that a supplier's or a colleague's use of them is correct.

Quick facts
- SAL 10-6 is a design requirement for terminally sterilized devices, set out in EN 556-1, whose current edition is EN 556-1:2024 - not the 2001 edition that still appears in older training material.
- SAL is a probability applied to one item. Log reduction is a change in a population. D-value is a property of a spore population under a defined set of conditions. They are not synonyms and cannot be substituted for one another in a written procedure.
- ISO 11139:2018 defines SAL as "probability of a single viable micro-organism occurring in a product after sterilization." It is the vocabulary standard for the field, first published in 2018 and replacing ISO/TS 11139:2006.
- ISO 11138-1:2017 gives two normative methods for determining a D-value - Annex C (survivor curve) and Annex D (fraction-negative) - so a D-value quoted without a stated method and set of conditions is incomplete.
- Steam sterilization process requirements sit in ISO 17665:2024, which withdrew and replaced ISO 17665-1:2006 and consolidated the former Part 2 and Part 3 technical specifications into a single document.
What is sterility assurance level
Sterility assurance level is the probability of a single viable micro-organism occurring on an item after sterilization. The ISO definition is deliberately worded as a probability attached to a single item, not to a batch, a load, or a working day. SAL 10-6 therefore means: if one million items were sterilized under the validated process, the expectation is that not more than one would carry a viable organism.
Two consequences follow, and both matter in practice.
First, sterility is never demonstrated by testing the product. A clinic cannot culture a sterilized instrument and conclude that the item was sterile because the culture was negative. A single negative culture has no statistical power against a 10-6 probability. That is why biological indicators, process challenge devices and physical parameters are used instead: they interrogate the process, not the item.
Second, SAL 10-6 is the acceptance criterion for terminally sterilized devices under EN 556-1. The current edition, EN 556-1:2024, retains the requirement; if a supplier's documentation cites "EN 556-1:2001" as the basis of a claim, the citation is out of date even where the conclusion is not.

Log reduction and SAL are not the same claim
Log reduction describes how much a population was reduced by a process. A 6-log reduction takes 1,000,000 organisms down to an expected 1. SAL describes the probability of an organism remaining on the final item. The arithmetic produces similar-looking exponent numbers, which is exactly why they get conflated.
The difference is what sits on each side of the calculation.
| Quantity | What it describes | Reference point | Typical use |
|---|---|---|---|
| Log reduction | Change in population across a process | Starting bioburden | Disinfection and cleaning claims; sterilant efficacy studies |
| D-value | Time or dose to reduce a population by 1 log under stated conditions | A defined spore population, carrier and conditions | Sterilizer validation, cycle development, BI selection |
| SAL | Probability of one surviving organism on one item | The final sterilized product | Sterility acceptance criterion for terminally sterilized devices (EN 556-1) |
| Z-value | Temperature change needed to change the D-value by a factor of 10 | Same population as the D-value | Comparing performance across cycle temperatures |
The practical rule: log reduction tells you what the process did to a population you can count; SAL tells you the risk that remains on the item you cannot test. A written procedure that mixes the two will mislead whoever follows it.
The arithmetic that ties D-value, bioburden and SAL together
For an overkill steam cycle the relationship is a small piece of arithmetic: SAL = N0 x 10^-(F/D), where N0 is the initial bioburden or the biological indicator population, D is the D-value of that population under the cycle conditions, and F is the exposure time at the reference temperature. The numbers show why biological indicators are designed the way they are.
| Scenario | N0 | D-value | Exposure (F) | D-value multiples | Resulting SAL |
|---|---|---|---|---|---|
| Spore test BI, complete kill expected | 1.0 x 10^6 | 1.5 min | 12 min at 134 C | 8 | 1.0 x 10^-2 on the carrier, then read against the survival-kill response in ISO 11138-1:2017 Annex E |
| Same BI, longer exposure | 1.0 x 10^6 | 1.5 min | 21 min at 134 C | 14 | 1.0 x 10^-8 |
| Clinic bioburden-based cycle | 1.0 x 10^3 | 1.5 min | 15 min at 134 C | 10 | 1.0 x 10^-6 |
Two readings of that table matter. A biological indicator with a 10^6 population and a 1.5-minute D-value reaches SAL 10-6 on the carrier at roughly nine D-values of exposure, which is why a "no growth" result is evidence that the cycle delivered a large multiple of the intended lethality rather than a marginal one. And the last row shows why a bioburden-based validation can reach the same SAL with less exposure than a 10^6 overkill approach: the starting population is smaller. Both are legitimate; they are different validation routes, not different acceptance criteria.
