Ophthalmic Procedure Kits in Canada vs the US: Which Class Applies?

Clear graduated specimen cups stacked in a clean clinical setting

Quick facts

  • A phaco kit is the sterile single-use pack used during phacoemulsification: tubing, a cassette, a test chamber, sometimes a blade or a cannula, packed for one procedure.
  • In the United States the kit sits inside the system entry 21 CFR 886.4670 at Class II, with product code QDW for the phaco kit; the vitrectomy kit QDY sits under 886.4150, also Class II.
  • Canada has no kit entry. A sterile single-use pack that enters the eye or contacts the surface of the eye is an invasive device, and Rule 1(1) or Rule 2(1) places it in Class II.
  • The packaging is regulated separately from the kit: ISO 11607-1:2019 governs the sterile barrier system and the shelf life the maker claims.
  • The sterilization method the maker used decides which validation standard applies: ISO 11135:2014 for ethylene oxide, ISO 11137-1:2006 for radiation.
  • Sections, classes and product codes were read from the eCFR and the openFDA classification database on 2026-10-17.

An ophthalmic procedure kit is the pack opened at the start of a cataract or vitrectomy case: tubing, a cassette, a test chamber, and the small consumables the surgeon needs for that one eye. The United States files the phaco kit inside the phacofragmentation system entry at Class II; Canada has no kit entry and classifies the sterile single-use pack as an invasive Class II device. This article sets out both routes and the receiving checks a surgery centre runs on every lot, with the 20-pack occlusion patch set that dresses the eye at the end of the case, and the eye test chart set used when vision is checked at the first post-operative visit.

What goes into an ophthalmic procedure kit?

A phaco kit is built around the machine it feeds. The irrigation and aspiration tubing connects the handpiece to the console; the cassette or pack holds the fluid path; a test chamber lets the nurse verify that the line holds vacuum before the patient is draped. Some kits add a blade, a cannula, a syringe or a viscoelastic cannula, and some are configured for a specific console model.

A vitrectomy kit follows the same pattern for a different machine: cutter tubing, infusion line, and the small instruments the surgeon asks for by name. Because the two procedures use different consoles and different consumable sets, a centre that runs both keeps two kit families and two sets of par levels.

The kit is a packaging problem as much as a clinical one. Everything inside has to arrive sterile, stay sterile until the field is opened, and be presented in an order that lets the scrub nurse hand items without turning away from the field. That is why kit configuration changes slowly: the sequence of the wrap is part of the product.

How does the United States classify a phaco or vitrectomy kit?

21 CFR 886.4670 covers the phacofragmentation system, and the openFDA classification API returns QDW for Phaco Kit and MXO for a phacolysis laser system inside it. The vitrectomy kit, QDY, is returned under 21 CFR 886.4150. Both sit at Class II.

Item United States
Phaco kit 21 CFR 886.4670, Class II, product code QDW
Phacolysis laser system 21 CFR 886.4670, Class II, product code MXO
Vitrectomy kit 21 CFR 886.4150, Class II, product code QDY
Route to market Class II; the applicable premarket submission depends on the product code
Read on eCFR and openFDA classification, both read 2026-10-17

The structure is worth noting. The kit is not a standalone entry; it is a product code inside the entry for the system it serves. A US maker who changes the tubing length but keeps the same intended use stays inside the same code, and a maker who adds an active component may move to a different code inside the same section.

How does Canada classify the same kit?

Canada has no kit entry, so the pack is judged on what it does. A phaco kit enters the anterior chamber through a surgical incision, which makes it surgically invasive; Rule 1(1) of Schedule 1 places a surgically invasive device in Class II. A kit that only contacts the surface of the eye falls under Rule 2(1), which also gives Class II.

The subrules matter for the exceptions. Rule 1(2) takes a device intended to be used in direct contact with the heart, the central circulatory system or the central nervous system to Class IV, and Rule 1(3) takes an absorbable device, or one intended to remain in the body for at least 30 consecutive days, to Class III. A single-use kit used for the length of one case meets neither branch, so Class II is the answer.

