Quick facts
- A plasma viscometer measures the resistance of a plasma sample to flow, so that a viscosity reading can be reported for a clinical purpose.
- In the United States it is filed under 21 CFR 862.2920, Plasma viscometer for clinical use, at Class I, product code JJL.
- In Canada the instrument is a non-invasive device at Class I under Rule 7(1) of the Medical Devices Regulations.
- Rule 7(2) raises it to Class II if it is intended as a calibrator, tester or quality control support to another medical device, or if it is connected to an active device already classified II, III or IV.
- A viscometer that only reports a raw reading and a viscometer delivered as part of a diagnostic system are not the same device on paper, even when the bench looks identical.
- Entries and rule text were checked against the current eCFR text, the Regulations and the openFDA classification database on 2026-10-11.
A plasma viscometer measures how readily a plasma sample flows, and it is used where a viscosity reading carries clinical meaning. In the United States it is a named Class I entry; in Canada the instrument is a non-invasive device whose class depends on how it is used and what it is attached to. This article covers both routes, the two situations that raise the Canadian class, and the laboratory standards a bench reads when it specifies an instrument of this kind.
What is a plasma viscometer?
A viscometer measures the internal resistance of a fluid to flow. In the clinical laboratory the sample is plasma, and the instrument reports either a relative viscosity or a value derived from the time a fixed volume takes to pass a capillary or to be sheared by a rotating element. The result is a number that a clinician reads alongside other results, and the instrument produces it from a sample that never returns to the patient.
That last point describes the device family. A viscometer is bench equipment: it holds a sample, measures it and reports a value. It does not enter the body and does not contact the patient, so in both countries the equipment is looked at as a laboratory instrument first, and the test that uses it is looked at separately.
Is a plasma viscometer a Class I device in the United States?
Yes. 21 CFR 862.2920 names a Plasma viscometer for clinical use and classifies it at Class I (general controls). The openFDA device classification API lists the instrument under product code JJL. A Class I device of this kind reaches the market under general controls: establishment registration and listing, labelling and good manufacturing practice, rather than a premarket review of the individual instrument.
The US entry is narrow and specific: it is written for a plasma viscometer for clinical use rather than for viscometers in general. That is characteristic of a named-entry system, and it is why a search by regulation number is the quickest route to the US answer.
How does Canada classify a plasma viscometer?
Canada looks at the device rather than the name. The instrument does not enter the body and does not contact the patient, so it is a non-invasive device and Rule 7(1) of the Medical Devices Regulations applies, placing all other non-invasive devices at Class I.
Rule 7(2) is the exception that matters. It raises a non-invasive device to Class II if the device is intended to act as a calibrator, tester or quality control support to another medical device, or if it is intended to be connected to an active device that is itself classified II, III or IV. A viscometer used as a stand-alone measuring instrument stays at Class I; one shipped as the measuring head of a larger analyser is read with the analyser.
| Item | Canada | United States |
|---|---|---|
| Named entry | None; classified by rule | 21 CFR 862.2920, plasma viscometer for clinical use |
| Product code | Not applicable | JJL |
| Default class | Class I (Rule 7(1)) | Class I (general controls) |
| Lift to Class II | Rule 7(2)(a) or 7(2)(b) | Not by class |
| Route to market, Class I | Establishment licence; no device licence | General controls; registration and listing |
| Route to market, Class II | Device licence under section 32 | Not applicable |
When does a viscometer become Class II in Canada?
The two triggers in Rule 7(2) are both about the role the instrument plays on the bench. The first is the calibrator, tester and quality control support limb. An instrument sold to set, check or verify another medical device is treated as part of that device's quality chain. The second is the connection limb: an instrument intended to be joined to an active device already classified II, III or IV is drawn up with the device it serves.
| How the instrument is used | Canadian rule | Canadian class | US entry |
|---|---|---|---|
| Stand-alone viscosity reading | Rule 7(1) | Class I | 862.2920, Class I |
| Bench research instrument, no clinical claim | Rule 7(1) | Class I | 862.2920 if clinical |
| Sold to calibrate or verify another device | Rule 7(2)(a) | Class II | 862.2920, Class I |
| Measuring head of a Class II analyser | Rule 7(2)(b) | Class II | 862.2920, Class I |
The lesson is the same one that runs through the laboratory family: the equipment answer and the assay answer are different answers, and the system answer can be higher than either. A laboratory should specify the instrument by the task it will perform, and keep the intended use in the method it writes.
Which standards apply to a clinical viscometer?
The class says what the instrument is; the standards say how the laboratory must work with it. ISO 15189:2022 sets out the requirements for quality and competence in a medical laboratory. A manufacturer building laboratory equipment to a management system does so under ISO 13485:2016, and the risk file follows ISO 14971:2019.
