What is the difference between Class B and Class N autoclaves?

Bowie-Dick test pack labelled for steam sterilization, used for the daily air removal check

What is the difference between Class B and Class N autoclaves?

A clinic replaces a fifteen-year-old benchtop sterilizer. Two quotes come back. One unit is described as Class B, with a price step that reflects it; the other is Class N, quicker to cycle, lower in quote, and described by the sales email as "suitable for dental instruments". Both statements are true. Only one of them answers the question the clinic actually has, which is whether the unit can sterilize the wrapped cassettes and handpieces that go through the practice every day.

The Class N, S and B labels are not marketing tiers. They are type designations in EN 13060, the standard for small steam sterilizers, which applies to units with a chamber volume under 60 litres that cannot accommodate a sterilization module of 300 mm by 300 mm by 600 mm. What separates the types is how the machine removes air, and therefore which loads it can be validated to process.

What do the Class N, S and B labels actually mean?

Class N means no vacuum stage. The sterilizer relies on gravity displacement, steam enters at the upper end of the chamber and pushes air out through a drain beneath the load, and the cycle is validated for unwrapped solid instruments. Class S means there is a vacuum stage, but the validation covers only the load types the manufacturer specifies in its instructions for use. Class B means there is a vacuum stage and the cycle is validated across the full range of load types the standard defines: solid, porous, hollow, wrapped and unwrapped.

The naming confuses people because Class B is not "better" in the sense of a grade. It is broader. A Class N unit that runs a solid, unwrapped load correctly is doing exactly what it was designed to do, and a clinic that only ever processes unwrapped solid instruments can run one without a compliance gap. The mismatch appears the first time someone puts a paper-and-film pouch into a gravity cycle and expects the same result.

Which loads can each class process?

Type Air removal Unwrapped solid Wrapped instruments Porous loads (textiles) Hollow or lumen instruments
N Gravity displacement Yes No No No
S Vacuum stage Yes Per manufacturer's declared load types Per manufacturer Per manufacturer
B Vacuum stage Yes Yes Yes Yes

Sterilization kit components: biological indicators, Bowie-Dick test packs, Class 5 integrators, pouches and autoclave tape

RCDSO points in the same direction without using the class letters: dynamic air removal steam sterilizers, including pre-vacuum and steam-flush pressure pulse units, are preferred over gravity displacement sterilizers (Infection Prevention and Control in the Dental Office, p.22). It also flags the instrument design feature that drives the choice, noting that long or narrow lumens, sharp bends, screws and hinges affect sterilization efficacy, and listing handpieces, surgical aspirators, implant drills and torque wrenches among them (p.22).

Monitoring consumables follow the same split, which is where the ordering usually goes wrong. A clinic with a pre-vacuum unit needs a pre-vacuum sterilization compliance kit that includes Type 2 Bowie-Dick packs, biological indicators and Class 5 integrators. A clinic still running a gravity unit needs the gravity and non-vacuum kit, because a Bowie-Dick test is not the right daily check for that machine. Sending the wrong kit to a satellite office is a common reason a gravity unit ends up with a vacuum test pack in the cupboard and no air removal test on the log.

Why do wrapped and hollow loads need a vacuum stage?

Air is the obstacle. Steam cannot sterilize a surface it never reaches, and trapped air in a hollow instrument or inside a sealed pouch sits there while the chamber reaches temperature. A gravity cycle can displace air from an open chamber reasonably well and still leave air in a lumen with a sharp bend. A vacuum stage removes air actively before steam is admitted, which is why the validated load range of a vacuum sterilizer is wider.

Packaging makes the same point from the other side. RCDSO notes that when items are packaged, the sterilizing agent takes longer to penetrate, because the packaging creates a barrier the steam has to be drawn or forced through (p.23). A pre-vacuum cycle has a mechanism for that; a gravity cycle has a drain and patience. Where a clinic runs a pre-vacuum unit, the daily air removal check is only repeatable if the test consumable is the right one: a Type 2 Bowie-Dick test pack intended for that check keeps the pack itself from becoming the variable.

Which class fits a Canadian clinic?

Bowie-Dick test pack labelled for steam sterilization, used for the daily air removal check

For a practice that sterilizes wrapped cassettes, handpieces or any hollow instrument, the working answer is a vacuum sterilizer, and in practice that means Class B or a Class S unit whose instructions for use cover the specific loads in question. For a small clinic processing only unwrapped solid instruments immediately before use, a Class N unit can be adequate, provided the clinic's written procedures reflect that scope and nobody quietly starts running pouches through it.

Two equipment rules sit above the class decision. RCDSO requires that sterilization be performed with medical sterilization equipment registered as a medical device in Canada, and that all tabletop sterilizers and reprocessing equipment be CSA approved and accompanied by the manufacturer's instructions for use (p.22, p.26). Neither statement is about class; both are about what an inspector will ask for when the unit is already installed.

