A screening clinic runs thirty otoacoustic emission tests before lunch. The tip comes off, the probe nose gets a wipe, and the next patient sits down. The wipe is the visible step; the decisions that make the screening result trustworthy sit either side of it.
Start with the hands. Nitrile examination gloves, 3.5 mil, 1,000 per case go on before the tip is removed, because the tip is the part that has been in a canal and the probe is the part that goes back in the drawer.
How are otoacoustic emission probes sanitised between patients in Canada?
Four steps, in order. Remove the ear tip and discard it, or set it aside for cleaning if it is reusable. Wipe the probe nose with a disinfectant the probe maker lists as compatible. Let it dry before the next ear, since a wet nose shifts the seal and the stimulus level. Then confirm the probe still passes its check cavity before the session continues. The probe body does not enter the canal; the tip and the sound tube do. That split is what makes the job a wipe rather than a soak, and it is also where clinics over-clean.
Does a wipe reach the sound tube?

No. The sound tube sits inside the probe nose, and cerumen drawn into it attenuates the stimulus. A probe with a partly blocked tube still produces a waveform, so the software returns a refer on an ear with normal cochlear function, and the clinic starts booking follow-ups that should not exist. A monthly check finds this late. Most makers supply a cleaning wire or a replacement nose, and the check cavity tells you which one you need.
Is an OAE probe semicritical?
Classify by contact rather than by reputation. The ear tip touches intact canal skin, which puts it in the non-critical group: cleaning plus low-level disinfection. A reusable tip that presses against the canal wall follows the maker's validated processing information under ISO 17664-1, and in Canada the reprocessing expectations in CSA Z314:23 cover it as reusable equipment. The probe nose does not enter the canal, so it does not become semicritical through proximity.
What goes on the between-patient record?

| Item | Contact | Step | Record |
|---|---|---|---|
| Probe nose | outside the canal | wipe with listed disinfectant, dry | time, product, initials |
| Single-use ear tip | canal skin | discard after one patient | date the box was opened |
| Reusable tip | canal skin | clean and disinfect per maker | cycle log |
| Check cavity | none | daily function check | date, pass or fail |
What about the tips and the booth?
An open box of tips on a bench is the usual contamination route, so tips stay in the box until the patient is seated. A booth shared with audiometry needs the same between-patient attention as the probe: a fresh length of perforated exam table paper, 21 inch on the ledge, and dry disposable washcloths for hands and headphone cushions that a wipe would leave damp.
Running an audiology or diagnostic clinic? See our diagnostic laboratory and clinic supplies.
Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers. Reselling? become a distributor.
Not sure how this applies to your own setup? Ask a sterilization compliance specialist for a written answer specific to your equipment and province, with the regulation or standard cited.
References
- Public Health Ontario — provincial guidance on cleaning, disinfection and sterilization of medical equipment in health care settings (checked 8 November 2026)
- ISO 17664-1:2021, Processing of health care products — information to be provided by the medical device manufacturer (checked 8 November 2026)
Printable companion: the free clinic log and checklist library includes a between-patient probe sheet with product, contact time and initials already ruled.
Booking screens all day? The clinic supply and cost planner turns a weekly test count into boxes of tips and canisters of wipes per quarter.
Related Reading
- Audiology Clinics: Between-Patient Hygiene for Hearing Test Rooms
- What Instrument and Waste Rules Apply to Cerumen Management in Canadian Audiology?
- What Consumables Does an Occupational Hearing Testing Program Use Per Canadian Worker?
Frequently Asked Questions
Should an OAE probe be wiped with alcohol between patients?
Only if the probe maker lists alcohol as compatible. Alcohol clouds some polycarbonate housings and dries silicone couplers, so the maker's list decides, not habit.
How often should the probe be checked in the test cavity?
At the start of each testing day, and after a probe is dropped or the nose is changed. The check takes seconds and it is the only warning before results start drifting.
Can a single probe be used on both ears of the same patient?
The tip is the contact item. A new tip per ear is the usual practice, and a tip that has been in one canal is not moved to the other without cleaning.
Does the probe have to be sterilized in Canada?
No. It does not enter the canal, so cleaning and low-level disinfection cover the body; a reusable tip that presses against canal skin follows the maker's validated processing instructions.
Last updated: October 2026. CliniEco Medical is a licensed medical device establishment (MDEL #35334).