The word sterile on a bag of pipette tips, tubes or plates is a process claim, and it means something specific: a validated sterilization process was applied to that lot, and the probability of a viable microorganism on an item after the process is at or below a stated level. The label usually does not tell you which level, which process or which standard. This article reads the claim clause by clause, so that a buyer can ask for the document behind it.
If you are buying a laboratory item sold as sterile, the documentation standard below is the one to ask a supplier for, whether the item is a tip, a tube or a plate, and the label does not usually tell you which of those documents exists.
What does SAL actually mean?
SAL is the sterility assurance level: the probability that a single item in the lot carries a viable microorganism after sterilization. A SAL of 10⁻⁶ means one item in a million is expected to be non-sterile; a SAL of 10⁻³ means one in a thousand. The number is a property of the validated process, not of the plastic, and it is established by the validation work described in the sterilization standard.
For medical devices, ISO 11137-1 requires a SAL of 10⁻⁶ unless a different level is justified and documented. Laboratory plasticware is often not a medical device in the regulatory sense, and a manufacturer may choose a SAL that fits the use, often 10⁻³. That is not a defect; it is a different claim, and the buyer needs to know which one is being made.
| Claim | What it means | Typical context |
|---|---|---|
| SAL 10⁻⁶ | One item in a million expected non-sterile | Medical devices under ISO 11137-1 |
| SAL 10⁻³ | One item in a thousand expected non-sterile | Laboratory consumables where the use allows it |
| Sterile, level not stated | A validated process was applied; the level is unknown to the buyer | Common on consumer-style lab packaging |
| Sterility assurance level not claimed | No SAL statement; the item may be clean, not sterile | Bulk bags without a sterile claim |


Why do lab plastics often claim 10⁻³?
Because the consequence of a non-sterile item is different. A pipette tip used to transfer a sample into a tube does not enter a sterile body cavity, and the laboratory's method usually includes its own controls. A manufacturer can therefore validate to a level that matches the use rather than to the medical device default, and the lower target allows a lower dose or a shorter cycle.
The buyer's question is not whether 10⁻³ is good enough in the abstract. It is whether the level claimed matches the use. For cell culture, for sterile media preparation, or for any workflow where the consumable is the only barrier, a 10⁻⁶ claim with documentation is the defensible choice.
How does ISO 11137-2 set a dose?
ISO 11137-2 describes how to establish the radiation dose for a given product and bioburden. The methods differ in what they assume and what they require in evidence. Method 1 establishes a dose from a bioburden estimate and a defined verification dose experiment. Method 2 uses a selected dose, and the VDmax approach substantiates a specific dose: VDmax25 substantiates 25 kGy, and VDmax15 substantiates 15 kGy for products whose bioburden and use support it.
The practical point for a buyer is that the dose is not chosen for convenience. It follows from the bioburden, the product's radiation tolerance and the SAL target, and it is supported by a validation report that names the method used. If a supplier cannot say which method substantiated the dose on the certificate, the claim is weaker than it looks.
| Label or certificate says | What you should be able to produce |
|---|---|
| Sterilized by gamma irradiation | Dose range, method used to set it, and the validation report reference |
| VDmax25 | Bioburden data and the substantiation report for 25 kGy |
| VDmax15 | Bioburden data, product justification, and the substantiation report for 15 kGy |
| SAL 10⁻⁶ | The dose-setting report that shows the chosen dose achieves it |
| Sterilized by EO | Cycle parameters, routine monitoring records and aeration details |
| Sterile, no method stated | Ask for the method; without it the claim cannot be audited |
What does an EO mark tell you?
Ethylene oxide sterilization is covered by ISO 11135. The standard addresses the validation of the process, routine monitoring and control, and the requirements that follow from the toxic nature of the gas. The part that matters on a lab consumable is residuals: EO and its reaction products remain in the material after the cycle and have to be reduced by aeration to a level that is safe for the intended use.
An EO mark therefore implies two documents, not one: a validation and routine-control record for the sterilization cycle, and a residuals assessment for the material. A supplier that can produce only the first has not finished the job. Where the product touches cells or media, the residuals question is the one to ask first.
What should a certificate of analysis contain?
A useful certificate names the product and lot, the sterilization method, the dose or cycle parameters, the SAL or the standard used to set it, the date of sterilization and the release decision. It should be traceable to the lot on the bag, not to a generic product page.
