What a Canadian Maker Must Satisfy Before Selling Devices in the United States: 21 CFR 880 and 21 CFR 820 Explained

Pair of 4.0 mil blue nitrile examination gloves on a white background

If you manufacture a device in Canada and want to sell it in the United States, two parts of the Code of Federal Regulations decide most of your work before you ship anything: 21 CFR 880, which classifies general hospital and personal use devices and says whether premarket notification applies, and 21 CFR 820, the quality management system regulation your facility has to run. Alongside those sit the registration and listing rules in 21 CFR 807, which require a foreign establishment to register, list its devices, and name exactly one United States agent — and which make the first importer of your product a regulated establishment in its own right. This guide walks the four steps in order, with the section numbers and the fee structure that apply to the current fiscal year.

Start with the classification question, because it decides everything else. A product such as a CliniEco 24-hour self-contained biological indicator — 50-pack is regulated under 21 CFR 880.2800, while disposable nitrile examination gloves, 3.5 mil, 1,000 per case sit under 21 CFR 880.6250 — two different sections, two different classes, and two different answers on premarket notification.

Which part of the US rulebook applies to what?

Instrument What it governs Who it binds Where to read it
21 CFR Part 880 Classification, identification and premarket notification exemptions for general hospital and personal use devices: sterilization process indicators, sterilization wrap, spore test strips, steam sterilizers, examination gloves, examination gowns and more Manufacturers and, through the exemptions, distributors and repackagers 21 CFR Part 880
21 CFR Part 878 Surgical and dental devices, including surgical apparel such as surgical masks and surgical gowns Manufacturers of those devices 21 CFR 878.4040
21 CFR Part 807 Establishment registration, device listing, the United States agent requirement for foreign establishments, and premarket notification under section 510(k) Foreign manufacturers, initial importers, repackagers and specification developers 21 CFR Part 807
21 CFR Part 820 The quality management system regulation: what the quality system must contain, which records must be kept, and which devices need design and development controls Manufacturers, contract sterilizers, relabellers, repackers, specification developers and initial distributors of foreign entities that perform those functions 21 CFR 820.10
Federal Food, Drug, and Cosmetic Act, section 738 Annual establishment registration fees and premarket submission fees Every registered establishment and every fee-paying submitter FY 2027 fee notice

Note how Part 880 describes itself. The identification of a device in a regulation "is not a precise description of every device that is, or will be, subject to the regulation," and a manufacturer filing a premarket notification "may not show merely that the device is accurately described by the section title" — it has to argue substantial equivalence. The class name on a competitor's label is not a classification decision.

Does the device need a 510(k) submission?

The table below maps the categories a Canadian sterilization and consumable supplier typically ships, against the section that names them. Confirm the product code for your specific device before relying on any row.

Category Section Class Premarket notification
Biological sterilization process indicator 21 CFR 880.2800(a) Class II (performance standards) The section carries no exemption from premarket notification
Physical or chemical sterilization process indicator 21 CFR 880.2800(b) Class II (performance standards) No exemption stated in the section
Spore test strip for a liquid chemical sterilant processing system 21 CFR 880.6887 Class II (special controls) Class II with named special controls for characterization, simulated use testing and labelling
Sterilization wrap, pack, bag or accessories 21 CFR 880.6850 Class II (performance standards) No exemption stated in the section
Steam sterilizer 21 CFR 880.6880 Class II (performance standards) No exemption stated in the section
Non-powdered patient examination glove 21 CFR 880.6250 Class I (general controls) Exempt from premarket notification when it is a finger cot, subject to 21 CFR 880.9
Examination gown 21 CFR 880.6265 Class I (general controls) Exempt subject to 21 CFR 880.9; if not represented as sterile, also exempt from the quality system requirements except records and complaint files under 21 CFR 820.35
Surgical mask and surgical gown 21 CFR 878.4040(b)(1) Class II (special controls) The section sets special controls and describes the conditions under which a surgical N95 respirator is exempt

Two mechanics follow from the table. First, a class II device needs premarket notification unless the regulation for that generic type grants an exemption, and many sections in Part 880 grant none — so the classification alone can commit you to a submission. Second, an exemption is not unconditional. Under 21 CFR 880.9, a manufacturer of an exempt class I or class II device must still submit a premarket notification when the device is intended for a use different from the intended use of a legally marketed device in that generic type, or when the modified device operates using a different fundamental scientific technology.

