CliniEco Dental Ultrasonic Cleaner — CD-4862, 6 L with 5-Preset Heating and Timer
Ships to Canada only
Specifications
| Tank capacity | 6 L (6,000 ml) stainless steel |
|---|---|
| Tank opening | 32.5 × 20.4 × 9.9 cm — accepts instruments up to 34 cm long |
| Frequency | 35 kHz, two industrial-grade transducers |
| SKU | CLI-ULTRA-CLEAN-CD4862 |
| Brand | CliniEco Medical |
| Category | Sterilization & Cleaning Equipment |
| Availability | In stock |
Pricing
| Order | Unit price |
|---|---|
| 1 unit | $462.00 CAD |
| Case and carton prices are in the size table below | |
| Request quote for clinic and volume orders | |
Read the full product guide 9 sections · specs, use, compliance
Quick answer: A 6 L tank — cassette size — with five fixed temperature presets between 40 °C and 60 °C and a full-range 1 to 30 minute timer, so the cycle is written down rather than guessed.
The 6 L class is the size at which a standard dental instrument cassette fits in one cycle, and this is the unit in the range that treats temperature as a fixed setting rather than a dial. Five presets — 40, 45, 50, 55 and 60 °C — plus a timer that runs the whole 1 to 30 minute range, and a three-colour LED display that shows the set-point while the cycle runs. Tank opening is 32.5 × 20.4 × 9.9 cm, which takes an instrument up to 34 cm long.
Scope of use. This is a cleaning device. It removes soil — cement residue, polishing paste, dried blood in hinges and threads — as a step in instrument reprocessing. It makes no microbicidal claim, and it does not replace any later step in the workflow set out in the instrument manufacturer’s instructions for use. The Health Canada guidance on reprocessing information is why the instrument manufacturer’s instructions — not the cleaner’s specification sheet — decide what happens to a given instrument after it comes out of the tank.
What are the CD-4862 specifications?
The presets are the operational difference between this unit and a tank with a rotary thermostat: the number you set is the number you record.
Tank capacity: 6 L (6,000 ml) stainless steel
Tank opening: 32.5 × 20.4 × 9.9 cm — accepts instruments up to 34 cm long
Frequency: 35 kHz, two industrial-grade transducers
Power: 260 W rated, AC 100–120 V, 60 Hz, North American plug
Timer: 1 to 30 minutes, full range, with three-colour LED display
Heating: Five presets: 40 / 45 / 50 / 55 / 60 °C, with the °F equivalent shown
| Field | Specification |
|---|---|
| Tank capacity | 6 L (6,000 ml) stainless steel |
| Tank dimensions | 32.5 × 20.4 × 9.9 cm (length × width × depth), SUS304 stainless steel tank |
| Tank opening | 32.5 × 20.4 × 9.9 cm — accepts instruments up to 34 cm long |
| Frequency | 35 kHz, two industrial-grade transducers |
| Power | 260 W rated, AC 100–120 V, 60 Hz, North American plug |
| Timer | 1 to 30 minutes, full range, with three-colour LED display |
| Heating | Five presets: 40 / 45 / 50 / 55 / 60 °C, with the °F equivalent shown |
| Safety features | Dual heating protection, moisture-resistant PCB, detachable power cord |
| Electrical certification | cETL — Intertek report 210716104GZU-001, to UL 73 and CSA C22.2 No. 68 |
| Included | Unit, one-year manufacturer warranty; stainless basket available as an accessory |
| List price | CAD $649.00 |
Does this unit log cleaning cycles?
No. The CD-4862 has no internal data logger and cannot export cycle data to a USB drive or a CSV file. If your reprocessing protocol requires machine-generated cycle records, choose a unit built for that purpose — this model is not one.
What the CD-4862 gives you instead is everything needed to keep a defensible manual record: a front-panel timer that counts the cycle, a temperature set-point you can read and write down, and a drain valve so the solution-change date is unambiguous. Under CSA Z314:23 the documentation obligation sits inside your reprocessing quality system rather than inside the appliance, so a signed paper log is not a lesser record — it is the record the standard asks for.
Five fields per cycle is enough: date, operator, temperature set-point, cycle length, and the date the solution was last changed. One line, end of shift.
Which tank size in the range fits your room?
