A Quebec dental office reprocesses a Gracey curette in the same order as an office in Ontario or British Columbia: chairside treatment, transport, cleaning, rinsing, drying, inspection, packaging, sterilization, storage. What differs is the document the office is measured against, and in Quebec that document is the joint guidance published by the Ordre des dentistes du Québec and the Ordre des hygiénistes dentaires du Québec on infection prevention and control and the reprocessing of medical devices. Its chapter 6 is written as a sequence of numbered protocols, which makes it unusually easy to audit — and unusually easy to fall behind on.
The guidance came into force when it was published, in October 2024 with a February 2025 revision, and it replaced the 2009 infection control information document while giving practices a two-year application period for the specific changes listed in chapter 6. Those changes concern the use of chemical indicators, the frequency of biological tests, the information added to sterilization packages, the content of the reprocessing register and the layout of the reprocessing unit. For a hygiene operatory the practical question is which of those five touches a curette or a piezo tip, and the answer is four of the five.
The equipment side of the loop is the same in Quebec as elsewhere: an
ultrasonic cleaner with a 7.5 L nominal tank for the mechanical cleaning step — with the caveat below that a mobile bath may have to be moved into the reprocessing unit — and a low-foam multi-enzyme ultrasonic solution dosed for it.
What does the Quebec framework add that a generic protocol does not cover?
Three things, and all three are record or layout questions rather than clinical ones. First, the reprocessing unit — the URDM — must not sit in a room where professional activities such as consultations, care, treatment or services are carried out; existing units can be rearranged rather than rebuilt, and a mobile or non-built-in ultrasonic bath may have to be relocated rather than replaced. Second, the reprocessing register is a required record with an example form in Annexe 4 of the guidelines, not a log book kept out of habit. Third, the guidelines allow an office to run its own biological tests, and recommend buying the equipment and keeping a register of results, because the laboratories that used to verify sterilizers have been closing.
There is also a language duty that is easy to miss. Section 51 of the Charter of the French Language requires that inscriptions on a product, its container or its wrapping be drawn up in French, so a package label, a register form or a procedure sheet printed by a Quebec office for its own use is a French-language document. Federal device labelling separately allows a single language for devices sold to health professionals, which is why a box can arrive in English and the office's own paperwork still has to be French.
| Chapter 6 section | What it covers | What it changes for a curette or a scaling tip |
|---|---|---|
| 6.1 Unit layout | URDM arrangement, unidirectional procedure | The bath has to sit on the dirty side; the rule is about surfaces, not about which way staff walk |
| 6.2 Sterilizers and monitoring | Physical parameters, biological and chemical indicators, process challenge devices | Indicator inside every package; a release decision that depends on what the load record shows |
| 6.3.1 Pre-cleaning, brief inspection, transport | Chairside treatment and movement of the device | Curettes leave the operatory in a closed container, not loose in a tray |
| 6.3.2 and 6.3.3 Manual and mechanical cleaning | Receiving, manual cleaning, mechanical cleaning | Brush work on the blade and knurling, then the bath in a basket so tips cannot touch |
| 6.3.5 Packaging | Packaging of the device before sterilization | Pouch sized so a sharp toe sits against film, and the chapter 6 package information added |
| 6.3.6 and 6.3.7 Sterilization and storage | Cycle, then storage of sterile devices | Closed, clean and dry storage; operatory cabinets closed during aerosol-generating procedures |
| 6.4 and 6.5 Failure and recall | What happens when a cycle fails | A failed load containing hygiene instruments has to be traced to the patients who received them |
| 6.6 Unpackaged instruments | Protocol for instruments sterilized without packaging | A short set is not an excuse; the practice needs the written protocol before it is used |


How does the chapter 6 loop run, section by section?
Read as a sequence, chapter 6 is the same ten steps every reprocessing room performs, with a decision attached to each. The pre-cleaning and transport step sets the condition of everything that follows: a curette that reaches the unit with dried blood in the shank junction will spend longer in manual cleaning and may still need a second pass. Manual cleaning under running water with a soft brush is where the blade, the knurled handle and the tip threads are dealt with, and it is the step the guidelines place before mechanical cleaning rather than in parallel with it.
Mechanical cleaning is where the bath configuration matters. Keep instruments in a perforated basket so that tips cannot vibrate against each other or against the tank, use the dilution, time and temperature on the solution label, and treat the bath as equipment that needs its own daily check — the standard behind the provincial guidance expects a commercial cleaning-verification method on each day of use and does not accept a foil test as validation, while the Quebec guidance itself sets no explicit efficiency test for the bath and notes the standard's daily recommendation. A practice that follows the stricter line has one routine to defend in both directions.
Lubrication, section 6.3.4, is the step most hygiene instruments skip: curettes and piezo tips have no hinge and no ratchet, so a lubricant that is appropriate for forceps is not appropriate here, and the manufacturer's instruction sheet decides whether anything is applied at all. Packaging then follows 6.3.5, with the chapter 6 information added to the package, and the cycle follows 6.3.6. Storage is one sentence worth memorising: sterile devices are kept in a closed, clean and dry place, and when they are stored in the operatory the cabinets and drawers are not opened during a treatment that generates aerosols.
