Sterile single-use lancets with a 33 gauge point, supplied 100 per box, for capillary blood sampling in glucose monitoring and point-of-care testing. Each lancet is used once and disposed of in a sharps container.

Gauge and depth both affect sampling: a thinner lancet is usually less painful but needs good capillary flow, so clinics with cold-hand or low-perfusion patients often keep a second size on hand.

Who it is for

  • Diabetes education and monitoring clinics
  • Family practices running point-of-care glucose tests
  • Long-term care homes and retirement residences
  • Pharmacies offering screening services

Specifications

Gauge 33G
Format Sterile, single use, 100 per box
Intended use Capillary blood sampling
Status Specification listing — availability pending supplier documentation

Frequently asked questions

What lancet gauge should a clinic stock?

Most clinics stock one mid-range gauge for routine use plus a smaller gauge for patients with reduced perfusion or when comfort is the priority. 33G sits at the fine end.

Are lancets reusable?

No. Lancets are single use for both infection control and skin-trauma reasons, and each one goes straight into a sharps container.

Do lancets need to be disposed of as sharps?

Yes. Lancets are a sharps stream even when the point is retracted into a housing, because the housing can be breached in waste handling.

How do I ask about volume supply?

Use our B2B enquiry form and we will confirm documentation status and availability.

What does 33G mean on a blood lancet?

Gauge describes the outside diameter of the point, not how deep it cuts. In the ISO 9626:2016 needle tubing series, 33G is the 0.20 mm designated size: one step finer than 32G at 0.23 mm, and the second-finest size in a series that stops at 34G (0.18 mm). A finer point parts less tissue, which is why fine gauges are associated with less discomfort at the puncture. Gauge alone does not set the sample, though, because the incision depth comes from the lancing device that holds and drives the lancet. A 33G lancet in a device set shallow can produce a smaller drop than a 30G lancet at a deeper setting, so clinics pair a fine gauge with an adjustable device instead of assuming that the finest lancet is the kindest one for every patient.

Why is a multi-patient lancing device not acceptable, even with a new lancet?

Health Canada addressed this directly in its notice of 26 September 2014 on licensing requirements for lancing devices and blood glucose monitoring systems (file 14-111696-560). The notice states that multi-patient use of lancing devices has been documented to pose a risk of disease transmission even if a new lancet is used for every patient. It also states that Health Canada will license lancing devices only if they are labelled as auto-disabling or for single-patient use, that point-of-care meters intended for multiple patients are licensed only with auto-disabling single-use lancing devices, and that the meter labelling must instruct health care personnel to change gloves between patients. The same notice records a 2012 Canadian incident in which a device designed for one person was used to obtain blood samples from 72 staff and students. A single-use lancet is therefore half of the control; the holder is the other half.

Which documents belong in a lancet specification file?

Sterility comes first: sterilisation validated to ISO 11135:2014 for ethylene oxide or to the ISO 11137 series for radiation, to a sterility assurance level of 10−6, with the sterile barrier system qualified to ISO 11607-1:2019 and a stated shelf life. Biological safety is assessed under ISO 10993-1:2018, and the manufacturer's quality system should hold ISO 13485:2016 certification. In Canada the device licence belongs to the manufacturer, while the importer or distributor holds a Medical Device Establishment Licence; CliniEco Medical distributes under MDEL #35334. For the glucose side of the workflow, Health Canada recognised ISO 15197:2013 for blood glucose monitoring systems for self-testing in the same 2014 notice, with a one-year transition for new devices. Marks belong to jurisdictions too: a CE mark is the manufacturer's declaration of conformity for the European market and is not evidence of Canadian licensing, so a Canadian file still needs the licence reference and the establishment licence number beside it. A specification file that holds those documents answers most of a procurement review before it starts.

How should used lancets be handled in a Canadian clinic?

A lancet is single use and goes into a sharps container at the point of use, point first, without recapping by hand. Waste rules are provincial rather than federal. In Ontario, guideline C-4, The Management of Biomedical Waste in Ontario, lists human blood waste in Table A with a yellow label and the universal biohazard symbol, and sharps go into a puncture-resistant, closeable container. A clinic in another province has to name its own rule instead of copying the Ontario one, and a policy written for sites in several provinces should cite the jurisdiction beside each requirement. Add the audit trail: container fill level, replacement date, and the person who closed and labelled it.

What should be checked when a box of lancets arrives?

Check the label against the requisition for gauge, box count and single-use wording; check that the sterile barrier is intact and the expiry date has not passed; confirm the carton has been stored dry and away from heat; and record the lot number on the receiving record so a supplier notice can be traced to a drawer. Keep one opened sample with the training file if the clinic trains new staff on lancing technique. Our blood glucose monitor kit, 1 cc insulin syringes and 1-quart sharps container are documented the same way, and volume requirements run through the wholesale enquiry desk.

33G blood lancet: what each specification changes
Specification As listed on this page What it changes at the bedside
Gauge 33G, 0.20 mm nominal outside diameter in the ISO 9626:2016 series Tissue trauma and discomfort at the puncture
Cutting depth Set by the lancing device, not by the lancet Drop size, and whether a second puncture is needed
Format Sterile, single use, 100 per box One lancet per patient, then straight into sharps
Device labelling Not stated on this page Auto-disabling or single-patient use, per Health Canada's 2014 notice
Sterilisation Not stated on this page ISO 11135:2014 or the ISO 11137 series, sterility assurance level 10−6
Sterile barrier Not stated on this page ISO 11607-1:2019 qualification and stated shelf life
Biocompatibility Not stated on this page ISO 10993-1:2018 evaluation for skin contact
Quality system — ISO 13485:2016 certificate held by the manufacturer
Distribution CliniEco Medical, MDEL #35334 Licence number recorded on the supplier file

CliniEco Medical supplies blood sampling consumables to Canadian clinics under MDEL #35334. This section is a specification reference; lancet and lancing device selection remains a clinical decision.

Related reading

Listing note: this page is published ahead of full catalogue onboarding so that specifications, documentation and sourcing details are available in one place. Pricing and ordering will be enabled once supplier documentation is complete. To be notified or to discuss volume requirements, use our B2B enquiry form.

Request a B2B Quote

Prefer a quote? B2B & facility pricing — bulk discounts and standing orders on request. We reply within 1 business day.

Frequently Asked Questions

What lancet gauge should a clinic stock?
Most clinics stock one mid-range gauge for routine use plus a smaller gauge for patients with reduced perfusion or when comfort is the priority. 33G sits at the fine end.
Are lancets reusable?
No. Lancets are single use for both infection control and skin-trauma reasons, and each one goes straight into a sharps container.