Two 6 L units sit in this range and they are not the same product. This one fixes the temperature in 5 °C steps and runs 35 kHz; the CS6.2 runs 32 and 48 kHz on one generator. If your protocol names a temperature, the preset ladder is the easier machine to audit.
| Model | Tank | Frequency | Built for | List price (CAD) |
|---|---|---|---|---|
| CS4.0 | 4 L | 32 / 48 kHz | Single chair, loose instruments | $449.00 |
| CD-4862 | 6 L | 35 kHz | Full cassette, five temperature presets | $649.00 |
| CS6.2 | 6 L | 32 / 48 kHz | Full cassette, dual frequency | $699.00 |
| CD-4875 | 7.5 L | 35 kHz | Larger trays, five temperature presets | $849.00 |
| CS8.0 | 8 L | 49 kHz | Two chairs into one cycle | $999.00 |
| CS20 | 20 L | 35 kHz | Several chairs or batch work | $1,199.00 |
The length limit is worth measuring before you buy. A 34 cm working length covers most hand instruments and many extraction forceps; anything longer belongs in a 20 L tank where the chamber is 42 cm long.
How do you run a repeatable cleaning cycle on the CD-4862?
This unit is built for the practice whose protocol already names a temperature. The presets exist so the number in the protocol and the number on the panel are the same number.
| Step | Setting | Why it matters |
|---|---|---|
| 1. Fill | Solution at the label dilution, to the working level in the tank | The same fill level every day keeps the load behaviour constant |
| 2. Preset | Choose 40, 45, 50, 55 or 60 °C to match the solution label | A fixed preset is a recordable number; a thermostat dial is an estimate |
| 3. Load | Instruments in a basket or tray, joints open, not stacked | Cavitation does not reach shielded surfaces |
| 4. Time | 1 to 30 minutes — set the interval your protocol names | The full range means the timer is not the constraint on the protocol |
| 5. Drain, dry, record | Drain through the valve, dry, then log temperature, time and operator | The three data points an inspector reconstructs a cycle from |
Keep the solution inside the temperature window printed on its label rather than at the highest preset. Hotter is not cleaner: above the solution manufacturer's range, the enzymes lose activity and the tank starts to leave a film on stainless.
What an infection-control inspection asks for is the record, not the machine. Ontario’s provincial reprocessing guidance asks for documentation and maintenance of records for each component of the cleaning, disinfection and/or sterilization process, based on the manufacturer’s recommendations, and repeats that records must be retained according to the facility’s policy (Public Health Ontario, “Best Practices for Cleaning, Disinfection and Sterilization of Medical Equipment/Devices In All Health Care Settings”, 3rd edition, checked 2026-11-07). A unit with a drain valve, a visible timer and a stable temperature set-point is what makes that record short enough to write at the end of a shift.
What does the cETL mark cover, and what does it not?
This unit is covered by a separate Intertek report from the CS range — 210716104GZU-001 rather than 220112139GZU-001 — which is worth knowing when a technician asks which family the certificate names. The mark on the unit is an ETL mark issued by Intertek, tested against UL 73 and CSA C22.2 No. 68. It is an electrical safety certificate. It is not a UL listing and it is not CSA certification, and it says nothing about how well the tank cleans. Canadian reprocessing practice sits on a separate standard, CSA Z314:23, whose scope puts documentation inside the reprocessing quality system rather than inside the appliance.
| Covers | Does not cover |
|---|---|
| Electrical safety of the appliance, assessed to UL 73 and CSA C22.2 No. 68 | Cleaning performance on your instruments, on your water hardness, or at your solution concentration |
| The specific unit family named on Intertek report 210716104GZU-001 | The condition of an individual unit after a drop, a repair, or a modification |
| Compliance of the supply, plug and cord assembly for the North American market | Any clinical outcome, or any claim that the unit removes or inactivates micro-organisms |
| The identity of the certificate holder and the factory it was issued to | Your own reprocessing records, which are yours to keep and produce |
What is in the box, and what should be ordered with it?
One cleaner, 6 L, 110 V with a North American plug and the manufacturer warranty. The stainless basket is an accessory. The power cord detaches, which matters in a room where the bench is wiped down between sessions.
Multi-enzyme ultrasonic cleaning solution is what goes in the tank, sold as a 1 L bottle or a 5 L pail and diluted 1:400 to 1:600. The 6 L CS6.2 is the unit to look at if your load grows into full cassettes. Read how to choose a tank size and what to record on the cleaning log before you commit to a size.
What does one tank fill cost in cleaning solution?