When premarket notification is required, the timing is fixed: under 21 CFR 807.81, the submission must reach the agency at least 90 days before the device is introduced or delivered for introduction into commercial distribution. The triggers include introducing a device that is not substantially equivalent to a device in commercial distribution before May 28, 1976, introducing a device for the first time by a person required to register, and a change or modification that could significantly affect safety or effectiveness. Narrow exemptions exist — a custom device, and a distributor or repackager placing a device into distribution under its own name where a notification was filed by another person (21 CFR 807.85) — but they are exceptions, not defaults.

If the classification itself is genuinely unclear, section 513(g) of the Act provides a route to ask the agency for a classification determination, and the fee notice lists that request as a fee-bearing submission.

What does 21 CFR 820 require of the manufacturer and its suppliers?

Part 820 is no longer the quality system regulation as it was written in the 1990s. It is now titled the quality management system regulation, and its requirements are built on the international standard:

  • A manufacturer must document a quality management system that complies with the applicable requirements of ISO 13485 and other applicable requirements of the part (21 CFR 820.10(a)).
  • ISO 13485:2016 is incorporated by reference and approved for several sections of Part 820, and ISO 9000:2015 supplies the definitions used in the part (21 CFR 820.7).
  • Specific regulatory requirements are attached to named ISO 13485 clauses: unique device identification under Part 830, traceability procedures under Part 821 where applicable, complaint reporting to the agency under Part 803, and advisory notices under Part 806 (21 CFR 820.10(b)).
  • Design and development controls under ISO 13485 clause 7.3 apply to class II and class III devices, and to the class I devices listed in the regulation (21 CFR 820.10(c)).
  • Failure to comply with an applicable requirement renders the device adulterated under section 501(h) of the Act (21 CFR 820.10(e)).

Scope is broader than a Canadian manufacturer might expect. The applicability provision names contract sterilization, installation, relabelling, remanufacturing, repacking and specification development among the activities that pull a party into the regulation, and it reaches devices imported or offered for import into the United States. A Canadian facility that sterilizes for a United States specification developer is therefore inside the scope of Part 820, not outside it.

A pair of 4.0 mil blue nitrile examination gloves photographed on a white background for a product listing
A patient examination glove is class I under 21 CFR 880.6250, but the premarket notification exemption in that section is limited to the finger cot form.

Who has to register, and who is the initial importer?

Registration is not optional and it is not annual paperwork you delegate to a broker without checking. Under 21 CFR 807.20, an owner or operator of an establishment engaged in the manufacture, preparation, propagation, compounding, assembly or processing of a device must register and submit listing information — and the definition of that activity explicitly includes the initial importation of devices manufactured in foreign establishments (21 CFR 807.3(d)). That is the hinge on which the initial importer duty turns.

Role in the chain Registration and listing duty United States agent Cite
Canadian manufacturer or specification developer Must register and list; must designate exactly one United States agent Required 21 CFR 807.20, 21 CFR 807.40
Initial importer (imports the finished device and furthers its marketing without repackaging or relabelling) Engaged in a regulated activity by virtue of initial importation, so registration and listing apply Not applicable — the foreign establishment names the agent 21 CFR 807.3(d), 21 CFR 807.3(g)
Wholesale distributor that does not manufacture, repackage, process or relabel Not required to register Not applicable 21 CFR 807.20(c)
Pharmacy or retail outlet making final delivery to the end user Exempt from registration Not applicable 21 CFR 807.65

For a foreign establishment the rules are specific. Under 21 CFR 807.40, the establishment must register and list using the electronic system, must submit the name, address and phone number of its United States agent as part of its registration information, and may designate only one agent — typically also its official correspondent. The agent must reside or maintain a place of business in the United States and must assist the agency in communications and inspection scheduling. No device may be imported or offered for import unless it is the subject of a device listing and was manufactured at a registered establishment. The establishment registration and listing information must be in English. Under 21 CFR 807.41, the foreign establishment must also submit the identity of each known importer of its devices, and each person who imports or offers for import those devices, together with the premarket submission number for the devices involved.

The registration calendar is fixed: initial registration within 30 days of entering into the operation, and annual registration for each fiscal year during the period beginning on October 1 and ending December 31 (21 CFR 807.22). The fiscal year runs from October 1 to September 30 (21 CFR 807.3(u)).

What the annual fees look like

Fees are set for each fiscal year and published in the Federal Register. The figures below are the FY 2027 rates published on 30 July 2026 and apply to the fiscal year that began on 1 October 2026.