What does a Quebec inspector read on the register?
Annexe 4 of the guidelines gives an example register and states the information that has to appear in it. The register is what turns a claim about a load into a traceable record, and for a hygiene set the failure mode is a line that says "instruments, one load" — which cannot answer the question an inspection actually asks, namely which set went through which cycle and when. Labelling each set by operatory or tray number costs a marker stroke and removes that ambiguity permanently.
| Decision | Quebec position | Ontario position |
|---|---|---|
| Where the reprocessing unit sits | Not in a room used for professional activities; existing units may be rearranged; a partition can separate dirty and clean areas | One-directional workflow with separated dirty and clean areas, and separate sections for each step |
| Unidirectional flow | A procedure to avoid recontaminating a device, not a rule about staff movement; one sink is sufficient and backward travel is permitted for rinsing after ultrasonic cleaning | One-directional workflow through distinct areas |
| Unpackaged cycles | A written protocol, section 6.6, before the practice uses them | Reserved for urgent situations where no other option is available, and not a substitute for instrument stock |
| Biological testing | Frequency set in chapter 6; an office may run its own tests, with a register of results | Each day of use and for each type of cycle, inside a process challenge device |
| Package and register language | French required for the office's own inscriptions under section 51 of the Charter of the French Language | No equivalent language requirement for package labels |
Where does the curette itself still catch people out?
The instrument-level traps do not change with the province. The cutting third of a Gracey blade holds calculus that a flat brush stroke slides over, the knurling on the handle holds dried blood, and the threaded joint of a piezo tip holds debris that is invisible while the tip is seated on its insert. Sharpening is the step that creates new soil: if the office sharpens between cleaning and packaging, the instrument has to be re-cleaned afterwards, because metal particles are soil and not dust. And the pouch has to be sized so the working end sits against the clear film, since a sharp toe pressing on paper is a puncture waiting for the next person who picks the package up.
The one Quebec-specific consequence is paperwork rather than technique. A failed cycle is not only a re-run: under section 6.4 and 6.5, a failed load has to be handled through the failure and recall protocols, which means the register has to be good enough to identify which patients received instruments from which load. The register is therefore written before the sterilizer door closes, not after the result is known.
Where does a clinic take this next?
Four things a clinic can act on rather than read: a printable compliance log sheet that records receiving dates, lot numbers and open-on dates and exports from the browser; the clinic supply planner that turns daily bench counts into cases per month; and, where a finding has to be documented for an inspector, Ask your compliance question which returns a written answer citing the regulation or standard it rests on.
Facility and multi-site buyers can also ask about wholesale and multi-site ordering for case pricing and account setup. Reselling? become a distributor.
Related reading
- How to clean and sterilize mouth mirrors and periodontal probes between patients in Quebec
- How to clean and sterilize Gracey curettes and ultrasonic scaling tips between patients
- What an infection control auditor checks about Gracey curettes and ultrasonic scaling tips
- How many Gracey curettes and ultrasonic scaling tips a clinic needs per operatory
Frequently Asked Questions
Does a Quebec office need a separate room for the reprocessing unit?
The guidance requires that the reprocessing unit not be located in a room where professional activities such as consultations, care, treatment or services are carried out. For an existing unit a rearrangement is sufficient rather than a rebuild, and where a sterilizer is not built into the cabinetry a partition such as a plexiglas screen may be used to separate the soiled area from the clean area.
Can a Quebec dental office run its own biological indicator tests?
Yes. The orders allow the office to carry out the verification itself, and the guidance recommends acquiring the equipment needed to run biological tests and creating a register in which results are recorded, because laboratories that offered sterilizer verification services have been closing. Annexe 4 of the guidelines gives an example of the register and states what has to appear in it.
What has to be written on a sterilization package in Quebec?
Chapter 6 adds information that has to appear on packages intended for sterilization, and the reprocessing register records the cycle that follows. Because section 51 of the Charter of the French Language requires inscriptions on a product, its container or its wrapping to be in French, a package label or register form printed by a Quebec office for its own use is a French-language document.
Is one sink enough in a Quebec reprocessing unit, and can staff walk backwards?
Yes to both in the sense the guidance uses the terms. Unidirectional flow is defined there as a procedure that avoids recontaminating a device during decontamination, not as a rule about the direction staff move; a single sink is sufficient, and returning in the other direction is permitted for rinsing devices after cleaning in the ultrasonic bath. What the rule does forbid is setting a decontaminated device down on a contaminated surface.
How long is a pouched curette considered sterile in Quebec?
The guidance treats storage as a condition rather than a countdown: sterile devices have to be kept in a closed, clean and dry place, and where they are stored in the operatory the cabinets and drawers should not be opened during a treatment that generates aerosols. The practice's own written procedure sets rotation, and the audit test is the condition of the pouch and the completeness of the register.