Item Canada United States
Named entry None; classified by rule 886.4670 (phaco kit) / 886.4150 (vitrectomy kit)
Class Class II (Rule 1(1) or Rule 2(1)) Class II
Sterile single-use pack Judged as a device in its own right Product code inside the system entry
30-day or absorbable branch Rule 1(3) would lift to Class III; not met by a single-case kit Not provided for in the entry
Licence route Device licence for the Class II device Class II premarket route for the product code
Product code Not applicable QDW / QDY / MXO

Why does the sterile barrier system need its own file?

Because the kit's sterility is a claim about packaging, not about the contents. ISO 11607-1:2019 sets out the requirements for the sterile barrier system - the pouch, the header bag, the Tyvek lid, the tray - and for the shelf life the maker claims. A kit that is sterile on the day it is packed but loses its barrier in a humid store room is a kit that fails at the point of use.

The sterilization method then determines the validation route. Ethylene oxide follows ISO 11135:2014; radiation follows ISO 11137-1:2006. Either way the maker has to establish the dose or the cycle, validate it against a bioburden, and keep the records that show the routine process stayed inside the validated window. The clinic does not repeat that work, but the clinic does rely on it, and the documents that prove it are the ones to ask for when a hospital tender requires evidence.

There is a practical consequence for storage. A kit box that sits against an outside wall in a Canadian winter, or in a room that reaches 30 degrees in July, is a box whose barrier is being tested by the building rather than by the maker. Storage conditions on the label are part of the product specification, not a suggestion.

Clear graduated specimen container with a white screw-on lid, marked in ounces and millilitres

What should a surgery centre check on receipt?

Five checks on every lot. First, the pouch or header bag integrity, inspected before the box goes into stores rather than when the case is on the table. Second, the sterilization indicator on the label, which tells the receiver that the lot went through the process the label claims. Third, the expiry date, recorded against the lot so the oldest stock is used first. Fourth, the kit configuration against the console model, because a kit built for one machine does not fit another. Fifth, the lot number, kept with the receiving record so a field action can be traced to the cases that used the lot.

Check What to look for Record
Barrier integrity Seal width, punctures, moisture Receiving inspection line
Process indicator Indicator on the label has changed Lot record
Expiry Date readable, within shelf life Stock rotation line
Configuration Kit matches the console and the procedure Catalogue cross-reference
Lot number Present on the box and on the inner pouches Receiving record and case record

A kit opened and not used is the case that tests the system. Once the barrier is open the kit cannot go back into stock, and the item has to be discarded or, where the maker's instructions allow, reprocessed. Most makers do not allow reprocessing of a sterile single-use kit, which makes accurate par levels the lower-cost control.

How does a clinic keep the kit record auditable?

By tying the case record to the lot. The surgeon's note describes the procedure; the implant or consumable record carries the lot numbers; the receiving record carries the delivery date. When the three are linked, a field action can be answered in an hour rather than a week. When they are not, the clinic has to reconstruct the link from purchase orders and hope the dates line up.

The same discipline applies to the backup packaging. A centre that keeps a sterilization roll for the occasional item that has to be re-packed before a case is keeping a product whose own sterility has to be maintained, and ISO 11607-1:2019 applies to that packaging too. Whichever pouch or roll stock a centre keeps for that step belongs in the same stock rotation as the kits.

What should a buyer compare between suppliers?

Four dimensions. The kit configuration against the console in use, because a lower-priced kit that needs an adapter is not lower-cost. The sterilization method and the validation documents, because a hospital tender will ask for them. The shelf life claimed and the storage conditions on the label, because a short shelf life raises the cost of holding stock. The quality system certificate to ISO 13485:2016, which is the baseline document for a device of this class.