The 6 L tank is the cassette size, and it is also the worked example on the solution page: a 6 L fill needs 15 mL of concentrate at 1:400 and 10 mL at 1:600. Those two figures are the whole chemistry cost of a cassette cycle — what is left for you to check is the pack price on the day you order, because that is the only variable that moves.
| Dilution on the label | Concentrate per full tank (6 L) | Fills from a 1 L bottle | Fills from a 5 L pail |
|---|---|---|---|
| 1:400 — heavier soil | ≈ 15 mL | 66 fills | 333 fills |
| 1:600 — light soil | ≈ 10 mL | 100 fills | 500 fills |
Cost per fill = the current pack price divided by the fills in that pack. Both halves of that sum move — the pack price with the market, the fills with how full you run the tank — which is why the page gives the arithmetic instead of a per-cycle figure. If your protocol changes the bath once per working day, five days a week, 66 fills is roughly 13 working weeks from one bottle.
Which cleaning solution goes in this tank, and when should it be changed?
Use the multi-enzyme ultrasonic cleaning concentrate — 1 L bottle or 5 L pail — diluted 1:400 to 1:600. It is low-foam by design, so the cavitation field does the work instead of a head of bubbles, and it is freely rinsable on stainless, anodised aluminium and plated instruments. The label window is 20–50 °C for 2 to 10 minutes; keep the bath inside it rather than at the highest preset this unit offers.
Change the bath on the schedule your label and protocol set, and log the change with the same five fields as the cycle: date, dilution, fill level, temperature, initials. The drain valve makes the job unambiguous, and the record is the part an inspector reads — Ontario reprocessing guidance asks for documentation and maintenance of records for each component of the cleaning, disinfection and/or sterilization process (Public Health Ontario provincial reprocessing guidance, 3rd edition, checked 2026-11-07).
What is the lead time, and what maintenance does the unit need?
This unit is built to order — allow 3–4 weeks, and we confirm the dispatch date before it ships. The manufacturer warranty runs one year and is supported in Canada. CliniEco Medical holds Health Canada establishment licence (MDEL) #35334 — importer and distributor — which is separate from the warranty terms. What the unit asks of you in return is short, and mostly about the bath rather than the electronics.
| Task | Interval | Why it matters |
|---|---|---|
| Drain and refill | At the interval the solution label and your protocol set | Soil load accumulates in the bath whatever the instruments look like when they come out |
| Run a fresh fill before the first load | Every refill | Dissolved air in new solution damps cavitation, and the bath should be at working temperature before an instrument goes in |
| Rinse tank and basket | Every change | Residue left on stainless carries soil into the next fill and films the tank over time |
| Wipe lid and check cord, plug and drain valve | Weekly | The three parts a drop or a wipe-down damages first |
| Keep the level above the working minimum | Every cycle | Running dry is the one failure mode that ends a tank’s service life early |
A presets-based unit invites a habit worth avoiding: turning the temperature up when a load comes out dull. Above the solution label’s range the enzymes lose activity and the tank starts to film over — the fix is a fresh fill at the correct dilution, not a hotter one.
Related reading
- Choosing an Ultrasonic Cleaner Tank Size for a Canadian Dental Clinic
- What Should a Canadian Clinic Record for Ultrasonic Cleaning?
- Enzymatic Cleaner Cost per Litre in Canada: Dilution Economics
CliniEco Medical is a licensed medical device establishment (MDEL #35334). We are the Canadian importer of record for this unit: duties and brokerage are handled on our side, and you receive it from our Toronto warehouse with a one-year manufacturer warranty supported in Canada. Questions on which size fits your room go to our sterilization desk.
Built to order. Allow 3–4 weeks — we confirm the dispatch date before the unit ships.
Common questions
What are the five temperature presets on the CD-4862?
40, 45, 50, 55 and 60 °C, with the Fahrenheit equivalent displayed at the same time. The preset should be chosen to sit inside the temperature window printed on the cleaning solution's label, not at the top of the ladder by default.
How long is the timer, and is it a full range?
1 to 30 minutes across the whole range, not a set of fixed steps. That matters if a protocol names an interval such as 12 minutes, which a stepped timer cannot reproduce exactly.
What is the longest instrument that fits?
About 34 cm, set by the 32.5 × 20.4 × 9.9 cm tank opening. Longer items — some extraction forceps, long-handled retractors — need the 20 L tank, whose chamber is 42 cm long.
Does the CD-4862 sterilize or disinfect instruments?
Neither. It cleans: it removes soil, and it makes no microbicidal claim. Whatever the instrument manufacturer's instructions for use require after cleaning still applies, and your provincial requirements for the rest of the workflow are unchanged.
Which cleaning solution should go in this tank?
A low-foam enzymatic concentrate diluted to the label ratio — ours runs 1:400 to 1:600 — because foam in a 6 L tank displaces the liquid the transducers need to work through.