Sterilization Compliance Kits & Monthly Supply Pack

These five options are the monitoring and packaging supply that a clinic's steam sterilization cycle is released on — biological indicators, Bowie-Dick test packs, Class 5 integrators, pouches, indicator tape and gloves — grouped so one order covers one sterilizer for about a month. Two of the five — the Assistant Kit and the Clinic Starter Bundle — stay available for a clinic that wants the incubator or the gloves in the same carton; the monthly pack is the current entry point, and Kit A and Kit B remain the option when only the compliance core is needed.

2 products
More about Sterilization Compliance Kits & Monthly Supply Pack

Choose on two facts about the machine, not on the size of the practice: whether the chamber pulls a vacuum (pre-vacuum, Class B) or works by gravity, and how many chambers you run. A vacuum machine needs a daily air-removal test and a gravity machine does not, and that difference decides between Kit A and Kit B. If you want the incubator and the record templates in the same carton, the Assistant Kit carries them; if you would rather the core be replenished without counting it yourself, the Monthly Pack does that.

Option What is in it How far one covers Who it suits
Kit A — pre-vacuum (Class B) 24-hour biological indicators (BIs) 50; Bowie-Dick Type 2 6-packs × 4; Class 5 integrators 100; Class 4 pouches 4 × 200; tape 3 rolls 25 testing days; 24 Bowie-Dick tests One pre-vacuum unit, a month
Kit B — gravity 24-hour BIs 50; Class 5 integrators 100; Class 4 pouches 4 × 200; tape 3 rolls; no Bowie-Dick 25 testing days; 800 pouches One gravity unit, a month
Assistant Kit — bundle 24-hour BIs 25; Bowie-Dick packs 6; Class 5 integrators 100; Class 4 pouches 400 + 50 m roll; tape 3 rolls; 55–60 °C incubator; log templates 25 spore tests; 400 pouches; incubator included Wants the incubator in the carton
Monthly Pack — 3 tiers Fixed core: Bowie-Dick 6-packs, 25-strip BIs, Class 5 integrators. Adjustable: pouches, nitrile gloves, Level 2 masks 1 / 2 / 3 sterilizers at ~4 / 8 / 12 cycles a day Core replenished monthly
Clinic Starter Bundle Nitrile gloves 1,000; 24-hour BIs 25; Class 4 pouches 200; tape 3 rolls About a month of daily spore testing; no incubator First sterilization cupboard

Which kit fits a gravity autoclave?

Kit B, and the reason is subtraction: a gravity unit has no air-removal step to verify, so the four Bowie-Dick packs that fill out Kit A would sit in the drawer. Everything else is the same — 50 24-hour biological indicators (one test plus one control per run day), 100 Class 5 integrators, 800 pouches and three-roll tape. The Assistant Kit also carries Bowie-Dick packs, so a gravity-only clinic fits it only if a pre-vacuum unit is coming.

What actually changes between a pre-vacuum cycle and a gravity one?

One test. On a pre-vacuum sterilizer the first cycle of each day is a Bowie-Dick test to ISO 11140-5 Type 2; a 6-pack covers six working days and Kit A ships four, for 24 test days. It measures air removal — the thing a vacuum pump does and a gravity chamber does not. Daily spore testing with a 24-hour indicator, incubated at 58 ± 2 °C before reading, and Class 5 integration inside every tray are common to both.

How do the Starter, Practice and Clinic tiers differ?

Only in how much of the fixed core they carry. The core is the same at all three — Bowie-Dick 6-packs, 24-hour BIs in 25-strips, Class 5 integrators — and is not substitutable, because each item answers a different release question. Quantity follows sterilizer count and daily cycles, not practice size; pouches, gloves and masks are the adjustable part.

Tier Sterilizers and cycles Fixed core each month What you can adjust
Starter 1 sterilizer, ~4 cycles a day Bowie-Dick 6-pack × 4; 24-hour BIs 25-strip × 1; Class 5 integrators 200/box × 1 Pouches 3.5" × 9" × 30 boxes; gloves × 1 case; masks × 1 box
Practice 2 sterilizers, ~8 cycles a day Bowie-Dick 6-pack × 8; 24-hour BIs 25-strip × 2; Class 5 integrators × 1 Pouches 3.5" × 9" × 45 boxes; gloves × 1 case; masks × 8 boxes
Clinic 3 sterilizers, ~12 cycles a day Bowie-Dick 6-pack × 11; 24-hour BIs 25-strip × 3; Class 5 integrators × 2; Bowie-Dick case × 1 Pouches 2.75" × 9" × 45 boxes and 3.5" × 9" × 52 boxes; gloves × 1 case; masks × 10 boxes

Does a monthly pack save anything over ordering item by item?

Three things change and one does not. The pack ships free every month, where item-by-item ordering in small batches carries a shipping charge, and it is billed below the same items bought separately. The counting comes from your sterilizer count and cycle volume, so you hear from us before an indicator box runs out — the price of the goods does not change. You can move between tiers or stop at any time: a monthly pack, not a term contract.

What is inside one carton, and which of these runs out first?

The biological indicators. Two vials a day — one test and one control — is 25 sterilization days from a 50-vial box, while the 800 pouches in the same carton are counted on a 3–4 chair practice using 400–750 a month, and the tape set goes by wrapped tray rather than calendar day. The box to re-check before a heavy month is the indicator box, not the packaging — a missing indicator box is an audit finding rather than a stockout.

That box carries the question the others cannot answer. The biological indicator says whether the spores were killed; the Class 5 strip whether time, temperature and steam reached the pack interior; the Class 4 pouch whether the wrapper was exposed; the tape whether the wrap stayed closed and went through; the Bowie-Dick pack whether the air came out.

What records come with each order?

Every shipment carries its lot trail: each lot ships with its own records — biological indicator lot and expiry, Bowie-Dick lot, and Class 5 integrator lot — so the sterilization log can be completed from what is in the box. Copy the number off the box and the entry writes itself. The daily record template is free and needs no account, with fields for the date, the cycle, the sterilizer, the vial lot number and the result.

What does an inspector look for?

One line per cycle day: the date, which sterilizer, the load or cycle number, the biological indicator lot with its control result, the integrator reading from the pack, and the initials of whoever reads it. An auditor follows that trail from the other end, and a gap in the log has to be explained by the practice — the obligation to complete the record stays with the practice whatever it buys. What a supplier carries is the other half: the article on the tray, its lot and expiry legible on the box, so the entry is made at the cycle rather than reconstructed at month end.

Not sure which matches your sterilizer? The sterilization self-check takes your sterilizer type and cycle volume and prints the answer, without an account. The five are Compliance Kit A, Kit B, the Assistant Kit, the Monthly Pack and the Clinic Starter Bundle. CliniEco Medical imports them under MDEL #35334.