How Many Class 5 Chemical Integrators Does a Diagnostic Laboratory in Quebec Use per Month, and How Many 200 per Pack Boxes Is That? One Strip per Load

CliniEco steam chemical integrator - Class 5, ISO 11140-1 Type, 200 strip pack

How Many Class 5 Chemical Integrators Does a Diagnostic Laboratory in Quebec Use per Month, and How Many 200 per Pack Boxes Is That? One Strip per Load

A diagnostic laboratory in Laval runs four loads on a weekday, two cycle types, and one sterilizer. The consumable line that gets under-ordered is not the pouch, because the pouches are counted every time a bench restocks. It is the Class 5 integrator, because the strip is pulled once per load and nobody multiplies loads by working days until the box is empty in the middle of a release.

The arithmetic is short. What makes it worth a page is that Quebec's count is per load through a process challenge device, while the indicator most people picture when they hear chemical indicator sits inside each package. Two different indicators, two different multipliers, and one of them is four to ten times larger than the other.

How many Class 5 chemical integrators does a diagnostic laboratory in Quebec use per month?

About ninety at four loads a day.

The build-up: four loads a day on 21 working days is 84 strips, because one load takes one strip through a process challenge device. A month contains about 4.3 weeks, so four or five test packs are run, and each test pack carries its own type 5 integrator alongside a biological indicator. That is 89 strips on a normal month, planned at 90.

The CliniEco Medical steam chemical integrator, Class 5, ISO 11140-1 type 5 is stocked in a 200-strip pack and in a case of five such packs, which is the unit the annual figure should be turned into before the order is placed.

What the figure is not: a function of package count. A laboratory sealing 1,200 pouches a month does not need 1,200 Class 5 strips. It needs one per load.

Why is the count per load and not per package in Quebec?

Because the provincial guidance separates the two jobs, and the two jobs are done by different indicator classes.

The per-package requirement is satisfied from inside the packaging system: each package carries a type 1 indicator on the outside and a type 4 indicator inside. Those two classes exist to show that the package was exposed to the process, and they are consumed with the pouch. A CliniEco Class 4 self-sealing pouch with an internal and external indicator is the package-level consumable in that calculation; the type 5 strip is the load-level one.

The per-load requirement is satisfied by a process challenge device carrying a type 5 integrator. The type 5 strip is the integrating indicator that responds to the combination of time, temperature and steam quality across the cycle, and it is what the load release argument rests on. One per load.

CliniEco steam chemical integrator - Class 5, ISO 11140-1 Type, 200 strip pack

Three levels, three classes, three consumption rates:

Level Indicator Consumed
Each package Type 1 outside, type 4 inside With the pouch
Each load Type 5 in a challenge device One per load
Each week, per cycle type Biological indicator plus type 5 One test pack

That structure explains the most common ordering error in a Quebec laboratory: buying Class 5 strips at the pouch replenishment cadence. The pouch is a package-level consumable; the integrator is a load-level consumable. A lab running 20 packages a load over-orders type 5 by a factor of twenty under that assumption.

What does the province require before a laboratory can operate at all?

A permit, and the permit period is worth knowing because it explains why load records line up the way they do.

Section 31 of the Act respecting medical laboratories provides that, apart from the Institut national de sante publique du Quebec, no person may operate a laboratory or an organ and tissue bank without a permit issued for that purpose. Section 37 sets the permit period at twelve months ending on 31 December each year, renewed on that date where the holder meets the regulatory conditions.

For a sterilization program, the practical consequence is documentary. Load release records, indicator results and biological indicator outcomes are part of the laboratory's quality evidence, and the operating year they are filed against ends on a fixed date. A consumable plan that runs on a fiscal year starting in April will not line up with the permit year, and the mismatch shows up as an awkward count at the wrong moment.

How do you convert loads per day into strips per month?

Multiply loads by 21 working days, add the weekly test packs, and round up. The table does that for four volumes, then converts the monthly figure into 200-strip packs for the year.

Loads a day Strips a month Strips a year 200 packs a year
2 47 564 3
4 89 1,068 6
6 131 1,572 8
8 173 2,076 11

Class 5 steam chemical integrator strip - REJECT and ACCEPT window comparison

The yearly column is the monthly figure multiplied by twelve, which assumes the laboratory does not close in August. A laboratory with a two-week summer shutdown and a week at the end of December should use 240 working days in the year rather than 252, and the boxes move down by one at the four-load volume.

