Two facts decide whether a clinic can buy a medical device lawfully: the class the device is placed in, and who holds the licence or clearance behind it. In Canada the class decides whether a device licence exists at all, and section 26 of the Medical Devices Regulations asks for it from the manufacturer rather than the seller. In the United States the class decides which premarket file exists — none, a 510(k), or an approval — while the establishment duties sit on the firm that registers. The carton looks the same on both sides of the border; the paper behind it does not.
The worked example in this article is a Class 4 self-sealing sterilization pouch, 200 pouches to a box, and the United States entry that covers a sterilization wrap is 21 CFR 880.6850 — Class II, performance standards, no exemption from premarket notification. One line is enough to show why a Canadian and a United States purchase order ask different questions about the same carton.
What actually decides the class of a device?
Not the material, not the price, and not the category name in a catalogue — the intended use the seller assigns. Canada states the rule in one sentence: medical devices are classified into one of Classes I to IV by the classification rules set out in Schedule 1, where Class I is the lowest risk and Class IV the highest (section 6), and where a device can be classified into more than one class, the class representing the higher risk applies (section 7).
Four rules carry most of a dental order. Rule 3(b) puts all surgical or dental instruments in Class I, which is why a scaler is not a licensed device. Rule 13(b) puts a device intended for sterilizing a medical device in Class II, which is why the sterilizer is. Rule 7(2)(a) puts a device intended to act as a quality-control support to another device in Class II, the paragraph that reaches a 24-hour steam biological indicator, 50 vials to a pack. Rule 2(2) returns a device placed in the oral cavity as far as the pharynx to Class I, which keeps a birch tongue depressor, 1,000 to a case on the simple side of the line.
The United States writes the same idea as a definition rather than a rule table. Classes I, II and III are defined by which controls are sufficient: general controls alone, general controls plus special controls, or premarket approval (21 CFR Part 860). The class of a specific device type is then published one entry at a time in the CFR parts that run from 862 to 892, and the entry is the place the answer is actually written down.
What does each route require before a clinic can buy?
| Route | Licence or clearance | Held by | What the premarket file contains | Importer duty | Written in |
|---|---|---|---|---|---|
| Canada, Class I | None — no device licence exists for Class I | Nobody; the manufacturer keeps objective evidence of sections 10 to 20 | No premarket review. The Minister may request an analysis after the fact (section 25) | Establishment licence to import or sell, unless a retailer or health care facility (section 44) | MDR ss. 6, 25, 44 |
| Canada, Class II | Device licence | The manufacturer (section 26) | Intended use, standards list, attestation, label copy, ISO 13485 certificate (section 32(2)) | Establishment licence; distribution records (section 52) | MDR ss. 26, 32(2), 44, 52 |
| Canada, Class III and IV | Device licence, plus a control number on the label | The manufacturer | As Class II, plus materials, sterilization method, study summary and clinical evidence (section 32(3)) | Establishment licence; distribution records | MDR ss. 21(1)(d), 26, 32(3) |
| US, Class I exempt | No 510(k) | Nobody files a premarket submission | None, but general controls apply: registration and listing, labelling, reports (21 CFR 860.3) | Foreign establishment registers, lists and names one US agent (807.40) | 21 CFR 860.3, 807.40, 872.4565 |
| US, Class II | 510(k) clearance | The owner of the submission (807.81) | Class, proposed labels, similarity and difference statement with data (807.87) | Registration and listing; US agent; reports under 803.50 | 21 CFR 807.81, 807.87, 880.6850 |
| US, Class III | Premarket approval | The applicant for the approval | An approved premarket approval application, device specific | Registration and listing; US agent; reports under 803.50 | 21 CFR 872.6730 |
Read the second and third columns together. In Canada the file sits with the manufacturer, so an importer cannot create one by asserting it. In the United States a 510(k) clearance belongs to whoever submitted it, and a distributor placing a device into commercial distribution under its own name may rely on another person’s submission only in the narrow case set out in 21 CFR 807.85(b). Neither system asks the clinic to hold a premarket file; both ask the clinic to name who does.
Who holds the paper in Canada?
Section 26 is the sentence a buyer should be able to recite: subject to section 37, no person shall import or sell a Class II, III or IV medical device unless the manufacturer holds a licence in respect of that device. The application is submitted by the manufacturer (section 32(1)). For Class II it must carry the intended use, the standards complied with, a senior official’s attestation that objective evidence exists, a copy of the label, and the quality management system certificate to CAN/CSA-ISO 13485 (section 32(2)). For Class III and IV it grows to include the materials, the sterilization method where the device is sold sterile, a summary of the studies relied on, and the clinical evidence the class requires (section 32(3)).