For a laboratory working to ISO 15189:2022, that certificate is part of the receiving record. Keeping it with the lot number is what allows the laboratory to answer the question "how do you know this box was sterile" without contacting the supplier.
Which claims are not sterility claims?
Several claims sit next to sterile on the same label and are often read as if they were part of it. DNase/RNase-free describes an assay for nucleases, not the absence of microorganisms. Non-pyrogenic describes endotoxin testing. Clean or virgin material describes the absence of added substances. None of them tells you whether the item was sterilized, and a sterile item is not automatically free of nucleases.
Read the label as a list of separate claims, each with its own test. A supplier that can name the test behind each claim is giving you something to compare; a supplier that bundles them into one word is not.
How do you audit a supplier's claim?
Ask three questions and keep the answers: which standard was used, which method within that standard set the dose or cycle, and which lot document carries the result. If the answers are consistent across lots, the claim is under control. If the method changes between lots without a change note, the process is not under control, whatever the certificate says.
Where the consumable is used in a regulated laboratory, the answers belong in the supplier file. In Canada, sterile laboratory consumables are distributed under the establishment licence held by the importer or distributor; CliniEco Medical distributes under MDEL #35334. In the United States, general-purpose laboratory equipment falls under 21 CFR 862.2050, and the sterilization of a consumable is a supplier process rather than an FDA clearance for that lot.
What belongs in the purchase table?
List the sterile status, the sterilization method, the SAL if stated, the standard used, the packaging format (bulk or individually wrapped), and the lot documentation that will be supplied. Add the storage condition, because a sterile barrier that is damaged in storage is no longer a barrier.
The current listing is in the laboratory consumables collection. Buyers comparing documentation across product families can read the PCR tube rules in Canada and the US and the standards and labelling question for microcentrifuge tubes.
Where does a lab take this next?
Three things a buyer can act on rather than read: a printable compliance log sheet that records receiving dates, lot numbers and open-on dates and exports from the browser; the clinic supply planner that turns bench counts into monthly quantities; and, where a sterile claim has to be documented for an auditor, Ask your compliance question which returns a written answer citing the regulation or standard it rests on.
Facility and multi-site buyers can Wholesale and multi-site ordering for case pricing and account setup. Reselling? become a distributor.
Related Reading
- Microcentrifuge tubes in Canada vs the US: standards and labelling
- PCR tubes in Canada vs the US: which specimen rules apply
- Pipette tips in Canada vs the US: which device class applies
Frequently Asked Questions
Is SAL 10⁻³ sterile enough for a laboratory?
It can be, if the use allows it and the method includes its own controls. For cell culture or sterile media preparation, ask for a 10⁻⁶ claim with documentation.
What is the difference between VDmax25 and VDmax15?
Both substantiate a selected dose under ISO 11137-2. VDmax25 substantiates 25 kGy; VDmax15 substantiates 15 kGy where the bioburden and product justify it.
Does gamma sterilization make tips DNase-free?
No. Sterility and nuclease status are separate claims with separate tests. Radiation does not replace a nuclease assay.
What does an EO mark require?
A validated cycle under ISO 11135, routine monitoring records, and a residuals assessment after aeration. Ask for both documents.
What should I keep with a sterile lot?
The certificate of analysis, the lot number and the receiving date. Together they let you answer how the sterile status was established.
Where can I see sterile and non-sterile lab consumables?
Our laboratory range is listed in the laboratory consumables collection, and facility buyers can request case pricing through wholesale ordering.
Sources
- ISO 11137-1:2025, radiation sterilization of health care products — evs.ee
- ISO 11137-2:2013, establishing the sterilization dose — evs.ee
- ISO 11135:2014, ethylene oxide sterilization — evs.ee
- ISO 11607-1:2019, packaging for terminally sterilized devices — evs.ee
- ISO 14937:2009, sterilizing agents and general requirements — evs.ee
- ISO 15189:2022, medical laboratories — evs.ee
- Medical Devices Regulations, SOR/98-282 — laws-lois.justice.gc.ca
- 21 CFR 862.2050, general purpose laboratory equipment — law.cornell.edu
- Health Canada, medical devices — canada.ca
- Health Canada, medical device guidance documents — canada.ca
- CLSI, standards portfolio — clsi.org
CliniEco Medical distributes clinic and laboratory consumables in Canada and ships across Canada and the US under MDEL #35334. Requirements that bind a specific laboratory remain the ones in its own licence, its quality programme and its written procedure.
Last updated: 23 November 2026.