How does the class change daily monitoring?

It changes one item. For pre-vacuum sterilizers, RCDSO requires an air removal test with a Type 2 indicator at the beginning of each day the sterilizer is used, with the process challenge device placed in the chamber of an empty sterilizer (p.23). The daily consumable that makes that test repeatable is a Type 2 test pack. Gravity units do not take this test, and a log that shows one every morning for a gravity unit is a sign the paperwork was copied from another site.

Everything else stays the same across classes. Physical parameters are checked and recorded for each load. Chemical indicators go on the outside and inside of each package. A biological indicator runs every day the unit is used and with each cycle type used that day (Public Health Ontario, PIDAC 2013, p.40). Loads containing implantable devices are quarantined until the biological result is known (RCDSO, p.23).

What should be on the spec sheet before ordering?

  • Chamber volume and usable dimensions measured against the cassettes the clinic actually uses, not against the old unit's footprint.
  • The cycle types offered, and whether each is a wrapped, unwrapped, porous or hollow cycle.
  • Whether the unit prints a cycle record or stores it electronically. RCDSO states that tabletop sterilizers with recording devices are preferred and that all new sterilizers must have this feature (p.22).
  • CSA approval and a manufacturer's instructions for use in hand before installation (p.26).
  • Water quality requirements, gasket and filter part numbers, and whether preventive maintenance is documented by the supplier.
  • Which indicators and test packs the cycle types require, ordered at the same time as the unit rather than a month later.

The wider benchtop sterilizer range shows how much varies behind the same class label, and multi-site groups standardising units across locations usually work through a wholesale and institutional account rather than one unit at a time. The class label is the first line of a specification, not the whole of it. Two units can both be Class B and differ in chamber shape, cycle duration, record keeping and what the manufacturer will support in year five. The useful question is not which class is the higher one; it is which load types this clinic actually runs, and whether the unit in front of you is validated for every one of them. Reselling? become a distributor.

References

  1. EN 13060:2014+A1:2018 Small steam sterilizers (checked 29 October 2026)
  2. Infection Prevention and Control in the Dental Office — Standard of Practice (RCDSO) (checked 29 October 2026)
  3. Cleaning, disinfection and sterilization of medical equipment and devices in all health care settings (Public Health Ontario) (checked 29 October 2026)
  4. Reprocessing (Public Health Ontario) (checked 29 October 2026)

Printable companion: the free sterilization log sheet is built around the cycle types a vacuum sterilizer runs, so the daily checks that a Class B unit needs are recorded in the same place as the load.

Matching consumables to the sterilizer you already own is the other half of the decision. The sterilization self-check confirms which indicators and tests your cycle types require before you order.

Not sure how this applies to your own setup? Ask a sterilization compliance specialist — you’ll get a written answer specific to your equipment and province, with the regulation or standard cited.

Related Reading

Frequently Asked Questions

Is a Class B autoclave always the right choice for a dental office?

Not automatically. Class B is the designation for a small steam sterilizer with a vacuum stage validated across the full range of load types, which covers wrapped, porous and hollow loads. RCDSO states that dynamic air removal sterilizers such as pre-vacuum and steam-flush pressure pulse units are preferred over gravity displacement sterilizers (p.22), and a clinic that routinely sterilizes handpieces or wrapped cassettes is working with hollow and packaged loads.

Can a Class N sterilizer sterilize wrapped instruments?

No. A Class N cycle has no vacuum stage, and its validated use is unwrapped solid instruments. Packaging and hollow lumens both depend on air removal to let steam reach the surface, which is the function a vacuum stage provides.

What does Class S mean?

Class S sits between the other two: the sterilizer has a vacuum stage, but it is validated only for the specific load types its manufacturer defines. A Class S unit may handle wrapped solid instruments and not porous loads, or the reverse. The manufacturer's instructions for use are the scope of the claim, and a buyer has to read them load type by load type.

Do I still need a Bowie-Dick test with a Class B unit?

Yes, if the unit uses dynamic air removal. An air removal test with a Type 2 indicator must be performed at the beginning of each day the sterilizer is used, with the process challenge device in an empty chamber (RCDSO, p.23). The test verifies air removal only; it is not a sterility test.

Does buying a higher class reduce the monitoring workload?

No. Monitoring is set by the standard and by the provincial requirements, not by the class label. Physical parameters are checked for each load, a biological indicator runs every day the unit is used and with each cycle type used that day (PHO, p.40), and implant loads stay quarantined until the biological result is known (RCDSO, p.23).

What should a clinic check on the spec sheet before ordering?

Chamber volume against the cassettes actually used, which cycle types the unit offers, whether it prints or stores a cycle record, and whether it carries CSA approval, which RCDSO requires for tabletop sterilizers (p.26). The class label alone does not answer any of those four questions.