Short answer: A Bowie-Dick test pack is a Type 2 chemical indicator that proves air was removed and steam reached the centre of the chamber in a pre-vacuum sterilizer. Canadian clinics buy them by the box of six and run one at the start of every day the sterilizer is used, plus after any repair, seal change or failed test. At five days a week and fifty working weeks, a single sterilizer needs roughly 250 packs a year.
CliniEco supplies a
Bowie-Dick test pack built to ISO 11140-5 Type 2 in boxes of six, which is the size most single-sterilizer dental and medical clinics reorder. It is a consumable, not a monitoring programme: the pack answers one question about the chamber, and the rest of the record comes from the chemical indicator on each pouch and the biological indicator run that day.
Where can a Canadian clinic buy Bowie-Dick test packs in case quantities?
Order by specifying four things, because suppliers price and ship them differently:
- Indicator type — Type 2 per ISO 11140-5 is the Bowie-Dick format; a Class 4 or Class 5 strip is not a substitute.
- Pack format — single-use disposable sheet packs versus reusable towel packs that a clinic assembles itself.
- Packs per box — six is the common Canadian box size, which suits one sterilizer for a working week.
- Storage — keep packs in the closed box, away from steam lines, heat and solvent fumes, and check the expiry date on receipt.
Clinics ordering for more than one sterilizer usually move to a full case and hold two to three months of stock, because a failed test stops the sterilizer until it is resolved.
How many Bowie-Dick packs does one sterilizer use in a year?
| Sterilizer use | Packs per week | Packs per year (50 weeks) | Boxes of six |
|---|---|---|---|
| 5 days a week, one pre-vacuum unit | 5 | 250 | 42 |
| 6 days a week, one unit | 6 | 300 | 50 |
| 5 days a week, two units | 10 | 500 | 84 |
| Repairs, seal changes, retests (allowance) | — | 10 to 20 | 2 to 4 |
The count is driven by days of use, not by the number of loads. A clinic that runs six cycles on a Monday still needs one pack that morning.
Does a passing Bowie-Dick test prove the load is sterile?
No. The pack proves air removal and steam penetration, nothing more. Sterility assurance comes from the biological indicator, which is why the two tests are logged side by side. In Ontario the RCDSO requires a biological indicator for each sterilizer on every day it is used — daily monitoring, not weekly. The weekly baseline is the CDC / AAMI ST79 standard in most US states, and clinics that copy that rhythm into an Ontario office leave gaps in the record.
What should happen when a pack fails?
Stop releasing loads from that unit, quarantine anything processed since the last passing test, check the door gasket and drain strainer, then repeat the test before the sterilizer returns to service. Record the failure, the corrective action and the repeat result on the same log page as the routine tests, so an inspector can follow the sequence without asking questions.
Download the printable sterilization log sheet, run the sterilization self-check before an inspection, and if a cycle has already failed, ask a compliance specialist before releasing the load. Clinics setting reorder points can use the clinic supply planner, and buyers comparing monitoring consumables can 24-hour biological indicator 25-pack.
Related reading
What does Class B mean on a benchtop autoclave?; Class 5 integrators in gravity displacement cycles; sterilization compliance hub.
Frequently Asked Questions
Is a Bowie-Dick test the same as a biological indicator?
No. A Bowie-Dick pack shows that air was removed and steam penetrated the load; a biological indicator uses live spores to show that the cycle actually killed them.
How often should a Bowie-Dick test be run?
Run one at the start of every day the pre-vacuum sterilizer is used, and again after a repair, a seal change or any failed test.
Where should the pack sit inside the chamber?
Place it on the bottom shelf toward the front of the chamber in an otherwise empty load, following the sterilizer manufacturer’s loading instructions.
Can a passing pack be reused?
No. Disposable Type 2 packs are single use, and a pack that has already been through a cycle cannot be read a second time.