A spore test log entry passes inspection when a reader who was not there can tell which sterilizer ran, which cycle, which lot of
biological indicators was used, what the control read, what the test read, and who signed it. Six fields, one line, every day the sterilizer is used. Most failed log reviews we see are not missing tests; they are missing the lot number or the control result, which makes the entry untraceable.


Which fields does the entry need?
| Field | Example | Why an inspector asks |
|---|---|---|
| Date and time | 2026-10-21, 08:10 | Proves the test ran on a day the unit was used |
| Sterilizer ID | Autoclave 2, service tag 4471 | Ties the result to one unit, not the room |
| Cycle type | Gravity steam, 134 C, 4 min | Each cycle type used that day needs its own test |
| BI lot and expiry | Lot 24J-118, exp 2027-03 | Traceability if the manufacturer recalls the lot |
| Control and test result | Control positive, test negative | Shows the indicator was viable |
| Operator and readout | Read at 24 h by R.N. initials | Readout time matches the format used |
Why is the control result not optional?
A negative test only means something if the same lot grew in the control. Without the control line, a failed indicator and a failed sterilizer look identical on paper, and the clinic cannot show that the test was capable of detecting growth. Record the control for every lot, on the first day that lot is used, and refer back to it on later entries rather than repeating the readout.
What is the frequency the log has to reflect?
Ontario's RCDSO standard requires a biological indicator for each sterilizer on each day it is used, and for each type of cycle used that day. That is daily monitoring, not weekly. The weekly baseline you will find in North American guidance is the CDC and AAMI ST79 position applied in most US states. A log built on a weekly schedule is internally consistent and still wrong for Ontario.
How do you write a positive result?
Write it as an event, not as a correction. Record the date and time the result was read, the unit taken out of service, the load recall decision, the corrective action (recalibration, repair, repeat test with a new lot), and the repeat result before the unit returned to service. Do not erase the original line, and do not rewrite the entry after the fact; an inspector who sees a clean page after a known failure reads it as a documentation problem rather than a resolved event.
Paper or spreadsheet?
Either is acceptable if it is contemporaneous, complete and retrievable. What fails is the reconstructed log: entries written at month end from memory, or a spreadsheet that shows a single bulk entry per week. If you use a spreadsheet, keep the file immutable after signing and export a monthly PDF, because the audit question is what the record looked like at the time, not what it looks like now.
Printable layer: the sterilization log center holds the printable daily spore test log, and the same page runs a log generator for clinics that record on screen.
Need a second opinion on your monitoring schedule? Ask a Compliance Specialist and bring your current log to the conversation.
Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers account setup and case formats, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.
Related reading
- Sterilization documentation for a dental office
- Log retention: Canada vs the US
- When records are lost to a computer failure
Frequently Asked Questions
Does an Ontario spore test log need the lot number?
Yes. Without the lot number the entry cannot be tied to a manufacturer recall, and a reviewer cannot tell whether the control result quoted belongs to the same batch of indicators.
Can the control result be recorded once per lot?
Yes, on the first day the lot is used, with later entries referring back to it. Repeating the control on every entry is not required and adds noise to the log.
What must be written when a biological indicator tests positive?
The read time, the unit taken out of service, the load recall decision, the corrective action and the repeat result before return to service. The original line stays in the record.
Is a spreadsheet log acceptable in Ontario?
Yes if it is contemporaneous, signed, retrievable and immutable after the fact. A month-end reconstruction of daily entries is a documentation failure even when the tests were run.
How long should sterilization records be kept?
Follow your provincial regulator and accreditation body, and keep the record at least as long as the sterilizer remains in service. When in doubt, keep the full service life of the unit plus one year.