For the method itself, laboratories read the Clinical and Laboratory Standards Institute documents, which carry validated-method and quality control material for measurement procedures. In Canada the provincial frame sits with Public Health Ontario, and the wider safety material with CCOHS.
How does a laboratory specify and record a viscometer?
Specify by the sample the bench handles and by the reading the clinician needs. A capillary instrument is chosen where a relative value is enough and the sample volume is small; a rotational instrument is chosen where shear behaviour matters. The specification should also name the calibration routine and the control material, because those decide whether the reading can be defended.
Record by the run. The consumables that travel with the bench - graduated sample cups as the specimen is received, cotton swabs for cleaning between runs and individually wrapped swabs where a single sterile swab is used - are restocked as a set. Multi-site laboratories order at case level through the wholesale and multi-site ordering page and the B2B wholesale collection. Reselling? become a distributor.
There is one more habit worth building. Because the Canadian class can rise with the way an instrument is joined to the rest of the bench, a laboratory should note in its method what the viscometer is attached to and what it is used to check. If the instrument is later moved from a stand-alone role into a calibrated role, or wired into a larger analyser, the intended use recorded for it changes, and with it the class. Writing that down when the equipment is installed saves a great deal of reconstruction later, and it is the sort of record a quality manager looks for first.
Sources
- openFDA device classification API
- Medical Devices Regulations, SOR/98-282, full text and Schedule 1
- US FDA, medical devices
- US FDA, medical device databases
- US FDA, overview of device regulation
- Health Canada, medical devices
- Health Canada, establishment licensing
- Standards Council of Canada
- CSA Group
- ISO 13485:2016, medical devices quality management
- ISO 14971:2019, risk management of medical devices
- CCOHS, OSH answers
- Clinical and Laboratory Standards Institute
- Public Health Ontario, infection prevention and control
- ISO 15189:2022, medical laboratories, requirements for quality and competence
- US CDC, laboratory quality
- World Health Organization
- 21 CFR 862.2920, Plasma viscometer for clinical use
- Clinical and Laboratory Standards Institute
- Public Health Ontario, infection prevention and control
A viscosity reading is only as good as the calibration record behind it. The printable blank calibration and control sheet at the compliance log centre carries columns for the instrument, the control lot, the acceptance range and the reviewer, and it prints from the browser without an account and no sign-up.
Where a bench calibrates several instruments on a cycle, the calibration log generator on that page lays out a sheet sized to the number of instruments and the interval the laboratory runs.
Have a question about your own facility? Send it in and you will get a written answer specific to your setup, with the regulation or standard it is based on cited. Ask a compliance specialist about your equipment and province.
Related reading
- Laboratory pH meters in Canada vs the US: which class applies
- Laboratory thermometers in Canada vs the US: which class
- Laboratory centrifuges in Canada vs the US: which class
CliniEco Medical supplies eye care, laboratory and dental consumables of the kind described in this article. MDEL #35334. This article is written for optometry, laboratory and dental procurement professionals and is not legal advice; the class of a device and the records kept by a practice belong to the manufacturer and to the requirements that apply to it.
Frequently Asked Questions
Is a plasma viscometer a Class I device?
In the United States it is Class I under 21 CFR 862.2920, Plasma viscometer for clinical use. In Canada it is a non-invasive device at Class I under Rule 7(1).
When does a viscometer become Class II in Canada?
Under Rule 7(2). It becomes Class II if it is intended to act as a calibrator, tester or quality control support to another medical device, or if it is intended to be connected to an active device already classified as Class II, III or IV.
What is the US product code for a plasma viscometer?
The openFDA classification database lists Plasma Viscometer For Clinical Use under product code JJL within regulation number 862.2920 at Class I.
Does a research viscometer fall under the same entry?
In the United States the entry is written for a plasma viscometer for clinical use. A research instrument with no clinical claim sits outside it. In Canada the class follows the device and its intended use, so a research instrument may be out of scope or Class I under Rule 7(1).
Does a Class II viscometer need a licence in Canada?
Yes. A Class II device needs a device licence under section 32, supported by a CAN/CSA-ISO 13485 quality management certificate. A Class I device needs no device licence of its own.
Which standard frames a clinical viscometer laboratory?
ISO 15189:2022 sets the requirements for quality and competence in a medical laboratory. ISO 13485:2016 covers a manufacturer's quality system and ISO 14971:2019 the risk file.
How should calibration be recorded?
With the instrument, the control lot, the acceptance range and the reviewer, on a sheet the laboratory keeps. The compliance log centre provides a printable calibration and control sheet and a generator sized to the laboratory.