Quick answer: an IQ document proves the sterilizer was installed as specified, an OQ document proves it performs across its operating range on an empty chamber, and a PQ document proves it sterilizes the loads you actually run. All three are usually delivered together, and a clinic that signs without checking the load list in the PQ is signing for work that was never done. Our 24-hour biological indicators are the usual challenge device inside both OQ and PQ cycles.
What belongs in the installation qualification?
Evidence that the unit is where the manufacturer expects it to be and supplied as specified: model and serial number, utility connections including voltage and drain, water supply and quality, chamber level, door operation, and the site conditions the manual requires. IQ is often treated as paperwork, but two of the most common long-term problems — poor feed water and a drain that cannot handle the discharge — are IQ findings.
What does the operational qualification actually test?
Empty-chamber performance across the range the unit will be used in. Expect leak-rate testing, a Bowie-Dick test for pre-vacuum units, temperature and pressure verification, and biological and chemical indicator challenges placed in the positions the validation protocol defines. If the OQ was run only at one cycle type and your clinic runs three, the document does not cover your operation.
| Document | Chamber state | What it proves | Clinic should check |
|---|---|---|---|
| IQ | Not applicable | Installed as specified | Utilities, water, drain, serial number |
| OQ | Empty | Performs across operating range | All cycle types used, BI/CI positions |
| PQ | Loaded | Sterilizes real loads | Load list matches your instrument sets |
What must the performance qualification contain?
The load list. PQ runs your actual instrument sets, at the maximum load and the most difficult configuration, with indicators in the positions hardest to penetrate. A PQ that validates a single lightly loaded tray does not cover a full cassette load. Look for the number of consecutive cycles per load type, the indicator placement map, and the acceptance criteria written before the run rather than after.
Who signs, and what does a signature mean?
The supplier signs for the tests it performed; the clinic signs acceptance. Acceptance means the clinic has read the load list and confirmed it reflects actual practice. If the clinic adds a new instrument set or a new cycle after validation, the PQ does not automatically cover it, and that gap is what auditors find.
Where do the results live afterwards?
With the equipment record, not in a supplier's portal alone. A clinic should hold its own copy of all three documents plus any revalidation after a repair that affects performance. A printable validation record cover sheet from our compliance log centre and the log generator keep the document set traceable when staff change.
Clinics preparing for a first validation, or adding a second sterilizer, can review wholesale and multi-site ordering for indicator and consumable volumes across sites, and the B2B wholesale collection lists the monitoring lines stocked for institutional buyers. A compliance review with our sterilization desk will check whether the load list in your PQ matches what your clinic actually runs. Reselling? become a distributor.
Related reading
- Autoclave Validation Frequency for Research Laboratories
- Documenting Sterilizer Maintenance for Inspection
- Preventive Maintenance Schedules for Steam Sterilizers
Frequently Asked Questions
Is IQ/OQ/PQ required by law in Canada?
It is required by accreditation and quality systems rather than by a single federal rule, and provincial expectations vary. Most clinics encounter it through accreditation, research contracts or manufacturer warranty terms.
Can one PQ cover every load type?
No. PQ should cover each load type at its maximum and most difficult configuration, with indicators in the positions hardest to penetrate.
What happens after a major repair?
Performance-affecting repairs require revalidation. Ask the service provider which parts of the OQ and PQ are repeated and get the results in writing.
Who should hold the validation documents?
The clinic, not only the supplier. Keep a complete copy with the equipment record so it survives supplier changes and staff turnover.