How many 200 per pack boxes is that in a year?

Six packs at four loads a day, which is 1,200 strips against a requirement of about 1,068.

The shape of the order matters more than the total. A 200-strip pack lasts a little over two months at 90 strips a month, so single packs mean six receiving events a year and six chances for a lot number to be recorded late. A case of five 200-strip packs covers 1,000 strips, which is eleven months at this volume, and the sixth pack covers the tail of the year.

That case arithmetic is also the reason a laboratory and a small clinic buy the same product in different units: the same strip, ordered in packs by a clinic running one load a day, and in cases by a laboratory running four.

What else goes in the test pack with the integrator?

A biological indicator.

The provincial guidance describes the test pack as a sterilization pouch holding a biological indicator and a type 5 integrator, run for each sterilizer per cycle type each week, with a daily frequency described as the recommendation. The CliniEco 24-hour self-contained biological indicator, 25 tests per pack is the consumable that supplies that half of the pack.

Two details follow from the same guidance and both affect the order. The air-removal test cannot share a cycle with the process challenge device, so a lab with two cycle types is not saving a cycle by combining them. And a load containing non-sterile implantable material takes its own challenge device with a biological indicator and a type 5 integrator, which is a second strip in addition to the routine per-load strip.

A laboratory that runs implant loads should therefore add a line to the arithmetic rather than a percentage: implant loads per month, times two, added to the base figure.

CliniEco Class 4 self-sealing sterilization pouch with dual indicator

How does the laboratory's quality system record all of this?

Every record has to be retrievable, which is why the count and the paperwork are the same problem.

ISO 15189:2022 sets out the quality and competence requirements for medical laboratories, and the sterilization records sit inside that system: which load, which cycle, which indicators, which result, who released it. The packaging side is governed by ISO 11607-1 for sterile barrier systems, and the sterilization process itself by ISO 17665-1 for moist heat. The indicator classes themselves come from ISO 11140-1.

None of those standards tells a laboratory how many strips to buy. What they do is define what has to be recorded for each strip that is used, which is why an under-order is not merely an inconvenience: it turns into a load that cannot be released on the evidence required.

What do the numbers look like when they break?

Four ways, in the order they usually appear in a laboratory running a mixed bench.

Cycle types. The test pack is per sterilizer per cycle type, so adding a second cycle type adds test packs and adds a weekly biological indicator, without adding a single routine load.

Implantable loads. Each one carries a second challenge device. A laboratory that starts taking implant work at a low volume can see its type 5 consumption rise by a quarter while its load count barely moves.

Holiday and shutdown weeks. The monthly figure assumes 21 working days. A month with a two-week closure is a month with 11, and a par level built on 21 leaves a laboratory short in the busiest weeks of the two adjacent months.

Storage and expiry. Strips held past their expiry or stored outside the manufacturer's conditions are not usable evidence. A year of packs bought at once is fine if the storage is dry and temperature-stable; a year of packs kept in a warm room beside the sterilizer is not.

Ordering for a laboratory or a multi-site network in Quebec

Laboratories buy on account, not per pack, and the two things that matter are lead time on the 200-strip pack and consistency of the lot trail across sites. A network that standardises one type 5 across its benches gets comparable release records; a network that buys whatever is in stock gets a folder of strips that are all legitimate and none of them comparable.

Ordering for a laboratory, a hospital bench or a multi-site network? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.

Running a diagnostic laboratory and rebuilding the load-release folder? The diagnostic laboratory supply pages group the pouches, indicators and biological indicators a bench orders together, and a sterilization compliance consultation covers the per-load and per-package requirements in writing.