Holding a licence is not the same as being allowed to import. Section 44(1) says no person shall import or sell a medical device unless the person holds an establishment licence, and section 44(2) exempts a retailer, a health care facility, and — for Classes II to IV — the manufacturer. That is why a dental office buying from a distributor needs no establishment licence of its own, and why the exemption does not travel: a clinic that imports directly is importing, not buying. The guidance behind the licence is published as GUI-0016.
Section 44(3) adds the duty that catches private importing: a person who imports a medical device must ensure that the person they import from holds an establishment licence. The 2026 amendments narrow this at the edge rather than removing it — see the next section.
What changes on 1 December 2026?
SOR/2026-110, the Regulations Amending the Medical Devices Regulations (Establishment Licences), was registered on 4 June 2026 and, under its section 10, comes into force on the 180th day after registration. The Regulatory Impact Analysis Statement published with it describes the same change as coming into force six months from registration, and Health Canada’s guidance documents carry their own effective dates; a supplier file that has to cite a date should state which of the three it is using. Three changes belong in a clinic’s buying file.
First, section 45 now asks for the name and address of any person other than the manufacturer who sells a device to the establishment, and the classes that person supplies (paragraphs 45(f.1) and (f.2)): a supplier list becomes a licensing input. Second, new section 58.1 requires the manufacturer, importer and distributor of a Class II, III or IV device to maintain documented procedures for handling, storage, delivery, installation, servicing and corrective action. Third, section 52(3), section 58, subsections 59(3) and (4) and 61.2(4) turn procedures that were previously attested into procedures that must exist and be maintained. The mechanics are set out in our companion article on supplier lists and written procedures under SOR/2026-110, with the Canada Gazette text of SOR/2026-110.
What does “exempt” mean in the United States?
Exempt from premarket notification is not exempt from regulation. A Class I device is subject to the general controls listed in the definition — adulteration, misbranding, registration and listing, banned devices, notification and other remedies, records and reports (21 CFR 860.3) — and its labelling obligations sit in 21 CFR 801. Where a 510(k) is required it is due at least 90 days before commercial distribution begins (807.81(a)), and the submission carries the class, the proposed labels, and a statement of similarity and difference against comparable products with the data behind it (807.87).
Two exemptions are commonly misread, and both are settled by the text of the entry rather than by the class. A dental hand instrument is Class I and exempt from premarket notification, but only where it is made of the same materials used before 28 May 1976 and only within the limitations of 21 CFR 872.9 (872.4565(b)). A non-powdered patient examination glove is also Class I, and only the finger cot is exempt (880.6250(b)) — so the glove box in the treatment room is a Class I device with a 510(k) behind it. The class and the exemption are two separate facts, printed in two separate sentences.
The establishment side is a third fact. A foreign establishment that manufactures a device imported into the United States must register and list, and must name one United States agent (21 CFR Part 807, 807.40(a) and (b)); no device may be imported unless it is listed and was made at a registered establishment (807.40(c)). A licensed practitioner who alters a device solely for use in his or her own practice is exempt from registration (807.65(d)), and a retail establishment making final delivery to the end user is exempt (807.65(e)) — the two exemptions that most closely mirror Canada’s section 44(2).


Where do the two systems disagree on the same device?
| Device | Canada | Canada rule | United States | US premarket |
|---|---|---|---|---|
| Dental hand instrument (scaler, curette) | Class I | Schedule 1, Rule 3(b) | Class I, 21 CFR 872.4565 | Exempt, if made of the same materials as before 1976 (872.4565(b)) |
| Steam sterilizer (autoclave) | Class II | Schedule 1, Rule 13(b) | Class II, 21 CFR 880.6880 | 510(k) |
| Dry-heat sterilizer, bench type | Class II | Schedule 1, Rule 13(b) | Class II, 21 CFR 880.6870 | 510(k) |
| Endodontic dry heat sterilizer, glass beads | Class II | Schedule 1, Rule 13(b) | Class III, 21 CFR 872.6730 | Premarket approval |
| Biological indicator | Class II | Schedule 1, Rule 7(2)(a) | Class II, 21 CFR 880.2800 | 510(k) |
| Chemical indicator tape | Class II | Schedule 1, Rule 7(2)(a) | Class II, 21 CFR 880.2800 | 510(k) |
The clearest disagreement is the glass-bead sterilizer. In Canada a device intended for sterilizing a medical device is Class II under Rule 13(b), whatever the heat source. In the United States a steam sterilizer (880.6880) and a dry-heat sterilizer (880.6870) are both Class II, but the endodontic dry heat sterilizer that supplies heat through glass beads is Class III and requires a premarket approval (872.6730). Same bench, same purpose, three levels of control depending on which entry the device is described under — and the description on the label and in the listing is what decides.