A fifth dimension sits outside the kit itself: the supplier's ability to hold par levels through a winter. A cataract list that runs four days a week consumes kits at a predictable rate, and a supplier who ships from a single warehouse with no buffer is a schedule risk that shows up as a cancelled case.

Rolls of paper medical tape stacked beside a single roll on a white surface

A surgery centre also reorders the backup packaging kept for a kit opened and not used. Clinics that stock for a whole practice can order through the wholesale programme for case pricing and account setup; resellers can become a distributor. The eye test chart set for the screening station sits in the same catalogue. The dressing side of the pack is where the CliniEco 20-pack occlusion patch set belongs.

Sources

  1. 21 CFR 886.1910, spectacle dissociation test system
  2. 21 CFR 886.4350, manual ophthalmic surgical instruments
  3. 21 CFR 886.4670, phacofragmentation system
  4. openFDA device classification API
  5. Medical Devices Regulations, SOR/98-282, full text and Schedule 1
  6. US FDA, medical device databases
  7. Health Canada, medical devices
  8. Health Canada, establishment licensing
  9. Standards Council of Canada
  10. CSA Group
  11. ISO 15004-1:2020, ophthalmic instruments, general requirements
  12. ISO 15004-2:2007, ophthalmic instruments, light hazard protection
  13. ISO 17664-1:2021, processing of health care products
  14. ISO 11607-1:2019, packaging for terminally sterilized medical devices
  15. ISO 11135:2014, ethylene oxide sterilization
  16. ISO 13485:2016, medical devices quality management
  17. ISO 14971:2019, application of risk management to medical devices
  18. Canadian Association of Optometrists
  19. College of Optometrists of Ontario, standards of practice
  20. Canadian Ophthalmological Society

Printable record sheet. A kit lot is only traceable if the receiving sheet and the case sheet agree. download the kit receiving and traceability sheet — the pack carries a receiving line, a lot column and a case-link box; it prints on plain paper with no account and no sign-up

Log generator. Enter the kit family, the console and your province and the page produces a dated reconciliation sheet. build a kit traceability sheet — it lines up the lot number, the expiry and the case reference in the order a recall search is run

Ask a compliance specialist. A classification question usually turns on how the device is used, not only on its name. Ask a compliance specialist — you get a written answer specific to your equipment and province, with the regulation or standard it is based on cited.

Related reading

CliniEco Medical supplies eye care, laboratory and dental consumables of the kind described in this article. MDEL #35334. This article is written for optometry, laboratory and dental procurement professionals and is not legal advice; the class of a device and the records a practice keeps belong to the manufacturer and to the requirements that apply to it.

Frequently Asked Questions

What class is a phaco kit in the United States?

It is a product code inside 21 CFR 886.4670, the phacofragmentation system entry, at Class II. The product code is QDW. A vitrectomy kit, QDY, sits under 21 CFR 886.4150, also Class II.

What class is an ophthalmic procedure kit in Canada?

Canada has no kit entry. A sterile single-use kit that enters the eye is a surgically invasive device classified as Class II under Rule 1(1); a kit that only contacts the surface of the eye is Class II under Rule 2(1).

Does a kit ever reach Class III in Canada?

Yes, through Rule 1(3), where the device is intended to be absorbed by the body or to remain in the body for at least 30 consecutive days. A single-use kit used for one case does not meet that branch.

Which packaging standard applies to a sterile kit?

ISO 11607-1:2019 sets the requirements for the sterile barrier system and for the shelf life the maker claims. The sterilization method then decides the validation standard, ISO 11135:2014 for ethylene oxide or ISO 11137-1:2006 for radiation.

What else is a phaco kit called?

Phaco pack, phacoemulsification kit, cataract procedure pack, irrigation and aspiration pack, and custom procedure tray. The regulatory question follows the intended use and the configuration, not the trade name.

What should be recorded on receipt of a kit lot?

Barrier integrity, the process indicator, the expiry date, the configuration against the console, and the lot number. The lot number should also appear in the case record so a field action can be traced.