References

  1. Ordre des dentistes du Quebec, Lignes directrices sur la prevention et le controle des infections et le retraitement des dispositifs medicaux (rev. February 2026) (checked 12 November 2026)
  2. Act respecting medical laboratories, organ and tissue conservation and the disposal of human biological material (chapter L-0.2), section 31, permit required (checked 12 November 2026)
  3. Act respecting medical laboratories (chapter L-0.2), section 37, permit period and renewal (checked 12 November 2026)
  4. CSA Z314:23, Reprocessing of reusable medical devices in health care settings (standards record) (checked 12 November 2026)
  5. ISO 11140-1:2014, Sterilization of health care products, chemical indicators, general requirements (checked 12 November 2026)
  6. ISO 11140-4:2007, Chemical indicators, class 2 indicators as steam sterilization process challenge devices for Bowie and Dick-type tests (checked 12 November 2026)
  7. ISO 11138-1:2017, Sterilization of health care products, biological indicators, general requirements (checked 12 November 2026)
  8. ISO 17665-1:2006, Sterilization of health care products, moist heat, requirements for the development, validation and routine control of a sterilization process (checked 12 November 2026)
  9. ISO 15189:2022, Medical laboratories, requirements for quality and competence (checked 12 November 2026)
  10. ISO 11607-1:2019, Packaging for terminally sterilized medical devices (checked 12 November 2026)
  11. ISO 11737-1:2018, Sterilization of health care products, microbiological methods, determination of a population of microorganisms on products (checked 12 November 2026)
  12. ISO 11135-1:2007, Sterilization of health care products, ethylene oxide, requirements for the development, validation and routine control of a sterilization process (checked 12 November 2026)
  13. Public Health Agency of Canada, Routine Practices and Additional Precautions for Preventing the Transmission of Infection in Healthcare Settings (checked 12 November 2026)
  14. Public Health Ontario, Infection Prevention and Control for Clinical Office Practice (checked 12 November 2026)
  15. Medical devices portal, federal regulator (device classes and licensing pathway) (checked 12 November 2026)
  16. Charter of the French Language (chapter C-11), section 51, inscriptions on products and their wrapping (checked 12 November 2026)

Printable companion: the printable compliance log and checklist library has load-release sheets built around the per-load integrator and the weekly biological indicator, so the count in this article can be logged as it is consumed.

Prefer a number to a table? The sterilization usage calculator converts loads per day and cycle types into strips per month and packs per year.

Related Reading

Frequently Asked Questions

How many Class 5 chemical integrators does a diagnostic laboratory in Quebec use per month?

About ninety at four loads a day: one strip per load through a process challenge device is 84 on 21 working days, plus roughly five weekly test packs, each of which carries its own type 5. Six loads a day takes the figure to about 131 and eight loads to about 173.

Is the Class 5 count based on packages or on loads in Quebec?

On loads. The provincial reprocessing guidance puts a process challenge device with a type 5 integrator in each load, while the per-package indicator requirement is satisfied by a type 1 indicator outside the package and a type 4 inside it. Budgeting type 5 strips against pouch consumption over-orders by a wide margin.

How many 200 per pack boxes is ninety strips a month?

Three boxes short of five. Ninety strips a month is 1,080 a year, and 1,080 divided by 200 is 5.4 boxes, so a five-box case of 1,000 strips covers eleven months and a sixth box covers the year with a little float.

What else goes in the test pack with the integrator?

A biological indicator. The guidance describes the test pack as a sterilization pouch containing a biological indicator and a type 5 integrator, and it applies to each sterilizer per cycle type each week, with daily described as the recommendation.

Can the air-removal test share a cycle with the challenge device?

No. The guidance states the two tests cannot be run in the same cycle, which is why a laboratory with two cycle types budgets two test packs a week and two separate cycles for them.

Do loads containing non-sterile implantables change the count?

Yes. A load with non-sterile implantable material gets its own challenge device with a biological indicator and a type 5 integrator, in addition to the routine per-load strip, so those loads consume two strips and one indicator.

What does the provincial permit have to do with consumable planning?

Section 31 of the Act respecting medical laboratories requires a permit to operate a laboratory in Quebec, and section 37 sets the permit period at twelve months ending on 31 December. Load-release records are kept against that operating year, which is why many laboratories close their consumable budget on the same calendar.

How long can a Class 5 integrator be stored before use?

Storage conditions belong to the manufacturer's instructions and to the specification the strip is qualified against, not to a general rule. Keep strips in the original packaging, away from heat and moisture, and check the expiry date on the pack at receiving rather than at the point of use.