Case Review: Rebuilding Sterilization Records After a Failed Bowie-Dick Test

A pre-vacuum sterilizer that fails a Bowie-Dick test does not always have a broken sterilizer behind it. Air removal failures come from worn door gaskets, a drain strainer that missed its cleaning, wet steam, or a chamber loaded so that air was trapped around the pack. Separating those causes is a records problem before it is a service-call problem.

This review is de-identified: a Canadian dental clinic read a fail on the first cycle of the day, with a load already sitting in the chamber.

What does a failed Bowie-Dick test actually tell you?

A Bowie-Dick test pack measures air removal and steam penetration in a pre-vacuum cycle. It is a cycle test, not a load sterility test. A fail means the vacuum stage did not clear air well enough for steam to reach the centre of the pack. ISO 11140-5 defines the Type 2 indicator class used for these daily vacuum checks, read against a defined pass pattern rather than a personal impression of the colour change.

Keeping one pack format in stock, such as a six-pack of ISO 11140-5 Type 2 Bowie-Dick test packs, gives a steadier baseline across the month.

Is it a cycle failure or a sterilizer failure?

Treat the first fail as a cycle failure. The two paths carry different records and different release decisions. A cycle failure is handled inside the clinic: hold the load, correct the obvious cause, re-run the pack, document the result. A suspected sterilizer failure takes the unit out of service until it is repaired and requalified. Writing "sterilizer broken" at the first fail closes the file too early.

Stage What the record must show Who confirms it Release decision
Fail detected Unit ID, date, cycle type, pack lot, result, time read Person who ran the cycle Stop; hold the load
Quarantine Which load is held, where, who was notified Clinical lead Hold until a pass is documented
Fault finding Gasket, drain strainer, steam quality, load pattern Person doing the check Correct the cause, then retest
Retest Repeat result, cycle type, readings Second operator Release only after a pass
Closure Who signed the review, where the unit went next Practice owner Return to the daily routine

Which loads are affected when a Bowie-Dick test fails?

Every load processed since the last passing test is in scope, and anything not already released should be held. The morning load in this case never left the bench. Where a load was already released, the decision turns on the biological indicator records for the same period. The Bowie-Dick and the biological indicator catch different failures, which is why both belong in the same log.

Bowie-Dick test pack and autoclave log sheet after a failed vacuum check in a dental clinic

How often does a sterilizer need monitoring in Ontario?

Bowie-Dick packs are a daily vacuum check on pre-vacuum units. In Ontario, the RCDSO requires a biological indicator for each sterilizer on every day it is used. The weekly baseline belongs to the CDC and AAMI ST79 reference followed in most US states; it is not the Ontario rule. Reading one frequency from the other framework is a common documentation gap; the sterilization compliance hub sets out the records a review works through.

What should the release entry look like afterwards?

After the retest passes, one entry should carry the unit, the cycle type, the date and result of the failed test, the corrective action taken, and the result of the retest. A reader should be able to follow that sequence without asking anyone what happened. Keep the failed pack and photograph it beside the passing retest: the pass pattern explains more than a paragraph.

→ Start with a documented pass: CliniEco 24-hour Self-Contained Biological Indicator — 5-Pack Trial ($12.99, shipping included).

Ordering for a clinic, lab or care home? Wholesale and multi-site ordering covers case pricing and account setup, and the B2B wholesale collection lists the lines stocked for institutional buyers.

Related reading

Frequently Asked Questions

How long should a load stay quarantined after a failed Bowie-Dick test?

Until a repeat test on the same sterilizer records a clear pass and that pass is written into the log. It needs a documented passing cycle and a named person signing the release.

Should we run a biological indicator when a Bowie-Dick test fails?

Ontario practice is a biological indicator for each sterilizer on every day it is used, so the daily entry continues. What changes after a fail is the release decision: the biological indicator record supports releasing a load when no vacuum test covers that period.

Should the failed test pack be kept?

Yes. Keep the failed pack and a photograph of its pass pattern beside the passing retest. The colour distribution answers a reviewer faster than a written note, and it shows a service provider what the unit was doing before the call.

What is the most common record gap after a failed vacuum test?

A missing corrective action column. Clinics log the fail and the later pass but never record what changed in between, so the sequence reads as an unexplained recovery.