What is the accepted between-patient cleaning step for retinoscopes and ophthalmoscopes in Canada?
An optometry practice in Ottawa runs three lanes and one handheld set that travels between them. The retinoscope and the ophthalmoscope are shared all morning, and the habit at the end of each patient is a quick pass with whatever cloth is on the desk. That habit is the problem, and not because the instrument is dirty: it is because the step has a defined level, a defined frequency and a defined record, and a dry cloth satisfies none of them. What the lane actually needs is the single box of 200 individually wrapped cotton swabs for the parts that touch a face, used with a 70% isopropyl wipe, and a written line that says who did it.
What does the provincial table say about an ophthalmoscope?
It names the instrument. Appendix 8 of the environmental cleaning guidance, in its third edition with the November 2025 revision, carries a table of minimum cleaning and disinfection level and minimum frequency for noncritical equipment. The ophthalmoscope sits in that table as clean plus low-level disinfection, between patients. Noncritical is the operative word: the device rests against a brow and a cheek, which is intact skin, so the level required is low-level disinfection and not the high-level processing a tonometer prism would take.
Where does the retinoscope fit?

The retinoscope is not listed by name, and that omission is not a gap to be filled with a guess. It is held the same way, against the same intact skin, and it is a handheld optical instrument rather than an instrument that enters the eye. ISO 12865:2006 describes hand-held streak and spot retinoscopes, and ISO 10942:2022 does the same job for direct ophthalmoscopes; both are product standards, not cleaning standards, so they tell a buyer what the instrument is and leave the between-patient step to provincial guidance and the manufacturer.
| Part of the instrument | Level | When | What is used |
|---|---|---|---|
| Handle, body, forehead rest | clean + low-level disinfection | between patients | hospital-grade wipe at label contact time |
| Lens aperture and eyepiece | manufacturer's cleaning instruction | when soiled or smeared | soft swab, per the maker's instruction |
| Reusable soft pouch or case | clean | when soiled; not shared between lanes | detergent wipe |
| Instrument that touched the ocular surface | level changes | isolate and follow the maker's instruction | not the same step as the brow rest |
Which product, and how long does it stay wet?

Low-level disinfection is a product and a time, not a gesture. A hospital-grade disinfectant is applied and left wet for the contact time on its label; a wipe that flashes dry in fifteen seconds has cleaned the surface without disinfecting it. Warm water and detergent come first when there is visible soil, because a disinfectant applied over soil is a wasted step. ISO 17664-1:2021 is the reason the optics are handled separately: the manufacturer has to state how the device is cleaned and disinfected, and no clinic policy overrides that for the coated lens.
What must the record show?
Four fields survive an inspection: the instrument, the agent and its concentration, the contact time that actually elapsed, and the initials of the person who did it. A case of 1,300 double-tipped cotton swabs at the lane covers the optical cleaning step, and the log covers the rest.
Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.
A 160-wipe canister of hydrogen peroxide disinfecting wipes stays at the lane for the handle and the rest. See the family practice clinic supplies page for the rest of the room.
References
- Public Health Ontario, PIDAC, Environmental cleaning for prevention and control of infections in all health care settings, 3rd edition, 1st revision November 2025, Appendix 8 (Table 11), page 143 (checked 11 November 2026)
- ISO 17664-1:2021, Processing of health care products — Information to be provided by the medical device manufacturer for the processing of medical devices (checked 11 November 2026)
Printable companion: the free equipment cleaning log takes the four fields an inspector checks: instrument, agent and concentration, contact time and initials.
Restocking the lane? The clinic supply and cost planner converts patients per day into pads, swabs and wipe canisters.
Related Reading
- Comparing Three Eye Exam Instruments: Slit Lamp, Ophthalmoscope and Retinoscope
- Cleaning a Slit Lamp and Tonometer Between Patients: A Step-by-Step Clinic Protocol
- Ophthalmoscopes in Canada vs the US: Which Device Class Applies?
Frequently Asked Questions
Do retinoscopes and ophthalmoscopes need to be sterilized between patients?
No. They touch intact skin, which makes them noncritical devices, so cleaning plus low-level disinfection is the requirement and sterilization is not asked for.
Which disinfectant is used on the handle and forehead rest?
A hospital-grade low-level disinfectant applied at its label contact time; the lens, aperture and eyepiece follow the manufacturer's cleaning instructions rather than a hospital disinfectant.
Is the retinoscope in the provincial table?
The table lists the ophthalmoscope explicitly and not the retinoscope, so the retinoscope is handled as the same class of handheld noncritical instrument, with the same level and the same between-patient frequency.