How to Clean and Sterilize Mouth Mirrors and Periodontal Probes Between Patients in Quebec

Three rolls of autoclave indicator tape used to seal wrapped instrument packs and to carry the external process indicator

Quebec reprocesses a mouth mirror and a periodontal probe the way every province does — through cleaning, packaging, sterilization and monitoring — but the paperwork and the room around that loop are set by a different instrument. The Ordre des dentistes du Québec publishes guidelines on infection prevention and control and on medical device reprocessing, and the chapter that governs sterilization monitoring is the one most practices have had to revisit, because it changed the indicators, the frequency of biological tests, what is written on a package and what is entered in the registry, with a two-year implementation window attached to those changes.

This page follows the loop in the Quebec sequence, flags where the requirements differ from Ontario's, and names the documents a Quebec office should be able to produce on request. The equipment that carries the loop is an ultrasonic cleaner with a 7.5 L working tank and a Class 4 self-sealing sterilization pouch, both chosen against the volume the reprocessing unit actually turns over.

Quick facts for a Quebec reprocessing unit

  • The ODQ guidelines apply from publication, consolidate the 2009 infection control document, and give a two-year window for the specific changes listed in chapter 6.
  • Chapter 6 covers the use of chemical indicators, the frequency of biological tests, the information added to sterilization packages, the content of the reprocessing registry and the layout of the reprocessing unit.
  • Unidirectional flow is defined as a procedure to avoid recontaminating a device during decontamination; it is not a rule about which way staff walk, and one sink is sufficient.
  • There is no requirement about the number of doors in the unit, for existing units or for renovated ones, and returning backward is permitted for rinsing devices after ultrasonic cleaning.
  • A new or loaned sterilizer needs three consecutive full-load cycles with negative biological results before it is used.
  • The guidelines set no requirement to test washer-disinfectors or ultrasonic baths for efficacy, while noting that the CSA standard recommends doing so each day the equipment is used.

How does the Quebec loop run from the chair to the shelf?

The sequence is the familiar one: contaminated instruments leave the operatory in a closed container, are cleaned on the dirty side of the unit, rinsed and dried, inspected, assembled into sets, packaged, sterilized on the clean side, and stored so that sterility is maintained. What differs in Quebec is how the unit is arranged and what is written down. The guidance treats the unit as a space with two working zones and a one-directional procedure, where the rule to be satisfied is that a decontaminated device is never set down on a contaminated surface. A small practice can meet that with distance rather than with a wall, and a mobile or freestanding ultrasonic bath may need to be repositioned rather than replaced.

Storage is specified in one sentence worth memorizing: sterile devices must be kept in a closed, clean and dry place. The guidance allows sterile devices to be stored in the operatory, with the qualification that cabinets and drawers should not be opened during an aerosol-generating procedure, because the point of the closed storage is defeated by an open door at the wrong moment.

Stage What Quebec asks Practical consequence for mirrors and probes
Receiving Closed transport and a defined dirty side A dedicated container per tray keeps the mirror from travelling loose
Cleaning Automated bath acceptable; ultrasonic bath may be relocated to suit the layout The tank has to sit on the dirty side, even if that means moving a fixed unit
Rinsing Backward travel permitted after ultrasonic cleaning The rinse sink does not have to be on the clean side to be compliant
Packaging Chapter 6 adds information to be written on the package Pouches need a writable field that survives the cycle
Sterilization and registry Chapter 6 sets what the reprocessing registry records Every load produces a traceable entry naming the devices
Storage Closed, clean and dry Opened cabinets during aerosol procedures defeat the requirement

What does chapter 6 change about indicators and testing?

Two of the five chapter 6 items concern monitoring: the use of chemical indicators and the frequency of biological tests. The practical effect is that a Quebec office cannot inherit its indicator routine from an older protocol without checking the interval, and that the indicator inside each package is a requirement rather than a good habit. Quebec also permits an office to run its own biological tests, which the guidance recommends in light of laboratory verification services closing, together with a register of results.

