A foot clinic buys the drill and the suction line for different reasons and then has to clean them to two different rules. The College of Chiropodists of Ontario writes both into one standard, and the first thing a new staff member has to learn is which part of the room each rule covers. Our six litre benchtop ultrasonic cleaner with dual frequency sits between the treatment chair and the sterilizer in most single-room clinics.

What does the College require for a podiatry drill and its burs?

The Infection Prevention and Control Standard of Practice is direct: all critical and semi-critical medical instruments, including drill burs or bits, are steam or gas sterilized in instrument wraps, cases or plastic peel pouches in an autoclave before use on a patient. The same paragraph names sterile blades and rotary tool disks as examples of single-use items that must not be reprocessed.

Item What it touches Category Accepted processing
Drill burr or bit nail, skin, sometimes soft tissue critical or semi-critical clean, then steam or gas sterilize before use
Rotary tool disk nail, skin labelled single use used once, not reprocessed
Suction tip fluid and debris from the field single use discarded after the patient
Suction liner and tubing fluid single use replaced, not washed and reused
Nippers, curettes, probes nail and skin critical or semi-critical clean, then sterilize

Product information diagram of the CliniEco CS6.2 six litre ultrasonic cleaner, front elevation with the dual frequency controls

How is a drill burr cleaned before it is sterilized?

Cleaning comes before sterilization and is where most failures start, because a burr put into the bath with debris packed in the flutes comes out looking clean. The standard asks for detergent or enzymatic solution and, wherever possible, mechanical cleaning, and it strongly recommends an ultrasonic washer for anything with joints, crevices, lumens or other areas that are hard to reach, which describes a burr and a hinged nipper. Our multi-enzyme ultrasonic cleaning solution is dosed for that step.

What is accepted for suction containers, liners and tubing?

Plan view and side elevation of the six litre ultrasonic tank, drawn to scale with the working depth marked

Disposable, on all three parts. The standard states that suction containers and liners should be disposable and that solidifiers should be used wherever possible, that reusable suction containers should not be used, and that suction tubing shall be disposable. The between-patient change is a liner and a length of tubing.

How often are the bath and the sterilizer checked?

The ultrasonic bath is tested for efficacy at least weekly, and the load monitoring is daily: a biological indicator is used to test the sterilizer each day it is used and with each type of cycle used that day, with a control vial run on each day routine indicators are incubated. Loads are not released until the indicator result is available unless a Type 5 or Type 6 chemical indicator and the cycle parameters justify it, and the College asks for the load log to be kept one year on site and five years on file. Refill the stock that feeds it with Class 4 self-sealing sterilization pouches.

Writing the room's first written routine? The sterilization compliance desk reviews load monitoring and record keeping with single-room clinics.

The podiatry and physiotherapy supplies page groups the cleaning, packaging and waste lines a treatment room restocks.

Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.

References

  1. Infection Prevention and Control Standard of Practice (checked 5 November 2026)
  2. PIDAC guidance on cleaning, disinfection and sterilization in health care settings (checked 5 November 2026)

Printable companion: the instrument cleaning and sterilization log carries a line for the ultrasonic bath change, the cycle type and the biological indicator result, which is the record the College asks a foot clinic to keep.

Work out how many loads a week the room actually runs before buying indicators: the sterilization monitoring usage calculator turns patient volume into loads and indicator counts.

Related Reading

Frequently Asked Questions

Do podiatry drills and burs have to be sterilized between patients?

Yes. The College of Chiropodists of Ontario standard names drill burs or bits directly: critical and semi-critical instruments, including burs and bits, are steam or gas sterilized in wraps, cases or peel pouches before use on a patient.

Can a rotary tool disk marked single use be cleaned and reused?

No. The same standard states that critical and semi-critical items labelled as single use, with sterile blades and rotary tool disks as its examples, must not be reprocessed.

Does a foot clinic have to change the suction liner between patients?

The standard says suction containers and liners should be disposable and that reusable suction containers should not be used, and it states that suction tubing shall be disposable, so the liner and the tubing are both one-patient items.

How often does the ultrasonic bath have to be tested?

Ultrasonic washers, which the standard strongly recommends for instruments with joints, crevices or lumens, must be tested for efficacy at least weekly or according to the manufacturer's instructions, and the bath itself is changed on a written schedule.