Fee type FY 2027 standard FY 2027 small business Cite
Annual establishment registration fee $13,785 $13,785 FY 2027 fee notice
510(k) premarket notification submission $28,653 $7,163 FY 2027 fee notice
Annual fee for periodic reporting on a class III device $22,286 $5,572 FY 2027 fee notice
513(g) request for classification information $8,596 $4,298 FY 2027 fee notice

Three features of the structure matter more than the numbers. The registration fee is payable by every establishment that is registered or required to register, and the published table shows no small business reduction for it. The small business rates for submissions depend on qualifying in advance: a business with gross receipts or sales of no more than $100 million for its most recent tax year may qualify for reduced fees, and one at or below $30 million may also qualify for a waiver of the fee for its first premarket application. Qualification materials must reach the agency at least 60 days before the fee is due. And the rates reset each fiscal year — the FY 2026 notice published on 30 July 2025 set the registration fee at $11,423 and the 510(k) fee at $26,067, which shows how quickly a budget line moves.

Front and back views of a blue disposable isolation gown with close-up detail panels showing the cuff and back reinforcement
Surgical apparel splits across two sections: surgical masks and surgical gowns are class II special controls, while other apparel such as caps, hoods and shoe covers is class I.

Five errors that cost Canadian manufacturers time

  1. Treating registration as clearance. Registration and listing are administrative; a registration number is not a finding of safety or effectiveness, and referring to registration in a way that suggests otherwise is addressed in 21 CFR 807.39.
  2. Assuming a quality system certificate covers the product. An ISO 13485-based certificate supports the quality system requirement; it does not clear a device that needs premarket notification.
  3. Missing the 90-day lead time. The submission has to arrive at least 90 days before commercial distribution, which changes a launch calendar.
  4. Naming more than one United States agent, or naming an agent that has no place of business in the United States. The regulation allows exactly one, and requires a United States presence.
  5. Building the specification in Canada without a written responsibility split. If a Canadian party initiates specifications for a device made by a second party, it is inside the registration and quality system rules rather than outside them.

Sources

Related reading

Frequently Asked Questions

Do I need a United States agent if I manufacture in Canada?

Yes, if your establishment has to register. Under 21 CFR 807.40 each foreign establishment required to register must submit the name, address and phone number of its United States agent, and may designate only one agent, which must reside or maintain a place of business in the United States.

Is a biological indicator exempt from premarket notification?

The biological sterilization process indicator is a class II device under 21 CFR 880.2800(a), and that section states no exemption from premarket notification. Confirm the product code and any applicable exemption before assuming a submission is unnecessary.

What is the difference between registration and 510(k) clearance?

Registration and device listing are establishment obligations under 21 CFR 807 subpart B. A 510(k) premarket notification is a submission about a specific device, required under 21 CFR 807.81 in defined circumstances such as a first introduction or a significant change.

When is the annual registration fee due?

Annual registration takes place during the period beginning on October 1 and ending on December 31 of each fiscal year (21 CFR 807.22). The fee that applies is the rate published for that fiscal year; the FY 2027 registration fee is $13,785.

Does an examination gown need a premarket notification?

Under 21 CFR 880.6265 the examination gown is a class I device that is exempt from premarket notification subject to the limitations in 21 CFR 880.9. If the gown is not labelled or otherwise represented as sterile, it is also exempt from the quality system requirements, except for records and complaint files under 21 CFR 820.35.

Which quality system standard does 21 CFR 820 require now?

The part requires a documented quality management system complying with the applicable requirements of ISO 13485, which is incorporated by reference, together with other applicable requirements of the part. Specific regulatory add-ons such as unique device identification, traceability, complaint reporting and advisory notices attach to named ISO 13485 clauses.

Can a distributor register on my behalf?

A wholesale distributor that does not manufacture, repackage, process or relabel a device is not required to register under 21 CFR 807.20(c). The foreign establishment's own registration and listing obligations, including the United States agent requirement, remain with that establishment.

Supplying Canadian clinics and laboratories — MDEL #35334. CliniEco Medical supplies biological indicators, sterilization packaging, gloves, masks and gowns, and maintains the documentation buyers ask for during supplier verification.

Records, purchasing and compliance support

Printable records: the sterilization log sheet generator and record downloads provide autoclave logs, spore test records and traceability fields in one place. The same log and record centre is used by dental, laboratory and long-term care teams preparing for an inspection.

Institutional orders: set up an account through wholesale and institutional supply, or request a sterilization compliance consultation when you need documentation and supplier evidence reviewed. Product families, standards and record templates sit under the sterilization compliance hub. Reselling? become a distributor.

Vous travaillez déjà avec des cliniques ? Devenez représentant régional.