The line that most surprises practices is the one about equipment tests. The guidelines set no requirement to test a washer-disinfector or an ultrasonic bath for efficacy, while noting that the CSA standard recommends testing each day the equipment is used. Ontario's standard, by contrast, asks for ultrasonic cleaners to be tested at least weekly and for automated washers to be tested per the manufacturer's instructions each day they are used. A practice that wants one routine for both provinces should keep the stricter calendar and document it.

How do Quebec and Ontario differ on the same loop?

The endpoints are the same — both instruments are sterilized — and the monitoring rhythm, the equipment checks and the records are where the two frameworks separate. A practice operating in both provinces, or a group with clinics on either side of the border, gains nothing from a single provincial compromise; the workable approach is one process, documented twice.

Item Quebec Ontario
Biological test frequency Set in chapter 6 of the ODQ guidelines, with a two-year implementation window Each day of use and for each type of cycle, inside a process challenge device
Chemical indicators Chapter 6 addresses their use; one belongs inside each package Type 4, 5 or 6 inside each package, with a Type 5 or 6 device for release decisions
Ultrasonic and washer tests No explicit requirement; the CSA standard recommends each day of use Automated washers per manufacturer; ultrasonic cleaners at least weekly
New or loaned sterilizer Three consecutive full-load cycles with negative results before service Monitoring and verification per the standard's requirements
Records Reprocessing registry, with an example register in the guidelines Sterilizer log book kept at least ten years
Facility layout Unidirectional procedure, one sink sufficient, no door count requirement One-directional workflow with separated dirty and clean areas and separate sections for each step

How does Canada differ from the United States on the same loop?

Against the federal guidance used in the United States the contrast is sharper on two rows. The CDC summary sets a weekly biological monitoring interval, and it allows a heat-sensitive semi-critical item to be processed with high-level disinfection at minimum, an option the Quebec and Ontario frameworks do not extend to an instrument that can be sterilized.

Item Quebec United States
Biological indicator interval Per chapter 6 of the ODQ guidelines At least weekly for each sterilizer
Heat-sensitive semi-critical item Sterilized; the item is bought so that it can be High-level disinfection acceptable at minimum
Who may run the test The office, with equipment and a register of results Office or contracted service, per facility policy and CDC guidance

Do Canada and the United States differ on what goes on the package?

Chapter 6 of the Quebec guidelines adds information to what a sterilization package must carry, and the reprocessing registry records the cycle that follows. The federal guidance used in the United States does not prescribe a package label field by field; it expects the facility to follow the manufacturer's instructions and its own policy, which is why an American pack can legally carry less written information than a Quebec one. A practice that operates in both jurisdictions should print the Quebec fields everywhere, because adding a date and a cycle number to a pack costs nothing and removing them later costs a relabelling exercise.

The same logic applies to the instrument file. The WHO material on medical devices and the instrument reprocessing guidance published in the United States converge on one point: the pack label and the cycle record are what make a sterilization claim checkable after the fact.

What goes on the package and in the registry?

Both are chapter 6 items, and both are record surfaces rather than clinical steps. The package carries the information the guidelines add to sterilization packs; the registry carries the trace of what was reprocessed, with an example form published in the guidelines. For a mirror and probe set the practical design is a pouch with a clear write-on field, filled in before the pouch enters the sterilizer, and a registry entry that names the set rather than the contents in general terms.

The instrument side of the record is the one people forget. A registry that lists "instruments, one load" cannot answer the question an inspection actually asks, which is which set went through which cycle and when. Labelling each set by operatory or by tray number costs a marker stroke and removes that ambiguity permanently.

What does the transport route add?

Offices that send instruments to an external sterilization service have a second set of rules to satisfy, covered by the provincial directives on external sterilization and the transport of medical devices. Those directives include a count sheet to reconcile devices sent and returned, which is the control that prevents a set from being quietly lost between two buildings. For mirrors and probes the reconciliation matters more than it does for a bulky cassette, because small items disappear from a count that is done by eye.

