Quick facts
- A clinical electrophoresis apparatus separates molecules or particles - plasma proteins, lipoproteins, enzymes and haemoglobins - on the basis of their net charge in a buffered medium.
- In the United States it is a named entry. 21 CFR 862.2485, Electrophoresis apparatus for clinical use, is Class I, product code JJN read from the openFDA classification database.
- Canada keeps no named entry. A device intended to examine specimens in vitro to aid a diagnosis is an in vitro diagnostic device, and Part 2 Rule 4 of Schedule 1 places it at Class II.
- Part 2 Rule 4 names six categories that lift an IVDD to Class III, including genetic testing and disease staging, both of which a stated haemoglobinopathy or paraprotein use can touch.
- Part 2 Rule 7 applies the higher class to both devices where the apparatus is intended to be used with another IVDD such as a reagent kit or a control.
- Sections, classes and product codes below were checked against the eCFR, the Regulations and the openFDA classification database on 2026-10-16.
An electrophoresis apparatus moves charged molecules through a buffered medium so that they separate into bands. The United States names the instrument and keeps it at the lowest class; Canada asks whether it is an in vitro diagnostic device and derives the class from the intended use. This article sets out both routes and the records that keep a gel comparable, with the 15 mL sterile centrifuge tubes that carry the samples.
What is a clinical electrophoresis apparatus?
It is a power supply, a chamber, a support medium and a buffer. A sample is loaded onto the medium, an electric field is applied, and each molecule travels at a speed set by its net charge and its size. When the run ends, the medium is stained and the pattern is read - by eye, by densitometer, or by a scanning system - and the fractions are reported as percentages.
The instrument is the load-bearing part of a larger method. The buffer pH, the run time, the voltage, the medium and the staining all change the pattern, which is why the apparatus matters less than the method it is used in. A laboratory that changes one of those variables without changing its validation has changed its result even though nothing about the box on the bench looks different.
The family is wider than the serum protein run that most people picture. Immunofixation, haemoglobin fractionation, lipoprotein analysis and enzyme isoform work all use the same physics. The ISO 15189:2022 framework treats the whole method - instrument, reagent, control and operator - as one examined system, which is the correct unit of comparison when a classification question is asked.
What class is an electrophoresis apparatus in the United States?
It is a named entry. 21 CFR 862.2485 is titled Electrophoresis apparatus for clinical use, and the identification describes a device intended to separate molecules or particles, including plasma proteins, lipoproteins, enzymes and haemoglobins, on the basis of their net charge in specified buffered media, used in conjunction with certain materials to measure a variety of analytes as an aid in the diagnosis and treatment of certain disorders. The regulation sets Class I (general controls).
The openFDA device classification API returns product code JJN against that section. The related instrumentation entry is 21 CFR 866.4500, Electrophoresis instrumentation, also Class I, with product code JZS - so both the clinical apparatus and the immunology instrumentation sit in the lowest class on the US side.
| Item | Detail |
|---|---|
| US regulation | 21 CFR 862.2485 - Electrophoresis apparatus for clinical use |
| US class | Class I (general controls) |
| Product code | JJN |
| Separations named in the entry | Plasma proteins, lipoproteins, enzymes, haemoglobins |
| Related entry | 21 CFR 866.4500 - Electrophoresis instrumentation, Class I, product code JZS |
| Premarket route | Exempt from premarket notification under subpart E of part 807, subject to section 862.9 |
| Read on | eCFR identification text and openFDA classification, both read 2026-10-16 |
How does Canada classify an electrophoresis apparatus?
Canada keeps no named entry, so the class is derived, and the derivation starts one step earlier than the US entry does. The question is whether the apparatus is an in vitro diagnostic device: a device intended to be used in vitro for the examination of specimens taken from the body. An apparatus that separates patient specimens so that a diagnosis can be aided fits that description, and the class then comes from Part 2 of Schedule 1 of the Medical Devices Regulations.
Part 2 Rule 4 places such a device at Class II unless one of six categories applies. Two of them are worth reading closely here. Rule 4(b) covers genetic testing, which is where a stated haemoglobinopathy use can land, and Rule 4(e) covers disease staging. Rule 4(d) covers a risk that an erroneous diagnostic result would cause death or severe disability. A general serum protein run does not touch any of them, so it stays at Class II; a stated use that does touch one moves the apparatus to Class III.
Part 2 Rule 7 is the other half of the picture. Where an IVDD, including its analyzers, reagents and software, is intended to be used with another IVDD, the class of both is the class of the higher-risk device. The apparatus and the reagent kit are therefore read together, not separately.
| Item | Canada | United States |
|---|---|---|
| Named entry | None; classified by rule | 862.2485 - Electrophoresis apparatus for clinical use |
| Class | Class II (Part 2 Rule 4, diagnostic IVDD) | Class I |
| Class III triggers | Six categories, including genetic testing, disease staging and risk of death or severe disability | Not provided for in the entry |
| Paired devices | Part 2 Rule 7 takes the higher class of the two | Not provided for in the entry |
| Near-patient use | Part 2 Rule 6 places a near patient IVDD at Class III | Not provided for in the entry |
| Product code | Not applicable | JJN; JZS on 21 CFR 866.4500 |
Why is the reagent part of the classification?
Because a separation without a stated interpretation is only a stain. The apparatus produces a pattern; the reagent kit and the control decide what the pattern means. Part 2 Rule 7 recognises that by applying the higher class of the pair to both, so a laboratory cannot lower its regulatory burden by buying the instrument from one supplier and the reagents from another.