A D-value without stated conditions is not a number you can use. ISO 11138-1:2017 provides two normative routes to it - Annex C, the survivor curve method, and Annex D, the fraction-negative method - and they are not interchangeable in a report. When a supplier quotes a D-value, the same document should state the organism, the carrier, the recovery medium, the reference temperature and the method annex used. If it does not, ask for the test report rather than reusing the figure in a validation file.
How do I read a biological indicator certificate of analysis
A certificate of analysis for a steam biological indicator carries a small set of figures, and five of them are enough to audit a purchase.
- Organism and strain. For moist heat, Geobacillus stearothermophilus is the organism in common supply; records should name it and give the strain designation, not only the genus.
- Population per carrier. Usually expressed as a range in CFU, for example a nominal 10^5 or 10^6. The figure that matters for the arithmetic above is the population attached to the carrier, not the number printed on the box front.
- D-value with its reference temperature. A D-value measured at 121 C cannot be read across to a 134 C cycle without a z-value.
- Survival-kill response window. ISO 11138-1:2017 Annex E covers this: it defines the survival time and the kill time for the population, which is what makes a "no growth" result interpretable rather than merely negative.
- Incubation conditions and the reference incubation time. ISO 11138-8:2021 exists because incubation times shorter than the seven-day reference incubation time in ISO 11138-1:2017 must be validated as a reduced incubation time rather than assumed. A supplier claiming a shorter readout is claiming a validated reduced incubation time, and the validation is what to ask for.
Where these numbers sit in Canadian dental practice
SAL is a device-manufacturing concept, and most Canadian dental and clinic settings do not calculate it - they buy it. A Class B or Class S benchtop sterilizer is marketed against EN 13060, and the classification of the sterilizer is a different question from the SAL of a sterilized device.
What Ontario clinics do have to get right is monitoring frequency, and here the vocabulary correction matters more than the arithmetic.
> Spore testing (biological indicators) must be done each day the sterilizer is used and the results of the test recorded in a logbook. A control test must also be used when spore testing is performed and those results also recorded.
That sentence is from the Royal College of Dental Surgeons of Ontario's published infection prevention and control article. The Ontario position is daily monitoring for each sterilizer on every day it is used. Weekly biological indicator monitoring is the CDC and AAMI ST79 baseline used in most United States jurisdictions, not the Ontario requirement. Public Health Ontario's PIDAC guidance goes one step further and ties the daily requirement to cycle type: a biological indicator is to be used each day that the sterilizer is used and with each type of cycle that is used that day. A clinic running a 134 C flash cycle and a longer gravity cycle on the same day is running monitoring for both.
Two consequences for ordering follow from that frequency. A clinic monitoring daily against the Ontario position consumes roughly 250 biological indicators per sterilizer per year, so the 24-hour self-contained 25-pack is a monthly line item for a single sterilizer, and the rapid reader seed trial lets a clinic run the 3-hour readout workflow before committing to a reader. Where a shorter readout suits the cycle mix, the 3-hour fluorescence indicator carries the same spore population on a different readout schedule.
Does a longer cycle produce a better outcome
No. Exposure time is one input to lethality, and it is not the input that fails most often in a clinic. Steam quality, air removal, load configuration and instrument preparation decide whether the validated exposure reached the load at all. A sterilizer that holds 134 C for the full cycle while carrying a pocket of air inside a hollow instrument delivers far less lethality at the instrument surface than the chamber sensor recorded, which is why hollow-load testing and process challenge devices exist, and why a passing cycle printout is not a sterility claim.