The workplace side is governed separately. Quebec's occupational health and safety material on biological hazards applies to the technician doing the cleaning, independently of the dental guidelines, and a unit that writes its procedure around the device rules alone usually misses the exposure control part of the file.

Three rolls of autoclave indicator tape used to seal wrapped instrument packs and to carry the external process indicator Self-sealing Class 4 sterilization pouch with steam and EO process indicators, shown empty before a set is loaded

What does the language requirement change in practice?

Quebec adds a labelling duty. Section 51 of the Charter of the French Language requires that inscriptions on a product, its container or its wrapping be drawn up in French, so a register, a package label or an instruction sheet printed by a Quebec office for its own use is a French-language document. Federal device labelling separately permits a single language for devices sold to health professionals, which is why imported pouches can arrive with English-only packaging and still be sold to clinics.

For the loop, the consequence is a small design constraint: a bilingual label needs more surface, and the write-on field competes with printed text. Offices that print their own set labels should choose a pouch one size up rather than compress the record.

Which consumables carry the Quebec loop?

Four lines do the work. Enzymatic solution is dosed per tank and changed daily; the multi-enzyme ultrasonic cleaning solution is concentrated for that routine. Pouches are sized to the set and need a writable field, which is why the Class 4 self-sealing pouch is the default in a unit that documents each package. Wrapped packs need tape that carries the external indicator, and the autoclave indicator tape covers both the seal and the indicator. Biological indicators follow the chapter 6 frequency rather than the patient count, and the 24-hour biological indicator pack suits an office testing in house.

The registry and count sheets that sit alongside the guidelines' example register are printable from the compliance log centre, which also carries the sterilization log generator for the unit wall. Offices standardising a single loop across several chairs will find the adjacent lines under supplies for dental clinics.

Rebuilding a reprocessing unit around chapter 6? Send the layout, the sterilizers, the cycle types and the registry you currently keep, and the answer will name the provision it rests on. Ask a Compliance Specialist.

Buying for several clinics or a facility account in Quebec? The institutional account enquiry page covers case quantities and account setup, and the institutional collection groups the lines stocked for facilities. Reselling? become a distributor.

Related reading

Sources and version notes

Frequently Asked Questions

Does Quebec require a separate room for reprocessing?

No. The guidance treats the unit as a space with a dirty side and a clean side and a procedure that prevents recontamination. One sink is sufficient, there is no requirement about the number of doors, and a mobile ultrasonic bath may need repositioning rather than replacement to satisfy the flow.

Is unidirectional flow about where staff walk?

No. The guidance defines it as a procedure to avoid recontaminating a device during decontamination, not as a rule about the operator's movement through the room. What matters is that a decontaminated device is never placed on a contaminated surface.

Can sterile instruments be stored in the operatory in Quebec?

Yes, if they are kept in a closed, clean and dry place. The guidance adds that cabinets and drawers should not be opened during an aerosol-generating procedure, because that exposes the sterile packs to contaminated air.

Do we have to test the ultrasonic bath every day?

The ODQ guidelines set no explicit requirement for washer-disinfector or ultrasonic bath efficacy testing, while noting that the CSA standard recommends testing each day the equipment is used. Many offices adopt the CSA calendar because it is also the stricter Ontario position.

What is required before using a loaned sterilizer?

Three consecutive cycles with a full load must give negative biological results before the unit is put into service. The same expectation applies to a newly purchased sterilizer, and the results belong in the reprocessing registry.

Where does the reprocessing registry example come from?

The ODQ guidelines publish an example register as an annex to the document. The registry records what was reprocessed and by which cycle, which is why a set label that identifies the tray is more useful than a generic entry.

Provincial guidance is cited as published at the time of writing and is not reproduced in full. Chapter 6 changes carry an implementation window, so confirm the version in force before relying on a clause. Cycle parameters come from device manufacturer instructions. CliniEco Medical distributes medical consumables under MDEL #35334.