Importing from a supplier who markets a device in the United States does not import the American class with it, and a Canadian licence does not travel south. The document that does travel is the label.
What can a clinic check before the purchase order?
A Canadian label carries ten items (section 21(1)): the device name; the manufacturer’s name and address; the identifier; the control number for a Class III or IV device; the contents where not apparent; the word Sterile where a sterile sale is intended; the expiry date where one exists; the conditions, purposes and uses; the directions; and any special storage conditions. The device licence number is not one of them, which is why the practical check is the manufacturer’s name on the box searched against the licence listing.
Two registers do the rest. In Canada, the Medical Devices Active Licence Listing can be searched by manufacturer and device name for active licences and their class. In the United States, the FDA product classification database and the device classification guidance give the product code, the regulation number, the class and whether premarket notification applies — the same three facts a supplier should be able to put in writing.
The deliverable is one page per supplier: manufacturer legal name, device name as licensed, class, licence or clearance number, and the register where it was checked, with the date. The SOP package for imported medical devices we publish for imported devices is a starting template; the certifications page records the standards our own lines are made to. A supplier that cannot produce the four facts is not necessarily selling unlawfully, but the buyer cannot tell the two cases apart — the only question an inspector will ever ask.
Where does a clinic take this next?
Three things a clinic can act on rather than read: a printable compliance log sheet that records dates, lot numbers and stock movements and exports from the browser; the clinic supply planner that turns procedure or case volume into monthly quantities; and, where a finding has to be documented for an inspector, Ask your compliance question.
Have a question about your own facility? Send it in and you will get a written answer specific to your setup, with the regulation or standard it is based on cited. Ask your compliance question.
Facility and multi-site buyers can Wholesale and multi-site ordering for case pricing and account setup. Reselling? become a distributor.
Related Reading
- What Is an MDEL? Canada’s Medical Device Establishment Licence Explained
- Health Canada MDL vs MDEL: Which One Applies to Your Product or Your Import?
- FDA 510(k) Compared with Canada’s MDEL: What Buyers Must Check for Sterilization Pouches
- Device Classes I vs II: Health Canada Rules for Supplies
- Supplier Lists and Written Procedures Under SOR/2026-110
Frequently Asked Questions
Is a sterilization pouch a Class II device in Canada?
The class follows Schedule 1, not the shelf it sits on. In the United States a sterilization wrap is expressly Class II (21 CFR 880.6850). In Canada the answer depends on which rule the pouch is placed under, and it should come from the licence listing or from the manufacturer in writing rather than from the category name in a catalogue.
Does a clinic need an establishment licence to buy supplies?
No. Section 44(2) exempts a health care facility, and a retailer, from the establishment licence requirement when it imports or sells a device. The exemption belongs to the buyer’s role, so it stops applying the day the clinic starts importing on its own account rather than buying from a Canadian establishment.
Is a Class I device unregulated in Canada?
No. A Class I device has no device licence, but the manufacturer still has to meet the safety and effectiveness requirements in sections 10 to 20, the label still has to carry the section 21 items, distribution records are still required under section 52, and incident reporting under section 59 applies, including reports about incidents that occur outside Canada (section 59(1.1)).
Does an MDEL mean the device itself is licensed?
No, and this is the most common conflation in supplier paperwork. An establishment licence is held by the importer or distributor for its activities; a device licence is held by the manufacturer for a Class II, III or IV device. Holding the first says nothing about whether the second exists.
Where do I find the device licence number?
Not on the label — the licence number is not a section 21 element. Search the Medical Devices Active Licence Listing by manufacturer and device name. If the device is Class I there is nothing to find, because no device licence is issued for that class.
Do all Class I devices skip the 510(k) in the United States?
No. Some Class I devices are exempt from premarket notification, and some are not. A dental hand instrument is exempt only where it uses the same materials as before 1976 (21 CFR 872.4565(b)); a patient examination glove is Class I and only the finger cot is exempt (880.6250(b)). The entry decides, not the class.
CliniEco Medical is a licensed medical device establishment (MDEL #35334).