The US side reaches the same place by a different road. The apparatus is a Class I entry with general controls, and the reagent, control and interpretation steps are regulated as separate IVDs. That is why a comparison built on the apparatus entry alone understates the Canadian position: the instrument is not the whole device, in either country, but only one of them says so inside the classification rule.
The practical consequence for a laboratory is that the validation file, the reagent certificates and the instrument log are one submission and not three. ISO 15189:2022 treats them that way, and the classification rules reward the same reading.
What should a laboratory check before ordering?
Four checks. First, confirm the intended-use statement and whether it names a category in Part 2 Rule 4, because that is what decides Class II or Class III in Canada. Second, confirm the separation types the maker supports and the reference materials for each, since a serum protein method and a haemoglobin method are validated separately. Third, confirm the reagents and controls the apparatus is intended to be used with, and their classes, because Part 2 Rule 7 applies the higher class to both. Fourth, confirm the sample handling and traceability chain: the container, the label, the transport condition and the run log.
| Check | What to ask for | Why it matters |
|---|---|---|
| Intended-use statement | Label text naming or excluding the Part 2 Rule 4 categories | Decides Class II or Class III in Canada |
| Separation types | Methods the maker validated and their reference materials | Each method is validated on its own |
| Reagents and controls | The paired IVDDs and their classes | Part 2 Rule 7 applies the higher class to both |
| Run conditions | Buffer, voltage, time and temperature | A changed variable is a changed method |
| Traceability | Sample identity from draw to report | A correct run on a misidentified sample is a wrong result |
| Records | Instrument log, run log and control results | The evidence an assessment asks for |
The bench that runs the gels uses the same consumables as the rest of the laboratory. The 15 mL sterile centrifuge tubes, the graduated 90 mL specimen cups and the 120 mL sterile screw-cap container come from the same ordering line.
A clinical laboratory also reorders the consumables used at the same separation bench. Clinics that stock for a whole practice can order through the wholesale programme and request a bulk quote for a standing or annual order; the 15 mL sterile centrifuge tubes sits in the same catalogue. Consumables for this line are listed under diagnostic laboratories. Reselling? become a distributor.
Sources
- openFDA device classification API
- Medical Devices Regulations, SOR/98-282, full text and Schedule 1
- US FDA, medical device databases
- US FDA, overview of device regulation
- Health Canada, medical devices
- Health Canada, establishment licensing
- ISO 15189:2022, medical laboratories, requirements for quality and competence
- ISO 17511:2020, metrological traceability of values assigned to calibrators
- ISO 13485:2016, medical devices quality management
- ISO 14971:2019, application of risk management to medical devices
- ISO 11137-1:2006, radiation sterilization of health care products
- ISO 20776-1:2019, susceptibility testing of infectious agents
- Clinical and Laboratory Standards Institute
- World Health Organization
- US CDC, laboratory quality
- Public Health Ontario, infection prevention and control
- Standards Council of Canada
Printable record sheet. A run log that records the buffer, the voltage and the control result is the only thing that makes two gels comparable. print the separation run log sheets — the pack carries a run log, a control result page and a reagent lot tracker; it prints from any browser on plain paper, and no account and no sign-up is required
Log generator. Enter the method, the buffer lot and your province, and the generator lays out a run sheet with those values fixed in the header. generate a separation run sheet — the generated page keeps the method, the lot numbers and the control line in the same place on every sheet, which is what makes a later comparison possible
Ask a compliance specialist. A classification question usually turns on how the device is used, not only on its name. Ask a compliance specialist — you get a written answer specific to your equipment and province, with the regulation or standard it is based on cited.
Related reading
- Clinical HPLC systems: class in Canada and the US
- Cytocentrifuges in Canada vs the US
- Microtomes: which class applies in each country
CliniEco Medical supplies eye care, laboratory and dental consumables of the kind described in this article. MDEL #35334. This article is written for optometry, laboratory and dental procurement professionals and is not legal advice; the class of a device and the records a practice keeps belong to the manufacturer and to the requirements that apply to it.
Frequently Asked Questions
What class is a clinical electrophoresis apparatus in the United States?
It is 21 CFR 862.2485 at Class I (general controls) with product code JJN, exempt from premarket notification under subpart E of part 807 subject to section 862.9.
What class is a clinical electrophoresis apparatus in Canada?
As an in vitro diagnostic device intended to be used in diagnosis or patient management it is Class II under Part 2 Rule 4 of Schedule 1 of the Medical Devices Regulations.
When does an electrophoresis apparatus reach Class III in Canada?
When its intended use falls into one of the six categories in Part 2 Rule 4, such as genetic testing, disease staging, or a risk that an erroneous diagnostic result would cause death or severe disability.
Is electrophoresis equipment for immunology in a different class?
In the United States, 21 CFR 866.4500 covers electrophoresis instrumentation at Class I with product code JZS. In Canada the class still comes from Part 2 of Schedule 1 and the stated intended use.
What else is a clinical electrophoresis apparatus called?
Electrophoresis unit, gel electrophoresis system, capillary electrophoresis instrument and protein electrophoresis analyser. The class follows the intended-use statement.
What records should the laboratory keep?
The instrument log, the run log with buffer and voltage, the reagent lot numbers, the control results and the sample traceability record. ISO 15189:2022 places all of these inside the laboratory quality system.