Sources
- ISO 11139:2018, Sterilization of health care products - Vocabulary (ISO catalogue entry) - checked 16 September 2026
- ANSI/AAMI/ISO 11138-1:2017, Biological indicators - Part 1: General requirements (AAMI preview copy, contents and annexes) - checked 16 September 2026
- ISO 11138-8:2021, Biological indicators - Part 8: Method for validation of a reduced incubation time - checked 16 September 2026
- BS EN 556-1:2024, requirement for devices designated "STERILE" - Part 1 - checked 16 September 2026
- ISO 17665:2024, moist heat (lifecycle note on ISO 17665-1:2006) - checked 16 September 2026
- RCDSO, Infection prevention and control in the dental office: some common issues - checked 16 September 2026
- Public Health Ontario, PIDAC guidance on cleaning, disinfection and sterilization in all health care settings, 3rd edition, May 2013 - checked 16 September 2026
- RCDSO standards, guidelines and advisories on infection prevention and control - checked 16 September 2026
- ISO 11140-6:2022, Type 2 indicators and process challenge devices for small steam sterilizers - checked 16 September 2026
- ISO 14937:2009, sterilization process validation - general requirements - checked 16 September 2026
- ISO 11135:2014 with Amendment 1, Ethylene oxide sterilization requirements - checked 16 September 2026
- EN ISO 11138-3:2017, Biological indicators for moist heat sterilization processes - checked 16 September 2026
- ISO 11138-5:2017, Biological indicators for low-temperature steam and formaldehyde sterilization processes - checked 16 September 2026
- ANSI/AAMI/ISO 11138-7:2019, Guidance for the selection, use and interpretation of results (AAMI preview copy) - checked 16 September 2026
- Kunyk et al., Daily Use of Biologic Indicators in General Dental Practice, J Can Dent Assoc 2021;87:l11 - checked 16 September 2026
- Vatanparast et al., Biological indicator failure rates in Saskatchewan dental practices, BMC Oral Health 2024 - checked 16 September 2026
Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers.
Related reading
The probability vocabulary used above is expanded in our explainer on what SAL 10-6 means for sterilization, and the same arithmetic applied to logged validation figures is in D-value, z-value and SAL in sterilization validation. For what the numbers mean at the bench rather than in a validation report, see biological indicator kill time explained, and the monitoring layers that sit around the spore test are set out in our sterilization compliance hub.
Related Reading
- Explainer on what SAL 10-6 means for sterilization
- D-value, z-value and SAL in sterilization validation
- Biological indicator kill time explained
- Sterilization compliance hub
- dental clinic sterilization supplies
- free Ontario sterilization compliance log
Related reading
Frequently Asked Questions
Is SAL 10-6 the same as a 6-log reduction
No, although the exponents look alike. A 6-log reduction describes a decrease in a countable population across a process. SAL 10-6 is the probability that a single viable micro-organism remains on a single item after sterilization, and it is the acceptance criterion set by EN 556-1 for terminally sterilized devices. A process can be described in log reductions; an item's sterility claim is expressed as a sterility assurance level.
Can I demonstrate sterility by culturing a sterilized instrument
No. Culture testing of the item has no statistical power against a 10-6 probability, and the sampling involved would not be valid even at much coarser targets. Sterility assurance is demonstrated by a validated process, monitored with physical parameters, chemical indicators and biological indicators. The biological indicator interrogates the process; it does not certify the individual instrument.
Why does my biological indicator certificate give two D-value determination methods
Because ISO 11138-1:2017 provides two normative methods: Annex C, the survivor curve method, and Annex D, the fraction-negative method. They are alternatives, and the certificate or test report should identify which one produced the figure. A D-value quoted without a method, an organism, a carrier and a reference temperature is not usable in a validation file.
Does ISO 11138-8 change the seven-day incubation
It does not change it; it provides the method for departing from it. The seven-day reference incubation time sits in ISO 11138-1:2017, and ISO 11138-8:2021 specifies the test method used to establish or confirm a reduced incubation time shorter than that reference. That is why a three-hour or 24-hour readout claim should come with a reduced incubation time validation rather than an assertion.
Is weekly biological indicator testing acceptable in Ontario
Not as an Ontario requirement. The RCDSO's published guidance states that spore testing must be done each day the sterilizer is used, with a control test recorded alongside it. The CDC and AAMI ST79 baseline in most United States jurisdictions is weekly monitoring; the Ontario requirement is each day of use and each cycle type used that day.
Which standard replaced ISO 17665-1
ISO 17665:2024 withdrew and replaced ISO 17665-1:2006. The 2024 edition consolidated the former Part 1 requirements with the former ISO/TS 17665-2 and ISO/TS 17665-3 technical specifications into a single standard, so documents that still cite ISO 17665-1:2006 as the current requirement are citing a withdrawn edition.
What is a z-value and do I need one
A z-value is the temperature change required to change the D-value by a factor of ten for the same population. You need one whenever a D-value has to be carried from one cycle temperature to another, for example from a 121 C gravity cycle to a 134 C vacuum-assisted cycle. Without a z-value, a D-value is locked to the temperature at which it was measured.
CliniEco Medical distributes in Canada under MDEL #35334.
0 commentaire